AIR FORCE A-A-58071 A-2010 TEST SET RADIO FREQUENCY POWER (50 MHz to 40 GHz)《射频功率(50MHz- 40GHz)测试装置》.pdf
《AIR FORCE A-A-58071 A-2010 TEST SET RADIO FREQUENCY POWER (50 MHz to 40 GHz)《射频功率(50MHz- 40GHz)测试装置》.pdf》由会员分享,可在线阅读,更多相关《AIR FORCE A-A-58071 A-2010 TEST SET RADIO FREQUENCY POWER (50 MHz to 40 GHz)《射频功率(50MHz- 40GHz)测试装置》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、 Beneficial comments, recommendations, additions, deletions, clarifications, etc. and any other data that may improve this document should be sent to: WR-ALC/742 CBSG/GBEC, Robins AFB, GA 31098-1611. Since contact information can change, you may want to verify the currency of this address informatio
2、n using the ASSIST Online database at https:/assist.daps.dla.mil . FSC 6625 A-A-58071A 19 April 2010 SUPERSEDING A-A-58071 26 June 1996 COMMERCIAL ITEM DESCRIPTION TEST SET, RADIO FREQUENCY POWER (50 MHz to 40 GHz) 1. SCOPE. This Commercial Item Description (CID) describes the minimum requirements f
3、or a Radio Frequency Power Test Set which consists of a dual-channel peak and continuous wave power meter, sensors, and applicable accessories. The equipment, as described in this CID, will have a minimum frequency range of 50 MHz to 40 GHz and will be used for measuring peak power and continuous wa
4、ve power over selected frequencies. 2. CLASSIFICATION. Equipment covered by this CID shall be commercially available equipment and may be modified to the extent necessary to meet the following description. The equipment shall be Class 3, in accordance with MIL-PRF-28800, except as specified herein.
5、3. SALIENT CHARACTERISTICS. The equipment shall be capable of operation within the accuracies, limits, and specifications herein. 3.1 Safety and Environmental. The equipment shall meet all safety and environmental requirements as specified in MIL-PRF-28800 for the classification as stated herein exc
6、ept as specified herein. 3.1.1 Temperature Operating. The equipment shall conform to its specified performance and accuracy while being operated in a temperature range of 0C to +50C. 3.1.2 Temperature Non-operating. The equipment shall conform to its specified performance and accuracy after being st
7、ored in an environment in the temperature range of -40C to +70C. 3.2 Electrical Power Sources. The equipment shall operate from nominal commercial, military and shipboard power sources of 110/220 volts AC (10%) at line frequencies of 50 to 60 Hz. 3.3 Reliability. The design of the equipment shall be
8、 such that under normal use and operation the equipment does not fail within 2500 hours of operation with a statistical certainty of 95%. 3.4 Calibration and Maintenance Adjustments. The design of the system shall provide for readily accessible calibration adjustments and maintenance adjustments. Th
9、e calibration adjustments, wherever possible, shall be accessible without removal of the equipment case or modules. The calibration interval shall be a period of one year or greater based on an operating time of 2000 hours. No special tools shall be required for alignment or adjustment except those
10、furnished with the equipment. 3.5 Performance Requirements. The equipment shall meet all requirements specified herein, after the warm-up period specified in MIL-PRF-28800, under all combinations of input power conditions, output signal conditions and operating service condition as specified in MIL-
11、PRF-28800. 3.5.1 Meter/Chassis. The peak and continuous wave power meter shall have a minimum of two measurement channels. INCH-POUND Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-58071A 2 3.5.1.1 Remote Operation. The peak and continuous wave
12、power meter shall have an IEEE-488 interface for remote set up, control, and measurement operation. 3.5.1.2 Frequency Measurement Range. The minimum frequency range of the peak and continuous wave power meter with sensors shall be 50 MHz to 40 GHz. 3.5.1.3 Power Measurement Range. The peak and conti
13、nuous wave power meter with sensors shall have, as a minimum, a range of -20 dBm to 20 dBm in the pulse and continuous wave modes. 3.5.1.4 Display/Time Base. The peak and continuous wave power meter shall have a digital display for all functions. The display shall provide an analog representation of
14、 the pulse being measured. The display shall provide, as a minimum, information on the amplitude and timing of the measured signal. 3.5.1.4.1 Timing Information Display. As a minimum, the following pulse timing information shall be provided by the display: rise time, fall time, pulse width, pulse re
15、petition interval or frequency, pulse duty cycle, and delay. 3.5.1.4.1.1 Time Base. The time base on the display shall be selectable, as a minimum, from 10 ns per division to 20 ms per division. The time base resolution shall be 0.2 ns or better. The time base shall be accurate to less than (0.01% o
16、f window) (1ns). 3.5.1.4.2 Amplitude Information Display. Amplitude information shall be available for display in logarithmic form. The equipment shall be capable of displaying the amplitude measurement range specified herein. As a minimum, the following amplitude information shall be available for
17、display: peak power, average power, pulse top amplitude, pulse bottom amplitude, and overshoot. 3.5.1.4.2.1 Range. The equipment shall include a display range of 99.99 dB to +99.99 dB. 3.5.1.4.2.2 Resolution. The display resolution shall be less than 0.05 dB. 3.5.1.4.2.3 Offset. The equipment shall
18、have an amplitude offset capability to include a range from 99.99 dB to +99.99 dB. 3.5.1.4.3 Triggering. 3.5.1.4.3.1 Trigger Sources. Trigger sources shall include internal, external, and manual. Internal triggering shall be available from any sensor channel. Manual triggering shall be available via
19、 the front panel. 3.5.1.4.3.2 Trigger Level. The trigger level range for internal triggering shall include a range of minus 15 dBm to plus 15 dBm. The external trigger range shall be compatible with TTL voltage levels. 3.5.1.4.3.3 Trigger Delay. The equipment shall have a trigger delay capability to
20、 include a range from 0.0 to 200 milliseconds. 3.5.1.5 Operating Modes. 3.5.1.5.1 Peak/Continuous Wave (CW) Selection. Selection of the peak or continuous wave modes of measurements, or equivalents, shall be available via the front panel. 3.5.1.5.2 High/Low Video Bandwidth Selection. Selection of hi
21、gh or low video bandwidth, or equivalents, shall be available via the front panel. 3.5.1.6 Power Reference Calibration Source. An internal power reference source, and associated operating mode, shall be provided to allow user-selected automatic calibration of the combination of the sensor(s) and the
22、 equipment. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-58071A 3 3.5.1.6.1 Frequency. The power calibration source frequency shall be 1.0 GHz 5% or 50 MHz 5%. 3.5.1.6.2 Amplitude. The power reference source amplitude shall be within 1.5% at 1
23、 milliwatt (0 dBm). VSWR shall be less than 1.20 at 0 dBm, 50 ohms. 3.5.1.7 Connectors and Cables. 3.5.1.7.1 Power Reference RF Connector. The power calibration source shall have a Type N RF connector mounted on the front panel of the equipment. 3.5.1.7.2 Interconnect cable. The power meter shall co
24、me with interconnect cables for connecting sensors to the equipment. Each sensor cable shall be detachable from the sensor head and shall be a minimum of 5 feet (1.5 meters) in length. An interconnect cable shall be provided for each channel. 3.5.1.7.3 Adapters. The equipment shall include adapters
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
2000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- AIRFORCEAA58071A2010TESTSETRADIOFREQUENCYPOWER50MHZTO40GHZ 射频 功率 50 MHZ40GHZ 测试 装置 PDF

链接地址:http://www.mydoc123.com/p-426995.html