DOD A-A-54028-1990 CLAMP ARTERY INFANT PARTIAL OCCLUDING (CASTANEDA)《部分闭塞动脉夹》.pdf
《DOD A-A-54028-1990 CLAMP ARTERY INFANT PARTIAL OCCLUDING (CASTANEDA)《部分闭塞动脉夹》.pdf》由会员分享,可在线阅读,更多相关《DOD A-A-54028-1990 CLAMP ARTERY INFANT PARTIAL OCCLUDING (CASTANEDA)《部分闭塞动脉夹》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、A-A- 54028 7 August 1990 COMIIERCIAL ITEM DESCRIPTION CLAW, ARTERY, INFANT, PARTIAL OCCLUDINQ ( CASTANEDA) The eneral Service6 Administration has authorized the use of this commercial item description. This commercial item description COVPB a gpoon-type, partial- occluding infant-gize artery clamp o
2、! the Castonedo style suitable for use in vascular procedures. Salient characteristics: Generz! The clamp shall be 4-l/e inche8 long with jaws hav1r.g interdi g: tating transverse tee th Termtnol ogy ek,a? 1 be in kcccrdance with ASTh! Designetion F64i. Material Materia! of the clamp shall be Class
3、4 mertensltic tainles stee ir. accordance with ASTM Designation F8Q. Iiardnese ?he clamp ,ohEl! have a Rockwei: IiardneRE C Scale (KRCI o! 40 to 47 wher. tested :n accordance with ASTM Designation E18 or E02 whichever is aplicable. Convereicr, of hardne8s values shall be In accoraancc with ASTM Desi
4、gnation E140 Style, aesiCn and dimensions The clamp shall be ir, acccrdance with Defense Personnel Support Center (DFSC) Crawing No 22369 ior stple, cesign and dimeneions Finigt;. The clkmp shail be =irpor, setin or matte The motte finish shall exhibit (L uniform nondirectional surface texture equiv
5、slent to that produced by lightly blasting the prepared surface with fine abuives undel. pre8su.e Paeefvetior,. The clamp shall bt properly parrlvattd Corrotrion-reEirt digtribution Is unlimited. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54
6、028 Performance. The clamp shall conform to applicable portions of ASTM Deaignation F1026 Identification marking and labelinR. In addition to the requirements specified fn ASTk! Designation FlO26, the clamp shall be legibly marked with the contractors name or registered trademark and the word rtainl
7、ess Clamps ehal? bear the name of the country of origin Workmanship The clamp shall be free from defects which detract fporn its appearance OF impair ita serviceability. Product Quality Level Visual and dimensional examination Clamps shall be examined tc verify that aesign, construction and workmans
8、hip are in accordance with pplicable requirements Materiels The supplier shall be responsible for performing teste tc assure that materiais are in accordance with r equ i re ment s Unit Each (EA) One clamp, a8 specified, constitutes one un? t W- Quality Asurance Provisions ReeDonsibilitv for nepecti
9、on Unless otherwise specified in the contract or purchase order, the contractor is responsible for th-e perfcrmance of all inspection requirement8 as apecified herein Except as otherwise specifred in the contract or pvrchase order, the contractor may uze his own or any facilities suitbble for the pe
10、rformance of the :nspectLon requirements specified hereln, unless cisapproved b) the Government The Qovernment Peserves the right tc perform my of the inspections set forth in the specification where such inspections are deemed neccosary to nt. time to time. Cuning tht performance of tht contract en
11、d !or a period of thiae yeers after delivery of tht supplies to which such records relate. 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A- A- 94028 Inapection Inspection, as used herein, is defined ar both examination (such as visual or suditory
12、 investigation without the use of special laboratory appliances or procedures) and teiting (determination by technical meanB of physical and chemical propertieel of the item Sampling for examination Sampling for examination Eholl be conducted fn accordancc with MIL-STD-105, end as tptcif ied herein
13、Unit of product for examination purposes ahail be one clamp. Sampling shall be Inspecticn Level II and an Acceptable Quality Level (AQL) of 1.0. Examinatior, Exanination shall be conducted to determine compliance with specificat+on products SamplinR for tests Sampling for tests shall be conducted ir
14、. acccrdance with MIL-STD-105 and as specified herein The unit of product for test purposes shcll be one clern; ACL shell be l O (percent defective) with an Inspection Level of S-2 Tests Tests shall be conducted to determine compliance with specif Ication requirements. Where feasible, the same sampl
15、e ohell be used :or the dttermfnstion of two or more test characterixticr. Contractor certification. The contractor shall certify that the product offered meets the rilient characteristics of thia description arid conforma to the produccr own drawings, rpecificationi, standardr, and quality atrsurin
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA540281990CLAMPARTERYINFANTPARTIALOCCLUDINGCASTANEDA 部分 闭塞 动脉 PDF

链接地址:http://www.mydoc123.com/p-701704.html