1、b A-A-53Ll57 65 7777774 003076b Li l A-A-53 459 15 December 1988 7 COMMERCIAL ITEM DESCRIPTION SUCTION APPARATUS, SURGICAL The General Services Administration has authorized the use of this Commercial Item Description. This Commercial Item Description covers a suction apparatus suitable for use in a
2、n ambulance, hospital, or in a stationary field hospital to aspirate secretions, blood, or vomitus as necessary. Salient characteristics: The apparatus shall be capable of being powered via manual operation (component hand operated pump), hospital wall vacuum, or vehicle vacuum system as an oral, na
3、sal, and tracheal suction device. The apparatus shall be suitable for use in aspirating secretions, blood, and vomitus during emergency treatment including treatment of apnea patients. In addition to requirements specified herein, all components, attachments, materials, and functions shall be in com
4、pliance with ASTM F960 (Standard Specification for Medical and Surgical Suction and Drainage Systems) as applicable for specified use. The apparatus shall consist of, at least, a double cycle hand pump, a collection/vacuum container with a container liner installed, vacuum gauge, a vacuum adjustment
5、 control, a patient suction tube, a vehicle vacuum hose, all necessary interconnecting parts, all assembled on a base plate with a removable hold down bracket, ready for use when equipped with a noncomponent catheter. A package of 24 vacuum container liners, and operating and maintenance manual(s1 s
6、hall also be included . The hand pump shall provide suctioning when operated in either direction (push-pull), and shall have a minimum displacement . of 12.5 cubic inches (0.205 liters) per stroke. AMSC N/A FSC 6515 DISTRIBUTION STATEMENT A. Approved for public release: distribution is unlimited. TH
7、IS DOCUMENT CONTAINS ;/ PAGES. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-b3Li59 b5 W 7Y 00307b7 b m A-A-53 459 c t O The collection/vacuum container, with a liner installed, shall have a minimum capacity of 2300 cc and the contents o the 13
8、nei: shall be clearly viewable during use (i.e, installed liner shall also have minimum capacity of 2300 cc). container liner shall have graduation markings. The vacuum gauge shall have a range from O to a minimum of 30 inches (63.5 cm) Hg. The The vacuum adjustment control shall be hand operated an
9、d easily accessed. It shall be capable of controlling the vacuum rom O to 28 inches (63.5 cm) Hg. The high limit of tree air flow shall be 38.5 LPM. The patient suction tube shall be 1/4 inch (0.635 cm) ID by a minimum o 5 feet (1.52 meters) and shall be sterilizable via ethylene oxide and liquid ag
10、ent (s) . The vehicle vacuum hose shall be a minimum 16 feet (4.88 meters) long by at least 1/4 inch (0.64 cm) ID. The complete apparatus, less packaging, package of liners, and manual(s) shall not weigh more than 7 pounds (3.18 kg), Performance, When manually operated at the rate o two double strok
11、es (2 full cycles) per 2.4 seconds, the apparatus shall provide a minimum of 27.0 inches (68.58 cm) Hg negative pressure with the patient tubing occluded and a minimum of 35 LPM Eree ait. flow at the patient end of the non-occluded tubing. When cycle rate is increased, the free air flow shall be lim
12、ited to a minimum of 38.5 EPM. O When operated from a noncomponent vacuum source (vehicle vacuum or electrical vacuum pump) which has an excess of vacuum and ree air flow, the vacuum adjustment control shall provide a range from 0 to a minimum of 27.5 inches (69.85 cm) Hg. The apparatus shall be cap
13、able of providing a minimum o 30 LPM ree air flow at the patient end of the tube. Materials. A1L pazts shall be corrosion-resistant or treated For corrosion-resistance (paint is not acceptable as a corrosion-resistant treatment), 2 Provided by IHSNot for ResaleNo reproduction or networking permitted
14、 without license from IHS-,-,-A-A-5 3 45 9 Instructions. The apparatus shall be-provided with manual(s1 describing the theory of operation, operating instructions, assembly and disassembly instructions, a list of replacement parts by manufacturers catalog number, descriptive sketches and appropriate
15、, st-atement of warranty, and manufacturers name, address and phone number. Workmanshi . The suction apparatus shall be free from defects Unit, Each (EA). One suction apparatus, as specified, which d etract from its appearance or impair its serviceability. constitutes one unit. Each suction apparatu
16、s shall be provided with operating and maintenance instructions for use. Quality Assurance Provisions. the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise Specified in the contract or purchase order
17、, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified hereinr unless disapproved by the Government. Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed nece
18、ssary to assure supplies and services conform to prescribed requirements. Responsibility for inspection. Unless otherwise specified in The Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the
19、 Governments request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate. -Ins ection. Inspection, as used herein, is defined as both examination + such as visual or auditory investigat
20、ion without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item. 3 , Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-I A-A-5 3 45 9 O Tests. Tests shal
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA534591988SUCTIONAPPARATUSSURGICAL 外科 吸力 装置 PDF

链接地址:http://www.mydoc123.com/p-701672.html