ASTM D5799-1995(2004) Standard Test Method for Determination of Peroxides in Butadiene《丁二烯中过氧化物测定的标准试验方法》.pdf
《ASTM D5799-1995(2004) Standard Test Method for Determination of Peroxides in Butadiene《丁二烯中过氧化物测定的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D5799-1995(2004) Standard Test Method for Determination of Peroxides in Butadiene《丁二烯中过氧化物测定的标准试验方法》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 5799 95 (Reapproved 2004)An American National StandardStandard Test Method forDetermination of Peroxides in Butadiene1This standard is issued under the fixed designation D 5799; the number immediately following the designation indicates the year oforiginal adoption or, in the case of
2、revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the determination of peroxidesin butadiene.1.2 This test method covers
3、the concentrations range of 1 to10 ppm by mass (ppmw) as available oxygen.1.3 The values stated in SI units are to be regarded as thestandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard
4、 to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D 1265 Practice for Sampling Liquefied Petroleum (LP)GasesManual MethodD 3700 Practice for Containing Hydrocarbons FluidSamples Us
5、ing a Floating Piston Cylinder3. Summary of Test Method3.1 A known mass of the butadiene sample is placed in aflask and evaporated. The residue is then refluxed with aceticacid and sodium iodide reagents. The peroxides react toliberate iodine which is titrated with standard sodium thiosul-fate solut
6、ion using visual end-point detection. Interfering tracesof iron are complexed with sodium fluoride.4. Significance and Use4.1 Due to the inherent danger of peroxides in butadiene,specification limits are usually set for their presence. This testmethod will provide values that can be used to determin
7、e theperoxide content of a sample of commercial butadiene.4.2 Butadiene polyperoxide is a very dangerous product ofthe reaction between butadiene and oxygen that can occur. Theperoxide has been reported to be the cause of some violentexplosions in vessels that are used to store butadiene.5. Apparatu
8、s5.1 Condensers, Liebig, with 24/40 standard-taperedground-glass joint connections.5.2 Cylinders, graduated, 100-mL capacity.5.3 Flask, Erlenmeyer, 250-mL capacity, with 24/40standard-tapered ground-glass connections with marking at100 mL.5.4 Heating Mantle, electric, for 250-mL Erlenmeyer flasks.5.
9、5 Microburette, 10-mL capacity, graduated in 0.02-mLdivisions.5.6 Water Bath, a thermostatically controlled liquid bathcapable of maintaining a water temperature of 60 6 1C (1406 2F).6. Reagents6.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. Unless otherwise indicated, it is
10、 intended thatall reagents shall conform to the specifications of the Commit-tee on Analytical Reagents of the American Chemical Society,where such specifications are available.3Other grades may beused, provided it is first ascertained that the reagent is ofsufficiently high purity to permit its use
11、 without lessening theaccuracy of the determination.6.2 Purity of WaterUnless otherwise indicated, referencesto water shall be understood to mean deionized or distilledwater.6.3 Acetic Acid, 94 % by volume. Mix 60 mL of water with940 mL of glacial acetic acid (CH3COOH). (WarningPoisonous and corrosi
12、ve. Combustible. May be fatal if swal-lowed. Causes severe burns. Harmful if inhaled.)1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the direct responsibility of SubcommitteeD02.D0 on Hydrocarbons for Chemical and Special Uses.Current edi
13、tion approved Nov. 1, 2004. Published November 2004. Originallyapproved in 1995. Last previous edition approved in 2000 as D 5799 95 (2000)e1.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards vol
14、ume information, refer to the standards Document Summary page onthe ASTM website.3Reagent Chemicals, American Chemical Society Specifications, AmericanChemical Society, Washington, DC. For suggestions on the testing of reagents notlisted by the American Chemical Society, see Analar Standards for Lab
15、oratoryChemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeiaand National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,MD.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6.4 Carbon Dioxide, s
16、olid (dry ice). (WarningUse glovesto avoid frostbite when handling.)6.5 Potassium Dichromate Solution, Standard (0.1 N)Dissolve 2.452 g of potassium dichromate (K2Cr2O7) in waterand dilute to 500 mL in a volumetric flask.(WarningAvoidcontact with eyes and skin and avoid breathing of dust.)6.6 Sodium
17、 Fluoride.6.7 Sodium Iodide.6.8 Sodium Thiosulfate Solution, Standard (0.1 N)Dissolve 12.5 g of sodium thiosulfate (Na2S2O33 5H2O) plus0.1 g of sodium carbonate (Na2CO3) in 500 mL of water (theNa2CO3is added to stabilize the Na2S2O3solution). Let thissolution stand a week or more before using. Stand
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便