ASTM E168-2016 5921 Standard Practices for General Techniques of Infrared Quantitative Analysis《红外线定量分析通用技术的标准实施规程》.pdf
《ASTM E168-2016 5921 Standard Practices for General Techniques of Infrared Quantitative Analysis《红外线定量分析通用技术的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E168-2016 5921 Standard Practices for General Techniques of Infrared Quantitative Analysis《红外线定量分析通用技术的标准实施规程》.pdf(18页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E168 16Standard Practices forGeneral Techniques of Infrared Quantitative Analysis1This standard is issued under the fixed designation E168; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A n
2、umber in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the U.S. Department of Defense.1. Scope1.1 These practices cover the techniques most often us
3、ed ininfrared quantitative analysis. Practices associated with thecollection and analysis of data on a computer are included aswell as practices that do not use a computer.1.2 This practice does not purport to address all of theconcerns associated with developing a new quantitativemethod. It is the
4、responsibility of the developer to ensure thatthe results of the method fall in the desired range of precisionand bias.1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 This standard does not purport to address all of thes
5、afety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific hazardstatements appear in Section 6, Note A4.7, Note A4.11, a
6、ndNote A5.6.2. Referenced Documents2.1 ASTM Standards:2E131 Terminology Relating to Molecular SpectroscopyE334 Practice for General Techniques of Infrared Micro-analysisE932 Practice for Describing and Measuring Performance ofDispersive Infrared SpectrometersE1252 Practice for General Techniques for
7、 Obtaining Infra-red Spectra for Qualitative AnalysisE1421 Practice for Describing and Measuring Performanceof Fourier Transform Mid-Infrared (FT-MIR) Spectrom-eters: Level Zero and Level One TestsE1655 Practices for Infrared Multivariate QuantitativeAnalysis3. Terminology3.1 For definitions of term
8、s and symbols, refer to Terminol-ogy E131.4. Significance and Use4.1 These practices are intended for all infrared spectrosco-pists. For novices, these practices will serve as an overview ofpreparation, operation, and calculation techniques. For experi-enced persons, these practices will serve as a
9、review whenseldom-used techniques are needed.5. Apparatus5.1 The infrared techniques described here assume that theequipment is of at least the usual commercial quality and meetsthe standard specifications of the manufacturer. For dispersiveinstruments, also refer to Practice E932. For Fourier Trans
10、formand dispersive instruments, also refer to Practices E1421 andE932 respectively, and for microanalysis with these instru-ments see Practice E334.5.2 In developing a spectroscopic method, it is the respon-sibility of the originator to describe the instrumentation and theperformance required to dup
11、licate the precision and bias of amethod. It is necessary to specify this performance in termsthat can be used by others in applications of the method.6. Hazards6.1 Users of these practices must be aware that there areinherent dangers associated with the use of electricalinstrumentation, infrared ce
12、lls, solvents, and other chemicals,and that these practices cannot and will not substitute for apractical knowledge of the instrument, cells, and chemicalsused in a particular analysis.7. Considerations for Quantitative InfraredMeasurements7.1 Quantitative infrared analysis is commonly done withgrat
13、ing, filter, prism, or interferometer instruments. The fol-lowing guidelines for setting up an analytical procedure areappropriate:1These practices are under the jurisdiction of ASTM Committee E13 onMolecular Spectroscopy and Separation Science and are the direct responsibility ofSubcommittee E13.03
14、 on Infrared and Near Infrared Spectroscopy.Current edition approved April 1, 2016. Published June 2016. Originallyapproved in 1964. Last previous edition approved in 2006 as E168 06 which waswithdrawn January 2015 and reinstated in April 2016. DOI: 10.1520/E0168-16.2For referenced ASTM standards, v
15、isit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-
16、2959. United States17.1.1 Always operate the instrument in the most stable andreproducible conditions attainable. This includes instrumentwarm-up time, sample temperature equilibration, and exactreproduction of instrument performance tests for both stan-dards and samples.After calibration, use equiv
17、alent settings foranalyses. For all infrared instruments, refer to the manufactur-ers recommendations for the instrument settings. Aftercalibration, use these same settings for analysis.7.1.2 The absorbance values at analytical wavenumbersshould fall within the acceptably accurate range of the parti
18、cu-lar spectrometer used. In general, a single absorbance measure-ment will have the best signal-to-noise ratio when it is in therange from 0.3 to 0.8 absorbance units (AU) (1).3Thesensitivity of Fourier transform (FT-IR) spectrometers is suchthat lower absorbance values can be used quite effectivel
19、y,provided that the baseline can be estimated accurately (seeSection 12). Absorbances greater than 0.8 AU should beavoided wherever possible because of the possibility ofinstrumentally-caused non-linearity, both for dispersive (2) andFT-IR (3,4) spectrometers. Variation of the concentration andsampl
20、e path length can be used to adjust absorbance values intothe optimum range. When multiple components are determinedin a particular sample, it is acceptable to use absorbance valuesoutside the optimum range, (5) however, absorbances greaterthan 1.5AU should be avoided (2-4). Weaker absorption bandso
21、f high concentration components may be selected to provideabsorbance values within the optimal range.7.1.3 The most accurate analytical methods are imple-mented with samples in solution. With liquid samples that arenot exceptionally viscous, best results are obtained if the cell isnot moved after th
22、e first sample is introduced into the instru-ment (the fixed-cell method). The reason is that sample cellposition is difficult to reproduce accurately by insertion intotypical cell holders. Suitable fittings and tubes can be attachedto the cell to allow sample changing in a flow-through manner.When
23、it is not practical to use a flow-through cell, the cellshould fit tightly in the holder so that lateral and tilting motionsare restricted.7.1.4 Unless there is reason to suspect deposition on orcontamination of the cell from the samples, it is generallypreferable to wash out the current sample with
24、 the next sample,if sufficient sample is available. The volume of sample used toflush the cell should be at least five times (and preferably more,for example, 20 times) the volume between the sample inletand cell exit points.7.1.5 For some bands, the wavenumber of the maximumabsorbance changes as a
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME16820165921STANDARDPRACTICESFORGENERALTECHNIQUESOFINFRAREDQUANTITATIVEANALYSIS 红外线 定量分析 通用 技术 标准

链接地址:http://www.mydoc123.com/p-527014.html