1、Designation: C408 88 (Reapproved 2016)Standard Test Method forThermal Conductivity of Whiteware Ceramics1This standard is issued under the fixed designation C408; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last rev
2、ision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the U.S. Department of Defense.1. Scope1.1 This test method covers a general proced
3、ure2for deter-mining the thermal conductivity of whiteware ceramics overthe temperature range from 100 to 300F (40 to 150C).1.2 The values stated in inch-pound units are to be regardedas the standard. The SI (metric) units given in parentheses arefor information purposes only.1.3 This standard does
4、not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Significance and Use2.1 This test m
5、ethod provides information useful in under-standing and quantifying such parameters as thermal shockresistance and ability to conduct or dissipate heat.3. Apparatus3.1 Essentially, the apparatus3shall consist of a heating unit,upper and lower copper “thermodes” (bars that serve tointroduce heat) to
6、which the ceramic test specimen is soldered,and a cooling jacket. The foregoing set of parts shall beenclosed by a 5-in. (127-mm) diameter brass cylinder, 12 in.(305 mm) long, which is sealed to the base with a rubbergasket. With the exception of the copper thermodes and thecopper cooling jackets, t
7、he instrument shall be made entirely ofbrass with all joints silver soldered. During operation, thesystem shall be evacuated to a pressure of approximately 1 mof mercury with an oil-diffusion pump which is backed up witha mechanical vacuum pump.3.1.1 Heating UnitThe heating unit, shown in detail inF
8、ig. 1, shall consist of a 500-W Nichrome element helicallysupported by ceramic insulators inside a 2-in. (50.8-mm)diameter brass cylinder. This brass cylinder shall be highlypolished to serve also as a reflector, since heat transmission tothe upper thermode is primarily by radiation. Line voltage to
9、the heater circuit shall be held constant to within 60.1 % bymeans of an electronic voltage regulator, and power input tothe heater shall be controlled by a variable transformer.3.1.2 ThermodesThe thermodes to which the ceramic testspecimen is soldered shall be machined from electrolyticallydeposite
10、d pure copper to tolerances of 60.001 in. (0.03 mm)on diameters. Thermodes of diameters from 0.250 to 0.500 in.(6.50 to 12.70 mm) may be used, but in all cases the diametersof the specimen, and that portion of the upper and lowerthermodes incorporating the thermocouple shall be equal indiameter to w
11、ithin 60.001 in. (0.03 mm). The section of theupper thermode above the upper cooling jacket shall be coatedwith carbon black to increase its emissivity.3.1.3 Cooling JacketsDetails of the lower cooling jacketare shown in Fig. 1. The lower thermode makes contact withthe lower cooling jacket by means
12、of a tapered fit. The upper18-in. (3.2-mm) brass plate which supports the heating unitshall also be water-cooled for the purpose of eliminatingradiation from the heating unit to the thermodes and testspecimen below. In addition, the upper thermode shall be fittedwith a copper reflecting shield to pr
13、ohibit any direct radiationfrom striking the thermodes and specimen in the chamberbelow. A water-circulating system shall be provided for sup-plying the cooling jackets with water at constant pressure andtemperature. The water temperatures shall not vary at a rategreater than 1F (0.5C)/h.3.1.4 Therm
14、ocouplesCalibrated glass-asbestos insulatediron-constantan thermocouples shall be permanently attachedto the thermodes in positions 1.75 in. (44.4 mm) apart as notedin Fig. 2. The diameter of the thermocouple wires used shallnot exceed the diameter of No. 28 AWG (0.321 mm). Thethermocouple wires sha
15、ll be permanently attached to thethermode by silver-soldering each wire of the thermocoupleinto No. 70 (0.028-in. diameter) (0.71-mm) holes drilled in thethermode to a depth of132 in. (0.8 mm) on opposite ends of adiameter. All thermocouple wires shall be taken through thebase plate without any brea
16、k in the continuity of the wire. An1This test method is under the jurisdiction ofASTM Committee C21 on CeramicWhitewares and Related Products and is the direct responsibility of SubcommitteeC21.03 on Methods for Whitewares and Environmental Concerns.Current edition approved July 1, 2016. Published J
17、uly 2016. Originally approvedin 1957. Last previous edition approved in 2011 as C408 88 (2011). DOI:10.1520/C0408-88R16.2This work was done under the sponsorship of the U.S. Signal Corps., SquierSignal Laboratory, Ft. Monmouth, NJ, and the U.S. Air Force, Wright-Patterson AirDevelopment Center, OH.3
18、A suitable type of apparatus is described by Edwin Ruh in “Improved Methodof Measuring Thermal Conductivity of Dense Ceramics,” Journal, Am. CeramicSociety, Vol 37, 1954, No. 5, pp. 224229.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United Stat
19、es1ice-water bath shall be used for cold junction reference. Theinstrument for measuring emf shall not have an instrumenterror greater than 60.001 mV.4. Test Specimens4.1 The ceramic test specimens used in the thermal conduc-tivity instrument shall be accurately ground cylinders. Thespecimens may be
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC40819882016STANDARDTESTMETHODFORTHERMALCONDUCTIVITYOFWHITEWARECERAMICS 卫生 陶瓷制品 热导率 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-466502.html