ASTM D1946-1990(2011) Standard Practice for Analysis of Reformed Gas by Gas Chromatography《气相色谱法分析重整气的标准操作规程》.pdf
《ASTM D1946-1990(2011) Standard Practice for Analysis of Reformed Gas by Gas Chromatography《气相色谱法分析重整气的标准操作规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM D1946-1990(2011) Standard Practice for Analysis of Reformed Gas by Gas Chromatography《气相色谱法分析重整气的标准操作规程》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D1946 90 (Reapproved 2011)Standard Practice forAnalysis of Reformed Gas by Gas Chromatography1This standard is issued under the fixed designation D1946; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers the determination of the chemicalcomposition of reformed gases and similar gaseous mixturescontaining
3、the following components: hydrogen, oxygen, ni-trogen, carbon monoxide, carbon dioxide, methane, ethane,and ethylene.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This standard does not purport to address all of thesaf
4、ety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E260 Practice for Packed Column
5、 Gas Chromatography3. Summary of Practice3.1 Components in a sample of reformed gas are physicallyseparated by gas chromatography and compared to correspond-ing components of a reference standard separated underidentical operating conditions, using a reference standardmixture of known composition. T
6、he composition of the re-formed gas is calculated by comparison of either the peakheight or area response of each component with the corre-sponding value of that component in the reference standard.4. Significance and Use4.1 The information about the chemical composition can beused to calculate phys
7、ical properties of the gas, such as heating(calorific) value and relative density. Combustion characteris-tics, products of combustion, toxicity, and interchangeabilitywith other fuel gases may also be inferred from the chemicalcomposition.5. Apparatus5.1 DetectorThe detector shall be a thermal cond
8、uctivitytype or its equivalent in stability and sensitivity. The thermalconductivity detector must be sufficiently sensitive to producea signal of at least 0.5 mV for 1 mol % methane in a 0.5-mLsample.5.2 Recording InstrumentsEither strip chart recorders orelectronic integrators, or both, are used t
9、o display the separatedcomponents. Although a strip chart recorder is not requiredwhen using electronic integration, it is highly desirable forevaluation of instrument performance.5.2.1 The recorder, when used, shall be a strip chart recorderwith a full-range scale of 5 mV or less (1 mV preferred).
10、Thewidth of the chart shall be not less than 150 mm. A maximumpen response time of2s(1spreferred) and a minimum chartspeed of 10 mm/min shall be required. Faster speeds up to 100mm/min are desirable if the chromatogram is to be interpretedusing manual methods to obtain areas.5.2.2 Electronic or Comp
11、uting IntegratorsProof of sepa-ration and response equivalent to that for the recorder isrequired for displays other than by chart recorder.5.3 AttenuatorIf manual methods are used to interpret thechromatogram, an attenuator must be used with the detectoroutput signal to keep the peak maxima within
12、the range of therecorder chart. The attenuator must be accurate to within 0.5 %between the attenuator range steps.5.4 Sample Inlet System:5.4.1 The sample inlet system must be constructed ofmaterials that are inert and nonadsorptive with respect to thecomponents in the sample. The preferred material
13、 of construc-tion is stainless steel. Copper and copper-bearing alloys areunacceptable.5.4.2 Provision must be made to introduce into the carriergas ahead of the analyzing column a gas-phase sample that hasbeen entrapped in either a fixed volume loop or tubular section.The injected volume must be re
14、producible such that successiveruns of the same sample agree within the limits of repeatabilityfor the concentration range as specified in 11.1.1.1This practice is under the jurisdiction of ASTM Committee D03 on GaseousFuels and is the direct responsibility of Subcommittee D03.07 on Analysis ofChemi
15、cal Composition of Gaseous Fuels.Current edition approved Nov. 1, 2011. Published December 2011. Originallyapproved in 1962. Last previous edition approved in 2006 as D1946 90 (2006).DOI: 10.1520/D1946-90R11.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Custome
16、r Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.4.3 If the instrument is calibrated
17、with pure components,the inlet system shall be equipped to introduce a sample at lessthan atmospheric pressure. The pressure-sensing device mustbe accurate to 0.1 kPa (1 mm Hg).5.5 Column Temperature Control:5.5.1 IsothermalWhen isothermal operation is used, theanalytical columns shall be maintained
18、 at a temperature con-stant to 0.3C during the course of the sample run and thecorresponding reference run.5.5.2 Temperature ProgrammingTemperature program-ming may be used, as feasible. The oven temperature shall notexceed the recommended temperature limit for the materials inthe column.5.6 Detecto
19、r Temperature ControlThe detector tempera-ture shall be maintained at a temperature constant to 0.3Cduring the course of the sample run and the correspondingreference run. The detector temperature shall be equal to, orgreater than, the maximum column temperature.5.7 Carrier GasThe instrument shall b
20、e equipped withsuitable facilities to provide flow of carrier gas through theanalyzer and detector at a flow rate that is constant to 1 %throughout the analysis of the sample and the referencestandard. The purity of the carrier gas may be improved byflowing the carrier gas through selective filters
21、before its entryinto the chromatograph.5.8 Columns:5.8.1 The columns shall be constructed of materials that areinert and nonadsorptive with respect to the components in thesample. The preferred material of construction is stainlesssteel. Copper and copper-bearing alloys are unacceptable.5.8.2 Either
22、 an adsorption-type column or a partition-typecolumn, or both, may be used to make the analysis.NOTE 1See Practice E260 for general gas chromatography proce-dures.5.8.2.1 Adsorption ColumnThis column must completelyseparate hydrogen, oxygen, nitrogen, methane, and carbonmonoxide. If a recorder is us
23、ed, the recorder pen must return tothe baseline between each successive peak. Equivalent proof ofseparation is required for displays other than by chart recorder.Fig. 1 is an example chromatogram obtained with an adsorp-tion column.(1) Because of similarities in thermal conductivities, he-lium shoul
24、d not be used as the carrier gas for hydrogen whenhydrogen is less than 1 % of the sample. Either argon ornitrogen carrier gas is suitable for both percent and parts permillion quantities of hydrogen.(2) The use of a carrier gas mixture of 8.5 % hydrogen and91.5 % helium will avoid the problem of re
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD194619902011STANDARDPRACTICEFORANALYSISOFREFORMEDGASBYGASCHROMATOGRAPHY 色谱 分析 重整 标准 操作规程 PDF

链接地址:http://www.mydoc123.com/p-510725.html