ISO 2072-1981 Aluminium oxide primarily used for the production of aluminium Determination of zinc content PAN photometric method《主要用于铝生产的氧化铝 锌含量的测定 PAN光度法》.pdf
《ISO 2072-1981 Aluminium oxide primarily used for the production of aluminium Determination of zinc content PAN photometric method《主要用于铝生产的氧化铝 锌含量的测定 PAN光度法》.pdf》由会员分享,可在线阅读,更多相关《ISO 2072-1981 Aluminium oxide primarily used for the production of aluminium Determination of zinc content PAN photometric method《主要用于铝生产的氧化铝 锌含量的测定 PAN光度法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、International Standard f! 2072 0 + 4 *!ig!B INTERNATIONAL ORGANIZATION FOR STANDARDIZATIONMEYtlAPOAtiAfl OPAHH3AL alternatively, a suitable plastics material may be used. Use only ground glass or plastics (not rubber) stoppers. The apparatus specified in IS0 804, clause 5, and 5.1 Burette, graduated
2、 in 0,05 ml divisions, complying with the requirements of ISO/R 385. 5.2 pH meter. 5.3 Spectrophotometer, or 5.4 Photoelectric absorptiometer, fitted with filters ensur- ing maximum transmission at a wavelength of about 560 nm. 6 Procedure 6.1 Test portion Take an aliquot portion of the principal so
3、lution P prepared in accordance with IS0 804, sub-clauses 6.1, 6.2 and 6.3, con- taining 10 to 50 ug of ZnO, and place it in a beaker of suitable capacity (for example 250 ml). 6.2 Blank test Carry out a blank test at the same time as the determination, following the same procedure, on a solution pr
4、epared without the addition of extra-pure aluminium oxide, in accordance with IS0 804, sub-clause 6.4.2, using the same quantities of reagents as used for the determination. 2 Copyright International Organization for Standardization Provided by IHS under license with ISONot for ResaleNo reproduction
5、 or networking permitted without license from IHS-,-,-IS0 2072-1981 (E) 6.3 Preparation of the calibration graph 6.3.1 Preparation of standard calorimetric solutions, for photometric measurements carried out in cells of optical path length 1 cm By means of the burette (5.11, transfer into a series o
6、f seven beakers of suitable capacity (for example 100 ml), the volumes of the standard zinc solution (4.13) shown in the following table. Standard zinc solution (4.131 ml 0” I.0 70 38 48 5.0 3.0 Corresponding mass of ZnO L where b) the reference of the method used; m. is the mass, in grams, of the t
7、est sample taken for the preparation of the principal solution P; ml is the mass, in micrograms, of ZnO found in the aliquot portion of the test solution (6.4.1); m2 is the mass, in micrograms, of ZnO found in a cor- responding aliquot portion of the blank test solution (6.2); cl the results and the
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ISO20721981ALUMINIUMOXIDEPRIMARILYUSEDFORTHEPRODUCTIONOFALUMINIUMDETERMINATIONOFZINCCONTENTPANPHOTOMETRICMETHOD

链接地址:http://www.mydoc123.com/p-1252287.html