1、_ A-A-58032 H 9999974 OLb5470 97T INCH POUND October 17, 1994 A-A-58032 COMMERCIAL ITEM DESCRIPTION TEST SET, ELECTRICAL POWER The General Services Administration has authorized the use of this Commercial Item Description ( CID ) for all federal agencies. 1. sco PE. This CID, describes a Test Set, E
2、lectrical Power. This CID is meant as a minimum requirement for the Test Set, Electrical Power in which only those manufacturers that meet or surpass the following requirements are qualified per this CID. The Test Set is to be used for making level, frequency, noise, and noise-to-ground measurements
3、 on telephone lines. 2. CLASSIFICATION, Equipment covered by this CID may be commercially available equipment modified to the extent necessary to meet the following description. The equipment shall be Type III, Class 5, Style E and Color R in accordance with MIL-T-28800. 3. SALIENT CHARACTERISTICS,
4、3.1 Desim and Co nstruction, The equipment shall be UL (Underwriters Laboratories) listed and approved and/or shall comply with the type III safety requirements of MIL-T-28 800. 3.2 Electrical Power Sources. The equipment shall operate from nominal commercial, military and shipboard power sources of
5、 115/230 volts AC (+-lo%) at line frequencies of 50 Hz to 60 Hz. Transient state conditions shall meet MIL-T-28800 Class 5 requirements. The unit shall also operate from an internal, rechargeable battery pack for a minimum of three (3) hours. Beneficial comments, recommendations, additions, deletion
6、s, clarifications etc., and any data which may improve this document should be sent to: The Resources and Logistics Services Division, SA-ALC/TILDD, Bldg. 171, Post C-12,485 Quentin Roosevelt Rd., Kelly AFB TX 78241 -6524. AMSC N/A FSC 6625 DISTRIBUTION S TATEM ENT A. Approved for public release; di
7、stribution is unlimited. Licensed by Information Handling ServicesProvided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,- A-A-56032 W 9999974 OLb5q7L 806 W A-A-58032 3.3 (a) ODeratine Temperature. The equipment shall meet its performance and accuracy requir
8、ements in an operating environment of 0C to +5OoC. (b) -. The equipment shall meet its performance and accuracy requirements in a non-operating environment of -40C to +60C. 3.4 1). The equipment MTBF shall be 2,000 hours. The lower test MTBF shall be 2,000 hoursand the upper test MTBF shall be 4,000
9、 hours. 3.5 Pre-conditionin- Requirement. Each deliverable equipment shall be capable of being subjected to the environmental screening effects of bum in and thermal cycling, with thermal cycling being accomplished in the sample testing phase. 3.6 Calibration and Maintenance Adjustments. The design
10、of the system shall provide for readily accessible calibration adjustments and maintenance adjustments. These adjustments shall be provided by variable value components which are adjustable by the use of simple means. The calibration by substitution of selected components or parts is unacceptable un
11、less specifically approved. The calibration. adjustments, wherever possible, shall be accessible without removal of the instrument case or modules. The calibration interval shall be a period of one year or greater based on an operating time of 2,000 hours. 3.7 Performance Characte ristics. The Test
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
2000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- AIRFORCEAA580321994TESTSETELECTRICALPOWER 电力 测试 装置 PDF

链接地址:http://www.mydoc123.com/p-426931.html