ASTM E1950-2010 Standard Practice for Reporting Results from Methods of Chemical Analysis《对化学分析方法得出的分析结果进行报告的标准惯例》.pdf
《ASTM E1950-2010 Standard Practice for Reporting Results from Methods of Chemical Analysis《对化学分析方法得出的分析结果进行报告的标准惯例》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1950-2010 Standard Practice for Reporting Results from Methods of Chemical Analysis《对化学分析方法得出的分析结果进行报告的标准惯例》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E1950 10Standard Practice forReporting Results from Methods of Chemical Analysis1This standard is issued under the fixed designation E1950; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A n
2、umber in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the approximate number of digitsrequired to express the expected precision of results reportedfrom standard methods
3、 of chemical analysis. This practiceprovides selection criteria and proper form and symbols forcoding results when necessary to indicate the relative reliabilityof results having small values.1.2 Specifically excluded is consideration of report formsand the associated informational content of report
4、s in whichresults are tabulated or transmitted. It is assumed that thereporting laboratory has established a report format to ensureproper identification of the materials tested, the nature andconditions of the test, the responsible personnel, and otherrelated information in accordance with existing
5、 regulations andgood laboratory practices.2. Referenced Documents2.1 ASTM Standards:2E29 Practice for Using Significant Digits in Test Data toDetermine Conformance with SpecificationsE135 Terminology Relating to Analytical Chemistry forMetals, Ores, and Related MaterialsE1601 Practice for Conducting
6、 an Interlaboratory Study toEvaluate the Performance of an Analytical MethodE1763 Guide for Interpretation and Use of Results fromInterlaboratory Testing of Chemical Analysis Methods3. Terminology3.1 Definitions:3.1.1 For definitions of terms, refer to Terminology E135.3.2 Definitions of Terms Speci
7、fic to This Standard:3.2.1 lower limit, L, nthe lower limit of the quantitativeanalyte concentration range (see Annex A1).3.2.2 low-level reproducibility index, KR, nthe reproduc-ibility index constant (for low analyte levels) determined inaccordance with Guide E1763.3.2.3 null limit, NL, nthe analy
8、te content below whichresults are so near zero that averaging is unlikely to yield avalue significantly different from zero.3.2.4 quantitative, adjrelating to results, having a numeri-cal value that includes at least one significant digit (see PracticeE29).4. Significance and Use4.1 Aresult must be
9、stated to a sufficient number of digits sothat a user receives both quantitative information and ameasure of the variability of the value reported.4.2 The range of application of most methods of chemicalanalysis is based upon the presumption that the quantitativeresults produced are to be used to co
10、mpare the analyte contentof the test material with specified limiting values. However,analytical results may be used legitimately for other purposes.If the same material is analyzed a number of times or a productis analyzed periodically during an interval of production, eachset of results may be ave
11、raged to yield an average result havingimproved reliability, provided nothing has been done betweenanalyses to modify the composition of the analyzed material.Results that fall below the lower limit, although not quantita-tive individually, contain compositional information and maybe reported. The r
12、eporting system in this practice permits theanalyst to indicate which values are likely to be renderedquantitative by averaging and which are not.4.3 The system is simple enough to be used routinely inreporting results from standard methods and assists thoseuntrained in statistics to apply results a
13、ppropriately.5. Rounding Calculated Values5.1 Use information from the precision section of themethod to determine the appropriate number of digits to reportas follows:5.1.1 Estimate the reproducibility index, R, at the analytelevel of the result, C, from an equation of R as a function ofconcentrati
14、on or from the table of statistical information.5.1.2 Calculate the percent relative reproducibility index:Rrel%5 100 3 R / C (1)1This practice is under the jurisdiction of ASTM Committee E01 on AnalyticalChemistry for Metals, Ores, and Related Materials and is the direct responsibility ofSubcommitt
15、ee E01.22 on Laboratory Quality.Current edition approved Oct. 1, 2010. Published December 2010. Originallyapproved in 1998. Last previous edition approved in 2003 as E1950 98 (2003).DOI: 10.1520/E1950-10.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Se
16、rvice at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.1.3 For results within the range of appl
17、ication specified inthe method, round the values to the number of digits specifiedin Table 1 (see A1.1.1 through A1.1.2).5.1.4 For results less than the lower limit, proceed inaccordance with Section 6 to establish the number of digits andappropriate coding for rounding and reporting the values.5.2
18、Calculated values shall be rounded to the requirednumber of digits in accordance with the rounding method ofPractice E29.5.2.1 The procedure is summarized as follows:5.2.1.1 When rounding a number to a specified number ofdigits, choose that digit that is nearest. If two choices arepossible, as when
19、the digits dropped are exactly a five or a fivefollowed only by zeros, choose that end in an even digit.6. Procedure6.1 Preliminary PrecautionFor a method to be used toanalyze materials with analyte content very near zero, theanalyst shall determine that it is capable of producing “unbi-ased” estima
20、tes of zero. If the method occasionally yieldsnegative results for low analyte levels, that capability isdemonstrated. Proceed in accordance with 6.2.6.1.1 Test for “Biased-Zero” MethodsPrepare the methodto perform determinations. Include all aspects of instrumentpreparation and calibration. Apply t
21、he method to a “blank”sample or one known to have negligible analyte content butthat meets the methods scope requirements in all otherrespects. If the method yields a negative result, it is not a“biased-zero” method; proceed in accordance with 6.2. If,during the course of at least ten replicate dete
22、rminations,several zeros but no negative values are observed, it is a“biased-zero” method. Apply the biased-zero rule of 6.4 inreporting results lower than NL (see 6.2.2).6.2 Critical Concentrations:6.2.1 From the method, obtain the value of the lower limit,L, to two digits (add a final zero, if nec
23、essary). Determine thedecimal place of the second digit.6.2.2 Calculate the null limit as follows:NL 5 L / 4 (2)6.3 Basic Rules:6.3.1 Numerical values shall be reported for every result(including negative values) obtained from a properly con-ducted method except as provided for certain results from“
24、biased-zero” methods in accordance with 6.1.1 and 6.4.6.3.2 Results Less Than LRound values to the seconddecimal place of L, and enclose in parentheses before report-ing. Examples: For L equal to 1.5, round to x.x and report (x.x);for L equal to 0.22, round to 0.xx and report (0.xx); for L equalto 0
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME19502010STANDARDPRACTICEFORREPORTINGRESULTSFROMMETHODSOFCHEMICALANALYSIS 化学分析 方法 得出 分析 结果 进行 报告

链接地址:http://www.mydoc123.com/p-529898.html