DOD A-A-55709-1996 PUMP INTRAVENOUS INFUSION《静脉输入泵》.pdf
《DOD A-A-55709-1996 PUMP INTRAVENOUS INFUSION《静脉输入泵》.pdf》由会员分享,可在线阅读,更多相关《DOD A-A-55709-1996 PUMP INTRAVENOUS INFUSION《静脉输入泵》.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、 A-A-55709 9999974 0372826 612 W I INCH-POUND 1 A-A- 557 O9 16 August 1996 - COMMERCIAL ITEM DESCRIPTION PUMP, INTRAVENOUS INFUSION The General Services Administration has authorized the use of this Commercial Item Description for all Federal Agencies. 1. SCOPE. This Commercial Item Description cove
2、rs a portabie, battery operated, rechargeable, volumetric intravenous infusion 2. SALIENT CHARACTERISTICS: 2.1. Shall be a portable, rechargeable, battery operated, volumetric. intravenous infusionpump, suitable for the safe administration of fluid therapy. 2.2 in operation. Shall be possible to rec
3、harge the battery while the pump is 2.3 volumetric method. Shall accurately measure the fluid administration by the 2.4 Shall be capable of preselecting a volume to be administered from 1 to 999 mls. The rate of flow should at least range from 1 ml per hour to 200 hours, in 1 ml increments. Benefici
4、al comments (recommendations, additions, deletions) and any pertinent data which may be of use in improving this document should be addressed to: Headquarters, Defense Personnel Support Center, Directorate of Medical Materiel, DPSC-MBP, 2800 South 20th Street, Philadelphia, PA 19145-0530, by using t
5、he self- addressed StandardizatLon Document Improvement Proposal, 1426) appearing at the end of this document or by letter. (DD Form AMSC N/A FSC 6515 DISTRIBUTION STATEMENT A: Approved for public release; distribution is unlimited. Provided by IHSNot for ResaleNo reproduction or networking permitte
6、d without license from IHS-,-,-A-A-55 7 09 9979974 0372827 559 . A-A-557 O9 2.5 Shall have the following alarms and alerting mechanisms: a. Audible and visible air-in-line alarm. b. Visible low battery alarm that will operate one hour prior to need for recharging battery. c. An occlusion alarm to in
7、dicate occluded tubing or clogged filter. d. An audible alarm to indicate completion of infusion. 2.6 standard intravenous solution pole. Shall be capable of attachment to the upright portion of a 2.7 solution s.ets or as secondary sets with compatible tubing. 2.8 assembly and associated tubing, the
8、 pump should be capable of being cleaned monly accepted methods. Standard administration sets may be used between the With the exception of the disposable cassette and plunger 2.9 lessen the effects of intentional removal and/or accidental removal of the set/cassette to protect against uncontrolled
9、gravity infusions. Shall be provided with features designed to prevent and/or 2.10 Varying fluid container height, the maximum distance. allowed by the set tubing above and below the pump, shall not cause the flow error to be greater than () 5 percent. 2.11 the following single phase, nominal voltag
10、e sources as specified by the end user: . Voltage characteristics shall permit operation on any of 115 VAC k 10 percent, 50 Hz f 2 Hz 115 VAC k 10 percent, 60 Hz f 2 Hz 230 VAC k 10 percent, 50 Hz f 2 Hz 230 VAC f 10 percent, 60 Hz It 2 Hz 2.11.1 Equipment which operates on multiple frequency, 50/60
11、 Hz is acceptable. Shall have a power on/off switch and there shall be a clear, visual indication of line power connection and operation. 2.12 The pump shall be provided with a rechargeable, integral battery that shall provide, as a minimum, 4 hours of normal operation at a flow rate of 125 ml/hr. 2
12、 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-557 O9 2.13 activated at least 30 minutes prior to degradation of pump performance. audible alarm signaling interruption of infusion. battery signal shall precede the battery discharged alarm. A vi
13、sual and/or audible low-battery signal shall be A depleted battery shall be accompanied by an .The low- 2.14 With the unit connected to line power and operating at 125 ml/hr, it shall go from full discharge to 100 percent charge within 24 hours. With the power switch “OFF“ and the unit connected to
14、. line power, it should go from full discharge to 100 percent charge within 14 hours. There shall be a clear, visual indication of line power connection to signify battery charge when the pump switch is off. The battery shall have continuous recharge capability. 2.15 Equipment leakage current for eq
15、uipment powered by a nominal 110 VAC source shall not exceed the more stringent of the limits established in the ANSI-AAMI ES1 Standard for Safe Current Limits for Electromechanical Apparatus, or NFPA 99, Safety Standard for Healthcare Facilities, when measured as prescribed therein. For equipment p
16、owered by a nominal 220 VAC Source, leakage current shall not exceed those limits established in IEC- 601-1, Safety of Medical Equipment. 2.16 The equipment offered, should in so far as practical, be in accordance with applicable Guidelines and Preferred Practice for the Design of Medical Equipment.
17、 Nothing in this.standard shall be considered as limiting the selection of hardware, materials, or processes to the specific items described therein. Unless otherwise stated in specific provisions, this standards is applicable to the design of equipment for both men and women. 2.17 The pump shall be
18、 easy to Set-up, operate and correct alarms. The unit shall provide digital displays exhibiting amount infused, rate to be infused, and cumulative infusion volume for both primary and secondary infusions. 2.18 Units shall have a tamper proof feature to prevent patient from changing the flow rate or
19、turning off the unit. This may be accomplished by providing a locking device, programmable lock-out mode or other means of preventing unauthorized tampering. 2.19 If the devices front panel provides push button, keypad or membrane type touch sensors for data entry, adequate separation shall be provi
20、ded for activation by a single finger. Activation of the data entry device should provide an audible/tactile feedback. Invalid key depression may be indicated by an audible tone. 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-55709 9999974 037
21、2829 321 A-A-55709 2.20 Labels or markings shall be durable, and shall clearly and concisely identify the function of all controls and indicators. Simple operating sequence instructions and important precautions shall be printed on or permanently displayed on the equipment. 2-21 Shall be designed to
22、 pass Aeromedical Certification requirements and must be in compliance with the following standards. Military standards shall be interpreted as though all units will be used aboard Military aircraft. 2.21.1 Electromagnetic Interference (EMI). The purpose of the EM1 tests is to verify Compliance with
23、 MIL-STD-461D and MIL-STD- 462D, “Electromagnetic Emission and Susceptibility Requirements for the Control of Electromagnetic Interference“, and Measurement of ENI Characteristics,“ respectively. 2.21.1.1 Radiated Emissions (RE-102) - This test measures radiated emissions generated by the unit. coul
24、d interfere with aircraft navigation and communication equipment. power source that it is configured for, such as internal rechargeable battery, 110VAC/60Hz, 110VAC/400Hzr and 28VDC. Excessive emissions The unit is tested while operating on each possible 2.21.1.2 Radiated Susceptibility (RS-103) - T
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA557091996PUMPINTRAVENOUSINFUSION 静脉 输入 PDF

链接地址:http://www.mydoc123.com/p-701808.html