ISO 19318-2004 Surface chemical analysis - X-ray photoelectron spectroscopy - Reporting of methods used for charge control and charge correction《表面化学分析 X射线光电光谱法.pdf
《ISO 19318-2004 Surface chemical analysis - X-ray photoelectron spectroscopy - Reporting of methods used for charge control and charge correction《表面化学分析 X射线光电光谱法.pdf》由会员分享,可在线阅读,更多相关《ISO 19318-2004 Surface chemical analysis - X-ray photoelectron spectroscopy - Reporting of methods used for charge control and charge correction《表面化学分析 X射线光电光谱法.pdf(18页珍藏版)》请在麦多课文档分享上搜索。
1、 Reference number ISO 19318:2004(E) ISO 2004INTERNATIONAL STANDARD ISO 19318 First edition 2004-05-01 Surface chemical analysis X-ray photoelectron spectroscopy Reporting of methods used for charge control and charge correction Analyse chimique des surfaces Spectroscopie de photolectrons Indication
2、des mthodes mises en oeuvre pour le contrle et la correction de la charge ISO 19318:2004(E) PDF disclaimer This PDF file may contain embedded typefaces. In accordance with Adobes licensing policy, this file may be printed or viewed but shall not be edited unless the typefaces which are embedded are
3、licensed to and installed on the computer performing the editing. In downloading this file, parties accept therein the responsibility of not infringing Adobes licensing policy. The ISO Central Secretariat accepts no liability in this area. Adobe is a trademark of Adobe Systems Incorporated. Details
4、of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In the unlikely event that a problem relat
5、ing to it is found, please inform the Central Secretariat at the address given below. ISO 2004 All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying and microfilm, witho
6、ut permission in writing from either ISO at the address below or ISOs member body in the country of the requester. ISO copyright office Case postale 56 CH-1211 Geneva 20 Tel. + 41 22 749 01 11 Fax + 41 22 749 09 47 E-mail copyrightiso.org Web www.iso.org Published in Switzerland ii ISO 2004 All righ
7、ts reservedISO 19318:2004(E) ISO 2004 All rights reserved iiiContents Page Foreword iv Introduction v 1 Scope 1 2 Normative reference. 1 3 Terms and definitions. 1 4 Symbols and abbreviated terms 1 5 Apparatus. 2 6 Calibration of binding-energy scale 2 7 Reporting of information related to charge co
8、ntrol 2 7.1 Methods of charge control. 2 7.2 Information on specimen . 2 7.3 Instrument and operating conditions . 3 7.4 General method for charge control. 3 7.5 Reasons for needing charge control and for choosing the particular method for charge control 3 7.6 Values of experimental parameters 3 7.7
9、 Information on the effectiveness of the method of charge control . 4 8 Reporting of method(s) used for charge correction and the value of that correction. 4 8.1 Methods of charge correction . 4 8.2 Approach 4 8.3 Value of correction energy. 4 Annex A (informative) Description of methods of charge c
10、ontrol and charge correction 5 A.1 Introduction. 5 A.2 Methods of charge control. 5 A.3 Methods of charge correction . 7 A.4 Bias referencing17 . 8 A.5 Auger parameter measurements32-34 . 9 Bibliography . 10 ISO 19318:2004(E) iv ISO 2004 All rights reservedForeword ISO (the International Organizatio
11、n for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has
12、 the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization. Inter
13、national Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2. The main task of technical committees is to prepare International Standards. Draft International Standards adopted by the technical committees are circulated to the member bodies for voting. Publicat
14、ion as an International Standard requires approval by at least 75 % of the member bodies casting a vote. Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. ISO shall not be held responsible for identifying any or all such patent righ
15、ts. ISO 19318 was prepared by Technical Committee ISO/TC 201, Surface chemical analysis, Subcommittee SC 5, Auger electron spectroscopy. ISO 19318:2004(E) ISO 2004 All rights reserved vIntroduction X-ray photoelectron spectroscopy (XPS) is widely used for characterization of surfaces of materials. E
16、lements in the test specimen (with the exception of hydrogen and helium) are identified from comparisons of the binding energies of their core levels, determined from measured photoelectron spectra, with tabulations of these binding energies for the various elements. Information on the chemical stat
17、e of the detected elements can frequently be obtained from small variations (typically between 0,1 eV and 10 eV) of the core-level binding energies from the corresponding values for the pure elements. Reliable determination of chemical shifts often requires that the binding-energy scale of the XPS i
18、nstrument be calibrated with an uncertainty that could be as small as 0,1 eV. The surface potential of an insulating specimen will generally change during an XPS measurement due to surface charging, and it is then difficult to determine binding energies with the accuracy needed for elemental identif
19、ication or chemical-state determination. There are two steps in dealing with this problem. First, experimental steps can be taken to minimize the amount of surface charging (charge-control methods). Second, corrections for the effects of surface charging can be made after acquisition of the XPS data
20、 (charge- correction methods). Although the buildup of surface charge can complicate analysis in some circumstances, it can be creatively used as a tool to gain information about a specimen. The amount of induced charge near the surface, its distribution across the specimen surface, and its dependen
21、ce on experimental conditions are determined by many factors including those associated with the specimen and characteristics of the spectrometer. Charge buildup is a well-studied1, 2three-dimensional phenomenon that occurs along the specimen surface and into the material. Charge buildup may also oc
22、cur at phase boundaries or interface regions within the depth of the specimen that is irradiated by X-rays. Some specimens undergo time-dependent changes in the level of charging because of chemical changes or volatilization induced by photoelectrons and secondary electrons, X-rays, or heating. Such
23、 specimens may never achieve steady-state potentials. There is, at present, no universally applicable method or set of methods for charge control or for charge correction3, 4 . This International Standard specifies the information that shall be provided to document the method of charge control durin
24、g data acquisition and/or the method of charge correction during data analysis. Information is given in Annex A on common methods for charge control and charge correction that can be useful for many applications. The particular charge-control method that may be chosen in practice depends on the type
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ISO193182004SURFACECHEMICALANALYSISXRAYPHOTOELECTRONSPECTROSCOPYREPORTINGOFMETHODSUSEDFORCHARGECONTROLANDCHARGECORRECTION

链接地址:http://www.mydoc123.com/p-1252138.html