ASTM D93-2016 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester《用宾斯克马丁闭杯闪点试验器测定闪点的标准试验方法》.pdf
《ASTM D93-2016 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester《用宾斯克马丁闭杯闪点试验器测定闪点的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D93-2016 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester《用宾斯克马丁闭杯闪点试验器测定闪点的标准试验方法》.pdf(18页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D93 16Designation: 34/99Standard Test Methods forFlash Point by Pensky-Martens Closed Cup Tester1This standard is issued under the fixed designation D93; the number immediately following the designation indicates the year of originaladoption or, in the case of revision, the year of last
2、 revision. A number in parentheses indicates the year of last reapproval. A superscriptepsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the U.S. Department of Defense.INTRODUCTIONThis flash point test method is a dy
3、namic test method which depends on specified rates of heatingto be able to meet the precision of the test method. The rate of heating may not in all cases give theprecision quoted in the test method because of the low thermal conductivity of some materials. Thereare flash point test methods with slo
4、wer heating rates available, such as Test Method D3941 (forpaints, resins, and related products, and high viscosity products in the range of 0 C to 110 C), wherethe test conditions are closer to equilibrium.Flash point values are a function of the apparatus design, the condition of the apparatus use
5、d, andthe operational procedure carried out. Flash point can therefore only be defined in terms of a standardtest method, and no general valid correlation can be guaranteed between results obtained by differenttest methods, or with test apparatus different from that specified.1. Scope*1.1 These test
6、 methods cover the determination of the flashpoint of petroleum products in the temperature range from40 C to 370 C by a manual Pensky-Martens closed-cupapparatus or an automated Pensky-Martens closed-cupapparatus, and the determination of the flash point of biodieselin the temperature range of 60 C
7、 to 190 C by an automatedPensky-Martens closed cup apparatus.NOTE 1Flash point determinations above 250 C can be performed,however, the precision has not been determined above this temperature.For residual fuels, precision has not been determined for flash pointsabove 100 C. The precision of in-use
8、lubricating oils has not beendetermined. Some specifications state a D93 minimum flash point below40 C, however, the precision has not been determined below thistemperature.1.2 Procedure A is applicable to distillate fuels (diesel,biodiesel blends, kerosine, heating oil, turbine fuels), new andin-us
9、e lubricating oils, and other homogeneous petroleumliquids not included in the scope of Procedure B or ProcedureC.1.3 Procedure B is applicable to residual fuel oils, cutbackresidua, used lubricating oils, mixtures of petroleum liquidswith solids, petroleum liquids that tend to form a surface filmun
10、der test conditions, or are petroleum liquids of such kine-matic viscosity that they are not uniformly heated under thestirring and heating conditions of Procedure A.1.4 Procedure C is applicable to biodiesel (B100). Since aflash point of residual alcohol in biodiesel is difficult to observeby manua
11、l flash point techniques, automated apparatus withelectronic flash point detection have been found suitable.1.5 These test methods are applicable for the detection ofcontamination of relatively nonvolatile or nonflammable ma-terials with volatile or flammable materials.1.6 The values stated in SI un
12、its are to be regarded as thestandard.1.6.1 ExceptionThe values given in parentheses are forinformation only.NOTE 2It has been common practice in flash point standards for manydecades to alternately use a C-scale or an F-scale thermometer fortemperature measurement. Although the scales are close in
13、increments,they are not equivalent. Because the F-scale thermometer used in thisprocedure is graduated in 5 F increments, it is not possible to read it tothe 2 C equivalent increment of 3.6 F. Therefore, for the purposes ofapplication of the procedure of the test method for the separate tempera-ture
14、 scale thermometers, different increments must be used. In this testmethod, the following protocol has been adopted: When a temperature isintended to be a converted equivalent, it will appear in parenthesesfollowing the SI unit, for example 370 C (698 F). When a temperature1These test methods are un
15、der the joint jurisdiction ofASTM Committee D02 onPetroleum Products, Liquid Fuels, and Lubricants and are the direct responsibilityof Subcommittee D02.08 on Volatility. In the IP, these test methods are under thejurisdiction of the Standardization Committee.Current edition approved June 1, 2016. Pu
16、blished July 2016. Originally approvedin 1921. Last previous edition approved in 2015 as D93 15a. DOI: 10.1520/D0093-16.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1i
17、s intended to be a rationalized unit for the alternate scale, it will appearafter “or,” for example, 2 C or 5 F.1.7 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate saf
18、ety and health practices and determine the applica-bility of regulatory limitations prior to use. For specificwarning statements, see 6.4, 7.1, 9.3, 9.4, 11.1.2, 11.1.4, 11.1.8,11.2.2, and 12.1.2.2. Referenced Documents2.1 ASTM Standards:2D56 Test Method for Flash Point by Tag Closed Cup TesterD3941
19、 Test Method for Flash Point by the EquilibriumMethod With a Closed-Cup ApparatusD4057 Practice for Manual Sampling of Petroleum andPetroleum ProductsD4177 Practice for Automatic Sampling of Petroleum andPetroleum ProductsE1 Specification for ASTM Liquid-in-Glass ThermometersE300 Practice for Sampli
20、ng Industrial ChemicalsE502 Test Method for Selection and Use of ASTM Stan-dards for the Determination of Flash Point of Chemicalsby Closed Cup Methods2.2 ISO Standards3Guide 34 Quality Systems Guidelines for the Production ofReference MaterialsGuide 35 Certification of Reference MaterialGeneral and
21、Statistical Principles3. Terminology3.1 Definitions:3.1.1 biodiesel, na fuel comprised of mono-alkyl esters oflong chain fatty acids derived from vegetable oils or animalfats, designated B100.3.1.2 biodiesel blends, na blend of biodiesel fuel withpetroleum-based diesel fuel.3.1.3 dynamic, adjin petr
22、oleum productsin petroleumproduct flash point test methodsthe condition where thevapor above the test specimen and the test specimen are not intemperature equilibrium at the time that the ignition source isapplied.3.1.3.1 DiscussionThis is primarily caused by the heatingof the test specimen at the c
23、onstant prescribed rate with thevapor temperature lagging behind the test specimen tempera-ture.3.1.4 equilibrium, nin petroleum productsin petroleumproduct flash point test methodsthe condition where thevapor above the test specimen and the test specimen are at thesame temperature at the time the i
24、gnition source is applied.3.1.4.1 DiscussionThis condition may not be fullyachieved in practice, since the temperature may not be uniformthroughout the test specimen, and the test cover and shutter onthe apparatus can be cooler.3.1.5 flash point, nin flash point test methods, the lowesttemperature o
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD932016STANDARDTESTMETHODSFORFLASHPOINTBYPENSKYMARTENSCLOSEDCUPTESTER 用宾斯克 马丁 闪点 试验 测定 标准 方法 PDF

链接地址:http://www.mydoc123.com/p-512536.html