ASTM F2423-2011 Standard Guide for Functional Kinematic and Wear Assessment of Total Disc Prostheses《总脊面假体的功能 运动学和磨损评估标准指南》.pdf
《ASTM F2423-2011 Standard Guide for Functional Kinematic and Wear Assessment of Total Disc Prostheses《总脊面假体的功能 运动学和磨损评估标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2423-2011 Standard Guide for Functional Kinematic and Wear Assessment of Total Disc Prostheses《总脊面假体的功能 运动学和磨损评估标准指南》.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2423 11Standard Guide forFunctional, Kinematic, and Wear Assessment of Total DiscProstheses1This standard is issued under the fixed designation F2423; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last r
2、evision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide provides guidance for wear and/or fatiguetesting of total disc prostheses under functional and kinematiccondition
3、s and, to this end, describes test methods for assess-ment of the wear or functional characteristics, or both, of totaldisc prostheses.1.2 Both lumbar and cervical prostheses are addressed.1.3 Load and kinematic profiles for lumbar and cervicaldevices are not identical and, therefore, are addressed
4、sepa-rately in the guide.1.4 Partial disc replacements, such as nucleus replacementsor facet joint replacements, are not intended to be addressed.1.5 Wear is assessed using a weight loss method in a testingmedium as defined in this guide.1.6 This guide does not address any potential failure modeas i
5、t relates to the fixation of the implant to its bony interfaces.1.7 It is the intent of this guide to enable comparison ofintervertebral disc (IVD) prostheses with regard to wear andfatigue characteristics when tested under the specified condi-tions. It must be recognized, however, that there are ma
6、nypossible variations in the in vivo conditions.Asingle laboratorysimulation with a fixed set of parameters might not beuniversally representative.1.8 Most IVD prostheses primarily fall into two classifica-tions: articulating ball-in-socket type prostheses, and elasto-meric or compliant type prosthe
7、ses. For the former, this guideprimarily addresses Mode 1 wear (defined herein); whereas forthe latter, this guide addresses potential failure of the prosthesiswhen the implant is subjected to a range of motion and/or loadsthat fall within the full range of possible physiologic motionsand loads.1.9
8、For articulating components, this guide predominantlydescribes a Mode 1 test. The user is cautioned that other modesof wear may occur and may have significant influence on thefunctionality and performance of an articulating IVD prosthe-sis, and therefore the user should consider the effects of other
9、wear modes on the performance of the prosthesis.1.10 In order that the data be reproducible and comparablewithin and between laboratories, it is essential that uniformprocedures are established. This guide is intended to facilitateuniform methods for testing and reporting of data for total discrepla
10、cement prostheses.1.11 Without a substantial clinical retrieval history of IVDprostheses, actual loading profiles and patterns cannot bedelineated at the time of the writing of this guide. It thereforefollows that the load and motion conditions specified by thisguide do not necessarily accurately re
11、produce those occurringin vivo. Rather, this guide provides useful boundary/endpointconditions for evaluating prosthesis designs in a functionalmanner.1.12 The values stated in SI units are to be regarded as thestandard with the exception of angular measurements, whichmay be reported in either degre
12、es or radians.1.13 This guide is not intended to be a performance stan-dard. It is the responsibility of the user of this guide tocharacterize the safety and effectiveness of the prosthesis underevaluation.1.14 This standard does not purport to address all of thesafety concerns, if any, associated w
13、ith its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2F561 Practice for Retrieval and Analysis of Medical De-vices, and
14、 Associated Tissues and FluidsF1582 Terminology Relating to Spinal ImplantsF1714 Guide for Gravimetric Wear Assessment of Pros-thetic Hip Designs in Simulator DevicesF1877 Practice for Characterization of ParticlesF2077 Test Methods For Intervertebral Body Fusion De-vices1This guide is under the jur
15、isdiction of ASTM Committee F04 on Medical andSurgical Materials and Devices and is the direct responsibility of SubcommitteeF04.25 on Spinal Devices.Current edition approved July 1, 2011. Published August 2011. Originallyapproved in 2005. Last previous edition approved in 2005 as F2423 05. DOI:10.1
16、520/F2423-11.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor D
17、rive, PO Box C700, West Conshohocken, PA 19428-2959, United States.2.2 ISO Standard:3ISO 181921 Implants for SurgeryWear of Total Interver-tebral Spinal Disc ProsthesesPart 1: Loading and Dis-placement Parameters for Wear Testing and Correspond-ing Environmental Conditions for Test3. Terminology3.1
18、DefinitionsAll functional and kinematic testing termi-nology is consistent with the referenced standards (for ex-ample, Test Methods F2077, Terminology F1582, and so forth),unless otherwise stated.3.1.1 axial load, nthe resultant force Faxialapplied to thesuperior or inferior fixture-end plate that
19、simulates the in vivoload that an IVD prosthesis (original healthy disc) must resist.3.1.1.1 DiscussionBased on a healthy disc, the primarycomponent would be an axial compressive force FZin thedirection of the negative global Z axis, and it would passthrough the origin of the IVD prosthesis. Shear c
20、omponents inthe XY plane would be FXand FY. Lateral bending moment MXand flexion/extension moment MYcomponents would be cre-ated about the origin when the axial load does not pass throughit.3.1.2 coordinate system/axes, nglobal XYZ orthogonalaxes are defined following a right-handed Cartesian coordi
21、natesystem in which the XY plane is to bisect the sagittal planeangle between superior and inferior surfaces that are intendedto simulate the adjacent vertebral end plates. The global axesare stationary relative to the IVD prostheses inferior end platefixture, which, in this guide, is also considere
22、d to be stationarywith respect to the test machines frame. Lower case letters,xyz, denote a local, moving orthogonal coordinate systemattached to the superior end plate fixture with directionsinitially coincident with those of the global XYZ axes, respec-tively. The 3-D motion of the superior relati
23、ve to inferior endplate fixture is specified and is to be measured in terms ofsequential Eulerian angular rotations about the xyz axes,respectively (z, axial rotation; x, lateral bending; and y,flexion-extension).3.1.2.1 origin, ncenter of the global coordinate system islocated at the initial positi
24、on of the total disc replacementsinstantaneous center of rotation (COR).3.1.2.2 DiscussionSome articulating devices do not havea single center of rotation, but instead have either a mobilecenter of rotation or multiple distinct centers of rotation,depending on the direction of movement. In this case
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF24232011STANDARDGUIDEFORFUNCTIONALKINEMATICANDWEARASSESSMENTOFTOTALDISCPROSTHESES 总脊面假体 功能 运动学 磨损

链接地址:http://www.mydoc123.com/p-538666.html