EN 61675-1-2014 en Radionuclide imaging devices - Characteristics and test conditions - Part 1 Positron emission tomographs.pdf
《EN 61675-1-2014 en Radionuclide imaging devices - Characteristics and test conditions - Part 1 Positron emission tomographs.pdf》由会员分享,可在线阅读,更多相关《EN 61675-1-2014 en Radionuclide imaging devices - Characteristics and test conditions - Part 1 Positron emission tomographs.pdf(44页珍藏版)》请在麦多课文档分享上搜索。
1、BSI Standards PublicationRadionuclide imaging devices Characteristics and test conditionsPart 1: Positron emission tomographsBS EN 61675-1:2014National forewordThis British Standard is the UK implementation of EN 61675-1:2014. It isidentical to IEC 61675-1:2013. It supersedes BS EN 61675-1:1998+A1:2
2、008which is withdrawn.The UK participation in its preparation was entrusted by Technical Com-mittee CH/62, Electrical Equipment in Medical Practice, to Subcommittee CH/62/3, Equipment for radiotherapy, nuclear medicine and radiation dosimetry.A list of organizations represented on this committee can
3、 be obtained onrequest to its secretary.This publication does not purport to include all the necessary provisions ofa contract. Users are responsible for its correct application. The British Standards Institution 2014.Published by BSI Standards Limited 2014ISBN 978 0 580 76378 6ICS 11.040.50; 35.240
4、.80Compliance with a British Standard cannot confer immunity fromlegal obligations.This British Standard was published under the authority of theStandards Policy and Strategy Committee on 31 July 2014.Amendments/corrigenda issued since publicationDate Text affectedBRITISH STANDARDBS EN 61675-1:2014E
5、UROPEAN STANDARD NORME EUROPENNE EUROPISCHE NORM EN 61675-1 June 2014 ICS 11.040.50 Supersedes EN 61675-1:1998 English Version Radionuclide imaging devices - Characteristics and test conditions - Part 1: Positron emission tomographs (IEC 61675-1:2013) Dispositifs dimagerie par radionuclides - Caract
6、ristiques et conditions dessai - Partie 1: Tomographes mission de positrons (CEI 61675-1:2013) Bildgebende Systeme in der Nuklearmedizin - Merkmale und Prfbedingungen - Teil 1: Positronen-Emissions-Tomographen (IEC 61675-1:2013) This European Standard was approved by CENELEC on 2013-10-30. CENELEC m
7、embers are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national standards may be obtained on applic
8、ation to the CEN-CENELEC Management Centre or to any CENELEC member. This European Standard exists in three official versions (English, French, German). A version in any other language made by translation under the responsibility of a CENELEC member into its own language and notified to the CEN-CENE
9、LEC Management Centre has the same status as the official versions. CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Ice
10、land, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom. European Committee for Electrotechnical Standardization Comit Europen de Normalisation Electrotechnique Europ
11、isches Komitee fr Elektrotechnische Normung CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels 2014 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members. Ref. No. EN 61675-1:2014 E BS EN 61675-1:2014EN 61675-1:2014 - 2 - Foreword The te
12、xt of document 62C/550/CDV, future edition 2 of IEC 61675-1, prepared by IEC/SC 62C, “Equipment for radiotherapy, nuclear medicine and radiation dosimetry“, of IEC TC 62, “Electrical equipment in medical practice “ was submitted to the IEC-CENELEC parallel vote and approved by CENELEC as EN 61675-1:
13、2014. The following dates are fixed: latest date by which the document has to be implemented at national level by publication of an identical national standard or by endorsement (dop) 2014-12-13 latest date by which the national standards conflicting with the document have to be withdrawn (dow) 2016
14、-10-30 This document supersedes EN 61675-1:1998. Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. CENELEC and/or CEN shall not be held responsible for identifying any or all such patent rights. Endorsement notice The text of the In
15、ternational Standard IEC 61675-1:2013 was approved by CENELEC as a European Standard without any modification. BS EN 61675-1:2014- 3 - EN 61675-1:2014 Annex ZA (normative) Normative references to international publications with their corresponding European publications The following documents, in wh
16、ole or in part, are normatively referenced in this document and are indispensable for its application. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. NOTE When an international publicatio
17、n has been modified by common modifications, indicated by (mod), the relevant EN/HD applies. Publication Year Title EN/HD Year IEC/TR 60788 2004 Medical electrical equipment - Glossary of defined terms - - BS EN 61675-1:2014 2 61675-1 IEC:2013 CONTENTS INTRODUCTION . 6 1 Scope 7 2 Normative referenc
18、es 7 3 Terms and definitions 7 4 Test methods . 13 4.1 General 13 4.2 SPATIAL RESOLUTION . 13 4.2.1 General 13 4.2.3 Method . 14 4.2.4 Analysis 15 4.2.5 Report 17 4.3 Tomographic sensitivity 18 4.3.1 General 18 4.3.2 Purpose 18 4.3.3 Method . 18 4.3.4 Analysis 19 4.3.5 Report 20 4.4 Uniformity 20 4.
19、5 Scatter measurement . 20 4.5.1 General 20 4.5.2 Purpose 20 4.5.3 Method . 20 4.5.4 Analysis 21 4.5.5 Report 22 4.6 PET COUNT RATE PERFORMANCE 23 4.6.1 General 23 4.6.2 Purpose 23 4.6.3 Method . 23 4.6.4 Analysis 24 4.6.5 Report 26 4.7 Image quality and quantification accuracy of source ACTIVITY co
20、ncentrations . 26 4.7.1 General 26 4.7.2 Purpose 26 4.7.3 Method . 27 4.7.4 Data analysis 31 4.7.5 Report 34 5 ACCOMPANYING DOCUMENTS 35 5.1 General 35 5.2 Design parameters . 35 5.3 Configuration of the tomograph 36 5.4 SPATIAL RESOLUTION . 36 5.5 Sensitivity 36 5.6 SCATTER FRACTION 36 5.7 COUNT RA
21、TE performance . 36 BS EN 61675-1:201461675-1 IEC:2013 3 5.8 Image quality and quantification accuracy of source ACTIVITY concentrations . 36 Bibliography 37 Index of defined terms 38 Figure 1 Evaluation of FWHM 16 Figure 2 Evaluation of EQUIVALENT WIDTH (EW) 17 Figure 3 Scatter phantom configuratio
22、n and position on the imaging bed . 19 Figure 4 Evaluation of SCATTER FRACTION . 22 Figure 5 Cross-section of body phantom 27 Figure 6 Phantom insert with hollow spheres . 28 Figure 7 Image quality phantom and scatter phantom position for whole body scan acquisition 29 Figure 8 Placement of ROIs in
23、the phantom background 32 BS EN 61675-1:2014 6 61675-1 IEC:2013 INTRODUCTION Further developments of POSITRON EMISSION TOMOGRAPHS allow most of the tomographs to be operated in fully 3D acquisition mode. To comply with this trend, this standard describes test conditions in accordance with this acqui
24、sition characteristic. In addition, today a POSITRON EMISSION TOMOGRAPH often includes X-RAY EQUIPMENT for COMPUTED TOMOGRAPHY (CT). For this standard PET-CT hybrid devices are considered to be state of the art, dedicated POSITRON EMISSION TOMOGRAPHS not including the X-ray component being special c
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- EN6167512014ENRADIONUCLIDEIMAGINGDEVICESCHARACTERISTICSANDTESTCONDITIONSPART1POSITRONEMISSIONTOMOGRAPHSPDF

链接地址:http://www.mydoc123.com/p-720374.html