DOD A-A-54403-1991 INTRAUTERINE DEVICE CONTRACEPTIVE《子宫内避孕器》.pdf
《DOD A-A-54403-1991 INTRAUTERINE DEVICE CONTRACEPTIVE《子宫内避孕器》.pdf》由会员分享,可在线阅读,更多相关《DOD A-A-54403-1991 INTRAUTERINE DEVICE CONTRACEPTIVE《子宫内避孕器》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、A-A-54 4 03 22 May 1991 COMMERCIAL ITEM DESCRIPTION INTRAUTERINE DEVICE, CONTRACEPTIVE The General Administration has authorized the use of this commercial item description. This commercial item description covers an intrauterine device suitable of use in intrauterine contraception. Salient characte
2、ristics: General. The system shall be a sterile, T-shaped, copper and polyethylene intrauterine device with an inserter and plunger. Materials. The T-shaped body shall be fabricated of Alathon 20 polyethylene. Copper wire shall be Grade 1 copper wire. Copper collars shall be Grade 1 copper 99.99% pu
3、re. The polyethylene IITIl shall contain 22% barium sulfate added to render it radiopaque. Strings shall be fabricated of polyethylene containing up to 1% titanium dioxide for visibility. Style and desisn. The system shall be a T-shaped polyethylene body with approximately 176 mg of copper wire woun
4、d around the vertical portion of the l1Tl1. Copper collars each weighing 66.5 mg shall be provided on each arm of the horizontal sections of the IlTIl. The bottom of the T shall be bulb shaped with a monofilament polythylene thread tied through the bulb. The thread shall allow protrusion through the
5、 os for conformation checks that the intrauterine device is insitu. An insertion tube and plunger shall be furnished with each device. The insertion tube shall be equipped with a moveable flange to aid in gauging the depth to which the insertion tube is inserted through the cervical canal and uterin
6、e cavity. The inserter shall be 8-1/8 distribution is unlimited. . Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,- - - A-A-54403 b5 9999974 00bL51b 4 - A-A-54 4 O 3 Plunger shall be 7.5 f 0.1 inches long from handle brace to tip with a body diameter
7、 of 0.10 inches and a distal end diameter of 0.11 inches. Plunger shall have a finger ring on the proximal end for use in manipulating the plunger. The IUD shall be 1.245 to 1.270 inches wide by 1.405 to 1.425 inches high with at least 4-1/2 inches of polyethylene string attached to the bulb. The pl
8、astic body shall be 0.064 inches in diameter. The copper bands shall be 0.198 inches wide each located 7/32 inches from the vertical portion of the Instructions. Instructions for use shall be provided with each IUD which shall also contain a minimum of one patient information brochure and one patien
9、t wallet card. Shall also contain additional pertinent information on the item for physicians use. BiocomDatibility. The IUD shall meet the requirements of the USP for Class II plastics. Sterility. The set components and inside surfaces of the containing envelope shall be sterile and have a minimum
10、quality assurance level of 0.000001 in accordance with AAMI guidelines. Shelf life. A minimum of 40 months shelf life shall remain on shipment of the IUD to the Government. WorkmanshiD. The IUD shall be free from defects which detract from its appearance or impair its serviceability. Shall be free f
11、rom pits, cracks, distortion, corrosion, burrs, sharp edges or foreign material. Unit. Package (PG). One package containing five individually packaged sterile intrauterine devices, with instruction for use, as specified, constitutes one unit. Each device shall be packaged in a peel back container fo
12、r aseptic removal. The individual package shall be a sealed peel-open container capable of maintaining sterility of contents unless the package is damaged or opened. 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54 4 O 3 QUALITY ASSURANCE PRO
13、VISIONS. - Remonsibilitv for inspec-ion. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except otherwise specified in the contract or purchase order, the contractor may use his own or
14、 any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and se
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA544031991INTRAUTERINEDEVICECONTRACEPTIVE 宫内避孕器 PDF

链接地址:http://www.mydoc123.com/p-701736.html