AIR FORCE A-A-59839-2009 DECADE RESISTANCE STANDARD《十进制电阻标准》.pdf
《AIR FORCE A-A-59839-2009 DECADE RESISTANCE STANDARD《十进制电阻标准》.pdf》由会员分享,可在线阅读,更多相关《AIR FORCE A-A-59839-2009 DECADE RESISTANCE STANDARD《十进制电阻标准》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、 FSC 6625 Beneficial comments, recommendations, additions, deletions, clarifications, etc. and any other data that may improve this document should be sent to WR-ALC/742 CBSG/GBEC, 460 Richard Ray Blvd. STE 200, Robins AFB, GA 31098-1813. Since contact information can change, you may want to verify
2、the currency of this address information using the ASSIST Online database at http:/assist.daps.dla.mil . A-A-59839 23 October 2009 COMMERICIAL ITEM DESCRIPTION DECADE RESISTANCE STANDARD The General Services Administration has authorized the use of the Commercial Item Description (CID) for all feder
3、al agencies. 1. SCOPE. This Commercial Item Description (CID) describes a wide range, four-terminal, Decade Resistance Standard with trimming capability. This device provides highly accurate resistance values from 10 m to 1.2 M with a minimum resolution of 20 . This CID lists the minimum performance
4、, design, functional environmental, reliability requirements and product conformance provisions for this Decade Resistance Standard. 2. SALIENT CHARACTERISTICS. The equipment shall be capable of operation and bid sample testing, if so requested, within the accuracies, limits, and specifications here
5、in. 2.1 Classification. Equipment covered by this CID may be commercially available equipment modified to the extent necessary to meet the following description. The equipment shall be Class 4 in accordance with MIL-PRF-28800, except as specified herein. 2.2 Safety and Environmental. The equipment s
6、hall meet all safety and environmental requirements as specified in MIL-PRF-28800 for the classification as stated herein. 2.2.1 Temperature Operating. The equipment shall conform to its specified performance and accuracy while being operated as a calibration standard at a temperature of 23 C 1 C. F
7、urthermore this equipment shall conform to the specified temperature adjusted performance and accuracy while being operated in a minimum temperature range of 18 C to 28 C. 2.2.2 Temperature Non-Operating. The equipment shall conform to its specified performance and accuracy after being stored in an
8、environment in a temperature range of -10 C to 60 C. 2.2.3 Humidity Range Operating. The equipment shall conform to its specified performance and accuracy while being operated in a non-condensing humidity range of 30% to 70%. 2.3 Reliability. The design of the equipment shall be such that under norm
9、al use and operation the equipment does not fail within 2,000 hours of operation with a statistical certainty of 95%. 2.4 Calibration and Maintenance Adjustments. The design of the system shall provide for readily accessible calibration and maintenance adjustments. These adjustments shall be provide
10、d by variable value components, which are adjustable, by the use of simple means. The calibration by substitution of selected components or parts is unacceptable unless specifically approved. The calibration adjustments, wherever possible, shall be accessible without removal of the instrument case o
11、r modules. The calibration interval shall be a period of one year or greater, based on an operating time of 2000 hours per year. INCH-POUND INCH-POUND Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-59839 2 2.5 Performance Requirements. The instr
12、ument shall meet all requirements specified herein, after the warm-up period specified in MIL-PRF-28800, under all combinations of input power conditions, output signal conditions, and operating service condition as specified in MIL-PRF-28800. 2.5.1 Range. This equipment shall have a minimum range o
13、f 10 m to 1.2 M. 2.5.2 Resolution. This equipment shall have a resolution of 20 or better. 2.5.3 Number of Decades: This equipment shall have a minimum of eight decade switch. 2.5.4 Accuracy. 2.5.4.1 Initial Accuracy. This equipment shall have an initial untrimmed accuracy of (20 ppm 1 m) or better.
14、 2.5.4.2 Trimmed Accuracy. This equipment shall have a trimmed accuracy of 2 ppm or better. 2.5.4.3 Long Term Accuracy. The Long Term Accuracy if this equipment, after adjustment, shall be (20 ppm + 0.5 m per year) or better. 2.5.5 Temperature Coefficient. This equipment shall have a Temperature Coe
15、fficient of 20 ppm/C or better. 2.5.6 Resistance Power Coefficient. This equipment shall have a Resistance Power Coefficient of 1.0 ppm/mW/Step, or better. 2.5.7 Maximum Power Rating. The Maximum Power Rating of this equipment shall be greater than or equal to 1.0 Watt/step up to 5 W total, or 2.0 A
16、mps current. 2.5.8 Breakdown Voltage. The breakdown voltage of this equipment shall be 1500 V Peak to the case. 2.5.9 Rack Mount. The equipment shall be capable of being rack-mounted in a standard EIA/ECA-310E equipment rack with an optional rack-mount conversion kit. 2.5.10 Dimensions. The width an
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
2000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- AIRFORCEAA598392009DECADERESISTANCESTANDARD 十进制 电阻 标准 PDF

链接地址:http://www.mydoc123.com/p-427187.html