DOD A-A-51975-1987 LIGHT HEAD SURGICAL《外科日光》.pdf
《DOD A-A-51975-1987 LIGHT HEAD SURGICAL《外科日光》.pdf》由会员分享,可在线阅读,更多相关《DOD A-A-51975-1987 LIGHT HEAD SURGICAL《外科日光》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、x COMMERCIAL ITEM DESCRIPTION A-A-51975 26 June 1987 LIGHT, HEAD, SURGICAL The General Services Administration has authorized the use of this commercial item description. This commercial item description covers a surgical headlight with transformer. Salient characteristics: c The light shall be a hi
2、gh intensity headlight assembly system for use in medical and dental examinations. It shall consist of the following items: a. A transformer supplying two Voltage outputs. b. A headlight assembly. c. An adjustable headband. d. An extra light bulb. e. A carrying case. The headband shall be fabricated
3、 of a pliable polyethylene or plastic with a top strap to distribute headlight weight. It shall have a hook and pile fastener for size adjustment. The headband shall have a universal clamp with ball and socket joint and a soft Leather pad. The headband shall have appropriate attachments for the ligh
4、t assembly. The headlight assembly shall be small, of lightweight rigid construction consisting of a removable lamphouse (with double contact bayonet socket and lamp) which fits into a focusing member. The socket in lamphouse shall be powered from a two conductor cord at least 3 feet (i meter) long
5、terminating with a male connector engaging the female connector at the end of the transformer output line. The male connector shall be designed to prevent insertion into a 120 volt, AC receptacle. The light shall provide a minimum of 2000 footcandles of illumination at 24 inches (61 cm), 3500 footca
6、ndles at 16 inches (41 cm). The surgical headlight shall provide brilliant, low-heat light. It shall be free of ultraviolet and infra-red rays. It shall be possible to adjust the light field by axial movement of the lamphouse with respect to the focusing member . AMSC N/A FSC 6515 DISTRIBUTION STATE
7、MENT A. Approved for public release; distribution 0 is unlimited. i THIS DOCUMENT CONTAINS 4 PAGES. - - Licensed by Information Handling ServicesL A-A-51975 The transformer shall be 20 watts and provide two voltages from a two position switch. The transEormer shall have a thermal protector and a thr
8、ee-prong hospital grade grounded plug, IC shall have a secondary output line of the two conductor type, approximately 5 Eeet (1.5 metecllong, terminating in a female plug connector. The transformer shall meet all appropriate OSHA and ANSI Standards. A carrying case that holds all items in place with
9、out damage shall be supplied and shal be suitable for carrying or storage. The extra bulb shall be compatible and for use wsth the asraembly . It shall be possible to make all headband, headlight adjustments and lamp repacemencs without the need for tools. The metal portion shall be fabricated of a
10、suitable corrosion- resistant steel alloy that has been properly hardened and passivated. The unit shall be Eurnished with a nDirections for Usen. orkmanship. The surgical headlight with transformer shall be free from deeetk which detract: from its appeatance or impair its aehviceability a _I_ Unit.
11、 Each (EA) . One sucgicai light, head, with transformer, as specified, constitutes one unit. Quality Aseurance Ppovisions. Responsibility sr inspection. Unless otherwise specified in the eoneraee 06 purchase order, the concractor is responsible for the performance o all inspection requirements as sp
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA519751987LIGHTHEADSURGICAL 外科 日光 PDF