ISO 7991-1987 Glass Determination of coefficient of mean linear thermal expansion《玻璃 热平均线膨胀系数的测定》.pdf
《ISO 7991-1987 Glass Determination of coefficient of mean linear thermal expansion《玻璃 热平均线膨胀系数的测定》.pdf》由会员分享,可在线阅读,更多相关《ISO 7991-1987 Glass Determination of coefficient of mean linear thermal expansion《玻璃 热平均线膨胀系数的测定》.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、INTERNATIONAL STANDARD INTERNATIONAL ORGANIZATION FOR STANDARDIZATION ORGANISATION INTERNATIONALE DE NORMALISATION MEXjJYHAPOjJHAR OPrAHM3AMR IlO CTAH/JAPTkl3AMM 1 d 11 ,j b : .y 2. ,- : n*.P* B .i , _, -;_, i.i.ji,*:fs *:Li t) : The ratio of the change in length of a specimen within a temperature i
2、nterval to that temperature interval, related to the initial specimen length. It is given by the following equation: I - I, cd,; t) = f x - . . . 0 t - to to is the initial or reference temperature; t is the actual (constant or variable) specimen tem- perature; I, is the length at temperature to of
3、the specimen of glass under test (usually a rod made from the glass); I is the specimen length at temperature t. For the purposes of this International Standard, the nominal reference temperature, to, is 20 OC; therefore the coefficient of mean linear thermal expansion is denoted by a(20 YZ; t). 4 A
4、pparatus 4.1 Device for measuring the specimen length, to an ac- curacy of 0,l %. 4.2 Push-rod dilatometer, capable of determining changes in length of the specimen of 2 x 10e5 L, (i.e. 2 pm per 100 mm). The contact force of, the extensometer should not exceed 1 N. That force shall act through conta
5、cts of planes with spherical faces the radii of curvature of which shall be not less than the rod diameter of the specimen. In some special assemblies (see figure I) parallel planes are needed. The specimen-holding assembly shall ensure that the specimen is held firmly in position and shall prevent
6、even small changes in its alignment with respect to the push-rod axis throughout the test (see examples given in the annex). If the specimen-holding assembly is made of vitreous silica, see the precautions given in 7.2. From time to time, a performance test shall be carried out using a reference mat
7、erial (see clause 8). 4.3 Furnace, compatible with the dilatometer assembly, for temperatures up to 50 “C above the expected transformation temperature. The working position of the furnace relative to the dilatometer assembly shall be defined with a repeatability of 0,5 mm in both the axial and the
8、radial directions. Within the range of testing temperatures (i.e. up to tempera- tures about 150 OC below the highest expected transformation temperature, ty and at least up to 300 Cl, the furnace shall be capable of marntaining a constant temperature to + 2 OC over the whole specimen length. 4.4 Fu
9、rnace control device, suitable for the desired rate of increase in temperature up to (5 + I) OC/min within the test range (see 6.1) and for a cooling rate of (2 + 0,2) “Urnin for the annealing procedure according to 5.2. 1 IS0 7991 : 1967 (El 4.5 Temperature-measuring device (e.g. a thermocouple of
10、type E, J or K in accordance with IEC 684-11, capable of determining the temperature of the specimen to f 2 OC in the temperature range between to and t. 5 Test specimen 5.1 Shape and size The test specimen is usually in the form of a rod. Its shape depends on the type of dilatometer used. The lengt
11、h I, shall be at least 5 x lo4 times the resolution of the dilatometers measuring device for the change in length. NOTE - The specimen may be, for example, a rod either with a cir- cular cross-section having a diameter of 5 mm or with a square cross- section 5 mm x 5 mm, and between 25 and 100 mm in
12、 length. In cer- tain cases, a cross-section of at least 100 mm2 is more convenient (see the annex). 5.2 Preparation The test specimen shall be annealed before the test by heating it to about 30 YZ above the transformation temperature and then cooling it to about 150 OC below the transformation temp
13、erature at a rate of (2 + 0,2) Wmin, followed by further cooling to room temperature in draught-free air. 5.3 Number The test shall be carried out with two test specimens (see also 7.4). 6 Procedure 6.1 Choice of the test range In accordance with clause 3, the nominal reference temperature is 20 OC.
14、 For practical reasons, however, the measurement may be started between 18 and 28 OC. The preferred final actual temperature is 290 OC t 310 OC. If this is not practical, then the alternative values 190 OC t 210 OC, or, in special cases, !XC t 105OC or 390C t 410C may be chosen. The corresponding no
15、minal values of t are 300 OC, 200 OC, 100 OC, and 400 OC, respectively. All readings of temperatures and temperature differences shall be taken with an accuracy to 2 OC. Though these actual values are used in the calculations in accordance with clause 7, the test range shall be expressed in terms of
16、 the nominal tem- peratures (see 7.4). For a given coefficient a(20 OC; t) ex- pressed in terms of the nominal temperature, no influence on the value of the coefficient can be detected within the limits specified for the preferred actual temperatures. 6.2 Determination of the reference length Determ
17、ine the reference length 1, of the annealed specimen (see 5.2) to an accuracy of 0,l % at the reference temperature t,. Subsequently insert the specimen into the dilatometer and wait for about 5 min before beginning the test as described in 6.3 or 6.4. 6.3 Test at increasing temperature Determine th
18、e position of ;he dilatometer at the initial temperature to and take this reading as zero for the uncorrected change in length, AI, which will be measured. Subse- quently set the furnace control device (4.4) to the desired heating programme and start the programme. Record the temperature t and the r
19、elated change in length Al, until the desired final temperature has been reached. NOTE - The rate of temperature increase should not exceed 5 Wmin. As the dilatometer readings of Al, are recorded during the increase in the temperature between to and t (values chosen in accordance with 6.11, it shoul
20、d be borne in mind that a temperature difference will exist between the hot junction of the thermocouple and the test specimen; therefore a correction shall be applied to the apparent temperature of the test specimen. NOTE - The magnitude of this correction depends on the rate of temperature change
21、and the rate of heat exchange between the fur- nace and the test specimen. It is essential that the correction is deter- mined experimentally by comparison with measurements at constant temperatures. 6.4 Test at constant temperature Determine the position of the dilatometer at the initial temperatur
22、e to and take this reading as zero for the uncorrected change in length, AI, which will be measured. Subse- quently heat the furnace to the selected final temperature t and hold it constant to f 2 OC for 20 min. Then take from the dilatometer reading the value of Al,. NOTE - Although the test at inc
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ISO79911987GLASSDETERMINATIONOFCOEFFICIENTOFMEANLINEARTHERMALEXPANSION 玻璃 平均 线膨胀 系数 测定 PDF

链接地址:http://www.mydoc123.com/p-1255674.html