ECA EIA-501-A-1990 Recommended Practice for the Measurement of X-Radiation from Raster-Scanned Direct-View Data Display Cathode-Ray Tubes.pdf
《ECA EIA-501-A-1990 Recommended Practice for the Measurement of X-Radiation from Raster-Scanned Direct-View Data Display Cathode-Ray Tubes.pdf》由会员分享,可在线阅读,更多相关《ECA EIA-501-A-1990 Recommended Practice for the Measurement of X-Radiation from Raster-Scanned Direct-View Data Display Cathode-Ray Tubes.pdf(18页珍藏版)》请在麦多课文档分享上搜索。
1、ANSI/EIA-501-A-1990 Approved: March 9,1990 Reaffirmed: February 28,2001 EIA STANDARD Recommended Practice for the Measurement of X-Radiation from Raster-Scanned Direct-View Data Display Cathode-Ray Tubes EIA-501-A (Revision of EM-501) MAY 1990 Electronic Components, Assemblies the TV X-I., or the Mo
2、del 49l,or equivalent. Thin window ion chamber survey meters may also be used. See ELA Consumer Products Engineering Bulletin No. 3 “Measurement Instnunentation for X-Radiation fiom Television Receivers“.) 3.1.2 X-Radiation Measuring Instrument The principal x-radiation measuring instrument shall co
3、mply with the requirements of Title 21, Code of Federal Regulations (CFR), Chapter I, Subchapter J, Section 1020.10, subparagraph (c)-(2), and with EIA Consumer Products Engineering Bulletin No3. The instrument shali provide x-radiation measurements (mRJh) or pA/Kg in the range of the applicable lim
4、it. The radiation volume shall have across-section parallel to the external surface of the hypothetical cabinet with an area of 10 cni! and no dimension larger than 5 cm. Measurement deviations due to instrument imprecision, energy dependence and other sources of error shall be commensurate with the
5、 current instrument art. An example of an instrument which hilfills these requirements is the 440 RF/C, or equivalent. Measurements with instruments having other areas must be corrected for spatial non-uniformity of the radiation field to obtain the exposure rate averaged over a 1 O cni! area. NOTE:
6、 A larger area detector may be used, prowed that the field is determined to be uniform over its aperture. Examples: 440 RF, 203A, and TV-150. 32 Test Equipment Stable electronic conditions must be provided for the operation of the tube under tbt and to measure anode voltage and current. 32.1 Anode V
7、oltage Metering System The recommended anode voltage divider and voltmeter with an overall accuracy of 2 0.1% and the digital readout using at least four significant tgures, having a sensitivity of at least 0.01 kV. me effect of the overall accelerating potential on the x-radiation characteristics i
8、s such that the x-radiation exposure rate could double for each 1 kV change of accelerator potential, thereby making accurate voltage measurements desirable.) If the recommended accuracy of the measurement system cannot be achieved, the inaccuracy shall be accounted for by reducing the recorded volt
9、age reading by the maximum possible measurement deviation. 32.2 Anode Current Metering System The recommended anode current metering system incorporates an overall accuracy of $I 1%. (The effect of the anode current on the x-radiation characteristics is that the x-radiation exposure rate is directly
10、 proportional to the anode current.) If the recommended accuracy of the anode current measurement system cannot be achieved, the inaccuracy shall be accounted for by reducing the recorded current reading by the maximum possible measurement deviation. EM-501-A Page 3 3.23 Power Supply for the Overall
11、 Accelerating Potential The recommended power supply for the overall acclerating potential incorporates a stabfiy of 0.05%. (The effect of the overall accelerating potential on the x-radiation characteristic iS according to paragraph 3.2.1, thereby making a stable power supply desirable.) If the rec
12、ommended shbility of the power supply cannot be achieved, the variation shall be accounted for by recording the lowest voltage reading during a specific x-radiation measurement. 32.4 Anode Connector Assembly See paragraphs 6.2 and 63. 4. CALIBRATION 4.1 The measuring instrument shall be calibrated b
13、y exposure to a uniform x-ray field having an exposure rate and energy representative of those to be measured. 4.2 The calibration shall be traceable to the National Institute of Standards and Technology (NIST). 5. TEST CONDITIONS 5.1 Tube Position The tube under test shall be positioned relative to
14、 the test equipment so that measurements may be made as specified in paragraph 6.0. 52 Extemg Components Ehternal components normaiiy mounted on the tube may be in place while obtainiag x-radiation characteristics. (Extemai components may indude deflection yokes, alignment coils, mounting hardware,
15、implosion protection mechanisms, etc.) 53 warm-up A warm-up period shali be provided for both the tube and test equipment in order to obtain stable operating conditions before proceeding with the measurement. Anode current shall not vary or drift more than $L 1% while measurements are bei made. 5.4
16、Display 5.4.1 Focus The tube under test shd be operated at optimum focus. 5.4.2 Blanking The tube shall be operated with dc voltage which result in an unb1anke-d raster. 5.4.3 Multiple Guns Multiple gun tubes may be operated using only one gun. . EIA-501-A Page 4 5.4.4 Raster Geometry The horizontal
17、 and vertid scan controls shall be adjusted so that the raster width and height are 90% of the screen width and height. The linearity of the scan shaii be checked and adjusted with a crosshatch pattern such that the &tance between crosshatch lines measured along the major and minor axes shaU be witb
18、iu 10% of the corresponding measurements at the center of the screen. The radius of mature of ali four raster sides shd not be less than 200 un. 6. PROCEDURE The x-radiation exposure rate shaU be measured as described below at the location of maximum intensity. The background radiation shall be veri
19、fied at the test position and corrections applied to the observed data as required. 6.1 Radiation from the Tube Face The radiation fiom the tube face shall be measured with the effective center of the detector 5 cm fiom the surface of the tube. NOTE: When the plastic spacers on the 440 RF/C Survey M
20、eter, or equivalent, are in contact with the tube face or the surface of the hypothetical cabmet, the distance from the surface to the effective center of the detector is 5 cm. The measured area of the tube face must be extended past the extremities of the actual faceplate to completely enclose the
21、effective volume of the hypothetid cabinet described in paragraph 6.4. This extension of the tube face is achieved by hypothetidy extending the major spherical curvature of the front face of the tube to the top, bottom and side surfaces of the hypothetical cabmet. The measuring instrument may be pla
22、ced anywhere over the tube face or over the hypothetid extension of it, provided that the center of the detectors front face does not overlap the top, bottom or sides of the hypothetical cabmet. 6.2 Radiation Emitted from the Funnel Portion Radiation emitted Erom the funnel portion of the entire tub
23、e (&clusive of the anode contad and tube face) shd be measured with the effective center of the detector 5 an from the surface of the hypohetica, non-absorbing cabinet that is desuii in paragraph 6.4 and shown ia Fie 2 The measuring instrument shai be positioned for maximum exposure rate with the fr
24、ont face of the detector pardel to the surface of the cabinet. The measuring instrument may be placed anywhere over the surface of the hypothetical cabinet desuibed in paragraph 6.4, provided that the center of the detectors front face does not overiap the outer edges of the surface of the hypotheti
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ECAEIA501A1990RECOMMENDEDPRACTICEFORTHEMEASUREMENTOFXRADIATIONFROMRASTERSCANNEDDIRECTVIEWDATADISPLAYCATHODERAYTUBESPDF

链接地址:http://www.mydoc123.com/p-704381.html