ECA EIA-364-109-2003 TP-109 Loop Inductance Measurement Test Procedure for Electrical Connectors (1 nH-10 nH)《TP-109 电连接器的电感(回路)测量试验程序(1 nH - 10 nH)》.pdf
《ECA EIA-364-109-2003 TP-109 Loop Inductance Measurement Test Procedure for Electrical Connectors (1 nH-10 nH)《TP-109 电连接器的电感(回路)测量试验程序(1 nH - 10 nH)》.pdf》由会员分享,可在线阅读,更多相关《ECA EIA-364-109-2003 TP-109 Loop Inductance Measurement Test Procedure for Electrical Connectors (1 nH-10 nH)《TP-109 电连接器的电感(回路)测量试验程序(1 nH - 10 nH)》.pdf(40页珍藏版)》请在麦多课文档分享上搜索。
1、 ANSI/EIA-364-109-2003 (2009) Approved: May 20, 2003 Reaffirmed: December 2, 2009 EIA STANDARD TP-109 LOOP INDUCTANCE MEASUREMENT TEST PROCEDURE FOR ELECTRICAL CONNECTORS (1 nH 10 nH) EIA-364-109 EIA-364-109May 2003 ELECTRONIC COMPONENTS, ASSEMBLIES see 2.2.4.1 for more detailed information. Unless
2、otherwise specified in the referencing document it is recommended that the following equipment settings be used: Smith chart format, set network analyzer to display inductance values, minimum of 401 measurement points, frequency span conduct both wideband and narrowband sweeps, no smoothing, averagi
3、ng set to 16 or higher. NOTE “Wideband” sweep is typically the full range of the network analyzer and “narrowband” sweep is over a limited range (for example 100 MHz wide). EIA-364-109 Page 7 4.3 Fixture measurement Position the probes to touch the interface pads of the test board characterization s
4、tructure. Measure and record the loop inductance of the fixture from the Smith chart at the frequency(s) of interest. NOTE A loop inductance vs. frequency graph may be generated through the use of data acquisition software and spreadsheet software, if specified in the referencing document. EIA-364-1
5、09 Page 8 4.4 Specimen measurement 4.4.1 Connect the probe to the fixture interface pad of the driven line with the specimen installed as shown in figure A.1. Terminate the far end of the driven line in an electrical short circuit to the return conductor(s). 4.4.2 Place the specimen a minimum of 5 c
6、m from any object that may introduce error into the measurement. 4.4.3 Measure and record the loop inductance over the specified test frequency range or discrete frequencies. 4.4.4 Calculate the specimen loop inductance by subtracting the fixture loop inductance, (see 4.3) from the specimen plus fix
7、ture loop inductance, see 4.4.3. NOTE If specified in the referencing document, a loop inductance vs. frequency graph may be generated through the use of data acquisition software and spreadsheet software. 4.4.5 If requested, repeat 4.4.1 through 4.4.4 on multiple conductors throughout the specimen.
8、 4.4.6 When additional measurements with different test frequencies or ranges are required perform the calibration step defined in 4.2, then repeat 4.4.1 through 4.4.5 as necessary. 5 Details to be specified The following details shall be specified in the referencing document: 5.1 Measurement freque
9、ncy range and/or discrete frequency(s) 5.2 Special requirements with respect to the fixture, and the short circuit, (see 2.2.1.4) construction and electrical properties of each. 5.3 Signal/ground pattern, including the number and location of signal and grounds. It is recommended that enough location
10、s within the specimen be measured to take into account the varying loop inductances within the specimen 5.4 Location of the drive signal connection point, location of the return signal connection point and connections to be made to adjacent pins, if any 5.5 Specimen environment impedance if other th
11、an 50 ohms 5.6 Plots, if desired, and Smith charts or loop inductance vs. frequency graphs EIA-364-109 Page 9 6 Test documentation Documentation shall contain the details specified in clause 5, with any exceptions, and the following: 6.1 Title of test 6.2 Test equipment used, and date of last and ne
12、xt calibration 6.3 Description of test fixture and associated calibration structures 6.4 Values and observations 6.5 Representative graphs, if available 6.6 Name of operator and date of test EIA-364-109 Page A-1 Annex A Loop inductance measurement setup (normative) A.1 Figure A.1 shows an example of
13、 a typical test equipment setup for loop inductance measurements including the equipment, cables, probes, and fixture. connectorfixturereferencestructureVector Network AnalyzerprobePort 1 Test cable driven linereturn pathshortcircuitFigure A.1 - Example of test equipment setup for loop inductance me
14、asurements EIA-364-109 Page A-2 A.2 Figure A.2 shows an example of the measurement set up for measuring the loop inductance of a fixture with a ground plane return path. The signal and ground vias are shorted together with a copper surface. The test fixture should be designed such that the signal tr
15、ace, pad interface, and copper surface provide the lowest inductance path from the probe tip to the signal and ground vias. This reference trace and pad interface should represent the same structures and geometries that will be used for the specimen measurement. Figure A.2 - Example of fixture loop
16、inductance measurement setup EIA-364-109 Page A-3 A.3 Figure A.3 shows an example of the measurement set up for measuring the loop inductance of the fixture plus specimen. The figure shows an example of the probe, pads, shorting block, and a specimen consisting of an edgecard connector. The ground p
17、ad is connected to the test board ground plane (not shown) that is in turn connected to the specimen ground contacts. It is important that the shorting block connect the specified return path conductors, but not adjacent signal pins. Figure A.3 - Example of loop inductance measurement setup EIA-364-
18、109 Page A-4 A.4 Figure A.4 shows a drawing of a microprobe contacting the bottom side of the PCB test fixture and the specimen mounted on the opposite side of the PCB. The test professional should be aware that this type of fixture may be used, but that all vias and traces should be taken into cons
19、ideration when conducting the calibration procedures. Figure A.4 - Diagram of microprobes contacting the bottom side of the printed circuit board test fixture Specimen Microprobe Fixture EIA-364-109 Page B-1 B Calibration standards and test board reference traces (informative) B.1 Calibration standa
20、rds B.1.1 For the equipment calibration, a traceable calibration impedance standard should be used for a reference baseline. Specific equipment calibration should be performed according to the manufacturers instructions. However, care should be taken as to what standards or other fixtures are used f
21、or the calibration procedure. NOTE The term “calibration” used in this document is not to be confused with the periodic factory equipment calibration. Calibration is used in the sense of characterizing the fixture so that when the “fixture plus specimen” measurement is taken, the characteristics of
22、the specimen alone can be accurately determined. B.1.2 When possible the fixture should be designed to allow the attachment of the calibration standard as close to the specimen as possible. Reflections from fixture imperfections increase measurement error. B.1.3 Printed circuit test boards should no
23、t be used as calibration standards. Because of different printed circuit board technologies, fabrication control, and material variations, it becomes difficult to insure that different board designs or fabrication techniques will have the same calibration reference for the impedance measurements. Th
24、e impedance value of “controlled impedance traces” on a printed circuit board is typically 10% or 5% of the target value. In measurements and applications, this may be an acceptable tolerance to hold, however, for calibration purposes, this should not be used as a baseline. B.1.4 The use of the trac
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ECAEIA3641092003TP109LOOPINDUCTANCEMEASUREMENTTESTPROCEDUREFORELECTRICALCONNECTORS1NH10NHTP109 连接器 电感

链接地址:http://www.mydoc123.com/p-704210.html