ARMY MIL-P-11570 C-1984 PYROMETERS INDICATING THERMOCOUPLES《热电偶指示高温计》.pdf
《ARMY MIL-P-11570 C-1984 PYROMETERS INDICATING THERMOCOUPLES《热电偶指示高温计》.pdf》由会员分享,可在线阅读,更多相关《ARMY MIL-P-11570 C-1984 PYROMETERS INDICATING THERMOCOUPLES《热电偶指示高温计》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、 MIL-P-11570C 66 W 9999706 0127979 2.- Beneficial comments (recommendations, additions, deletions) and any pertinent data which may be of use in improving this document should bo addresaed to: Belvoir Research and Development Center, ATTN: 22060-5606 by using the self-addressed Standardization Docum
2、ent improvement Proposal (DD Form 1426) appearing at the end of this document or by letter. USA STRBE-DS, Fort Belvoir, VA 30 July 1984 SUPERSEDING MIL-P-,11570B 16 October 1974 MILITARY SPECIFICATION PYROMETERS, INDICATING, THERMOCOUPLES This specification is approved for use by all Departments and
3、 Agencies of the Department of Defense. 1. SCOPE 1.1 Scope. This specification covers direct reading, indicating thermocouple pyrometers with a temperature range up to 2700 F. 1.2 Classification. as specified (see 6.2): Pyrometers shall be of the following types and classes, Type I - Portable. Type
4、II - Wall-mounted. Class I Class 2 Class 3 - Range up to 1600 F. - Range up to 2300 F. - Range up to 2700 F. 2. APPLICABLE DOCUMENTS 2.1 Government documents. 2.1.1 Specifications and standards. Unless otherwise specified (see 6.21, . the following specifications and standards of the issue listed in
5、 that issue of the Department of Defense Index of Specifications and Standards (DoDISS) specified in the solicitation, form a part of this specification to the extent specified herein. SPECIFICATIONS FEDERAL PPP-B-60 1 PPP-B-636 - Boxes, Wood, Cleated-Plywood. - Boxes, Shipping, Fiberboard. THIS DOC
6、UMENT CONTAINS - %- PAGES Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-MIL-P-31570C bb U 77797Ob 0327780 7 U MILITARY MIL-P-116 MIL-M-16034 STANDARDS MIL-STD-105 MIL- STD-129 MIL-STD- 130 HANDBOOKS MIL I TAY MIL-HD BK-7 5 9 MIL-P-11570C - Preserva
7、tion, Methods of. - Meters, Electrical-Indicating (Switchboard and Portable Types) . - Sampling Procedures and Tables for - Marking for Shipment and Storage. - Identification Marking of US Military Inspection by Attributes. Property. - Military Standardization Handbook, HFE Design for Army Materiel.
8、 (Copies of specifications, standards, and drawings required by contractors in connection with specific acquisition functions should be obtained from the contracting activity or as directed by the cqntracting officer.) 2.2 Other publications. The following document(s) form a part of this specificati
9、on to the extent specified herein. The issues of the documents which are indicated as DoD adopted shall be the issue listed in the current DoDISS and the supplement thereto, if applicable. AMERICAN SOCIETY FOR TESTING AND MATERIALS (ASIM) D 3951 - Standard Practices for Commercial Packaging. E 220 -
10、 Calibration of Thermocouples by Comparison Techniques. E 230 - Temperature Electromotive Force (EMF) Tables for Thermocouples. (Application for copies should be addressed to the American Society for Testing and Materials, 1916 Race Street, Philadelphia, PA 19103.) AMERICAN NATIONAL STANDARDS INSTIT
11、UTE (ANSI) MC96.1 - Temperature Measurement Thermocouples. (Application for copies should be addressed to the Instrument Society of America, 67 Alexandria Drive, P.O. Box 12277, Research Triangle Park, NC 27709.) (Industry association specifications and standards are generally available for referenc
12、e from libraries. using Federal agencies.) They are also distributed among technical groups and 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-MIL-P-LL570C 66 7777906 0327983 O MIL-P-11570C 2.3 Order of precedence. In the event of a conflict betwe
13、en specification and the references cited herein, the text of this shall take precedence. 3. REQUIREMENTS 3.1 Description. The pyrometer shall consist essentially of and a thermocouple. the text of this spec if icat ion a millivoltmeter pyrometers for 3.2 First article. The contractor shall furnish
14、one or more examination and testing within the time frame specified (see 6.2) to prove prior to starting production that his production methods and choice of design detail will produce pyrometers that comply with this specification. test shall be as specified in section 4; and, unless otherwise spec
15、ified herein, all examinations and tests shall be conducted by the contractor subject to surveillance and approval by the Government (see 6.3). 6.21, the Government will conduct any or all of the first article examination and tests, as specified (see 6.2). Examination and When specified (see 3.3 Mat
16、erial. Material shall be as specified herein. Material not specified shall be selected by the contractor and shall be subject to all provisions of this specification (see 6.6). 3.4 Meter. The meter shall be a millivoltmeter and the movement of the meter shall be shielded against demagnetization. 6.2
17、1, an automatic reference junction compensator shall be provided with means for converting to manual operation. The meter shall conform to MIL-M-16034, type MR51WO75DCMV with accuracies as specified in 3.8 and 3.9, and shall be contained in a lightweight, shockproof housing. to MIL-HDBK-759, 1.2.3.1
18、 through 1.2.3.1.5.2. The housing shall protect the meter and the reference junction compensator from atmospheric conditkons normally encountered in foundries. marked as to the thermocouple for which the meter is calibrated, Unless otherwise specified (see The meter scale shall conform The meter sha
19、ll be permanently and legibly 3.5 Thermocouple. The thermocouple shall consist of 2 dissimilar metal wires. Means for identifying the wires shall be provided in accordance with ANSI MC96.1. The thermocouple shall be either protected or metal sheathed as specified (see 6.2). grounded end, as specifie
20、d (see 6.2). Protected thermocouples shall have the type and number of protection tubes provided as specified (see 6.2). A shielded or unshielded connecting block shall be provided for connecting the thermocouple to the millivoltmeter or to connecting wires or cables. be adjustable for use in any po
21、sition from the horizontal through any angle to not less than 90 degrees downward. The length of the thermocouple shall be as specified (see 6.2). The metal sheathed thermocouple shall have an exposed or The thermocouple shall 3.5.1 Connecting wires. Connecting wires shall be composed of materials t
22、hat are compatible to the thermocouple material as defined in MC96.1 with limits of error as specified in ASTM E 230. 3.6 Identification marking. Each pyrometer shall be identified in accordance with MIL-STD-130. 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license
23、from IHS-,-,-I MIL-P-33570C bb W 999790b 0327982 2 W I MIL-P-11570C 3.7 Finishing. Each pyrometer and its case shall be finished in accordance with the contractors standard practice. 3.8 Type I - portable. Type I pyrometers shall be an integral unit and shall be provided with a comfortable handle. p
24、lus or minus 1 percent of reading based on an impressed voltage input to an indicated reading. Type I meters shall have an accuracy of 3.9 Type II - wall-mounted. a thermocouple and a meter connected by an electric cable. shall be provided with a means for mounting on a wall. specified (see 6.21, th
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ARMYMILP11570C1984PYROMETERSINDICATINGTHERMOCOUPLES 热电偶 指示 高温 PDF

链接地址:http://www.mydoc123.com/p-446071.html