1、11-11 10 25 13 - 1 SECTION 10 25 13 PATIENT BED SERVICE WALLS SPEC WRITER NOTE: Delete between / - / if not applicable to project. Due to the very specialized nature of Patient Wall Systems, caution must be exercised before deleting any other item or paragraph not between / - /. When paragraphs are
2、deleted, renumber remaining paragraphs. Coordinate this specification section with the drawings. PART 1 GENERAL 1.1 DESCRIPTION This section specifies the furnishing, installation and connection of the patient wall systems both horizontal and vertical. Patient wall systems are also referred to as pr
3、efabricated bedside patient units or PBPUs. 1.2 RELATED WORK A. Section 09 06 00, SCHEDULE FOR FINISHES: Color and finishes of the patient wall units. B. Section 22 62 00, VACUUM SYSTEMS FOR LABORATORY AND HEALTHCARE FACILITIES AND Section 22 63 00, GAS SYSTEMS FOR LABORATORY AND HEALTHCARE FACILITI
4、ES: Requirements for air, oxygen and vacuum outlets in the patient wall units. C. Section 26 05 11, REQUIREMENTS FOR ELECTRICAL INSTALLATIONS: General electrical requirements that are common to more than one section of Division 26. D. Section 26 05 33, RACEWAY AND BOXES FOR ELECTRICAL SYSTEMS: Racew
5、ays and outlet boxes for wiring. E. Section 26 05 21, LOW-VOLTAGE ELECTRICAL POWER CONDUCTORS AND CABLES (600 VOLTS AND BELOW): Cables and wiring. F. Section 26 27 26, WIRING DEVICES: Wiring devices to be installed in the patient wall units. G. Section 26 24 16, PANELBOARDS: Panelboard requirements
6、for patient wall units with a panelboard. H. Section 26 05 26, GROUNDING AND BONDING FOR ELECTRICAL SYSTEMS: Requirements for personnel safety and to provide a low impedance path to ground for possible ground currents. I. Section 26 51 00, INTERIOR LIGHTING: Lighting fixture requirements when instal
7、led in or connected to the patient wall units. J. Section 27 52 23, NURSE CALL/CODE BLUE SYSTEMS: Nurse Call and Code One requirements for installation in the patient wall units. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-11-11 10 25 13 - 2 1.3
8、SUBMITTALS A. In accordance with Section 26 05 11, REQUIREMENTS FOR ELECTRICAL INSTALLATIONS, submit the following: B. Shop Drawings: 1. Sufficient information, clearly presented, shall be included to determine compliance with drawings and specifications. 2. Include electrical ratings, dimensions, m
9、ounting details, front view, side view, equipment and device arrangement, wiring diagrams, material, and connection diagrams. 3. Determine final layout of each style of patient wall system at this stage. Provide configuration drawings showing all possible device (nurse call, medical gases, electrica
10、l receptacles and switches, etc.) locations to the Resident Engineer. The Resident Engineer will provide by return of submittal the desired configuration of each style of patient wall system. Limit the number and type of devices allowed for each style of unit to the number and type of devices specif
11、ied for that style below. C. Manuals: Two weeks prior to the final inspection, deliver four copies of the following to the Resident Engineer. 1. Complete maintenance and operating manuals including wiring diagrams, technical data sheets, and information for ordering replacement parts: a. Include com
12、plete “As installed“ diagrams which indicate all items of equipment, their interconnecting wiring and interconnecting piping. b. Include complete diagrams of the internal wiring for each of the items of equipment, including “As installed“ revisions of the diagrams. c. Identify terminals on the wirin
13、g diagrams to facilitate installation, maintenance and operation. D. Certifications: Two weeks prior to the final inspection, deliver four copies of the following certifications to the Resident Engineer: 1. Certification by the manufacturer that the equipment conforms to the requirements of the draw
14、ings and specifications. 2. Certification by the Contractor that the equipment has been properly installed, adjusted, and tested in accordance with the manufacturers recommendations. 1.4 APPLICABLE PUBLICATIONS: A. Publications listed below (including amendments, addenda, revisions, supplements and
15、errata) form a part of this specification to the extent Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-11-11 10 25 13 - 3 referenced. Publications are referenced in text by the basic designation only. B. National Fire Protection Association (NFPA):
16、70-11 National Electrical Code (NEC) 99-12 Health Care Facilities C. Underwriters Laboratories, Inc. (UL): UL listed in product category SECTIONS AND UNITS (QQXX). This standard used to investigate listed products in this category is NFPA 70 (NEC). PART 2 - PRODUCTS 2.1 PATIENT WALL SYSTEMS A. Shall
17、 be UL listed. B. Shall consist of a structural framework, removable panels and removable equipment console units, factory assembled to house all permanent bedside services including but not necessarily limited to fixtures, grounding jacks, power outlets, telephone outlet, nurses call patient statio
18、n, medical gas outlet(s) and other fittings or devices. C. Shall conform to the following: 1. Applicable requirements in NFPA 70 (NEC) and NFPA 99. 2. Assembly and all components shall be UL listed or labeled. D. Coordinate the mounting space provisions for the nurse call equipment with Section 27 5
19、2 23, NURSE CALL/CODE BLUE SYSTEMS. SPEC WRITER NOTE: Due to the extremely sensitive nature of patent infringement the following paragraph may be modified only with consultation with VA Central Office (VHA). E. Compressed Air, Oxygen and Vacuum System Equipment: Furnish, install and test the equipme
20、nt in accordance with the drawings and Section 22 62 00, VACUUM SYSTEMS FOR LABORATORY AND HEALTHCARE FACILITIES and Section 22 63 00, GAS SYSTEMS FOR LABORATORY AND HEALTHCARE FACILITIES. 1. Fixed medical gas outlets are permanently installed in one location and may not be moved without special too
21、ls and shutting off the gas involved. 2. Movable medical gas outlets: a. Hose connected to gas manifold type: 1) The hoses connected to gas manifold shall be UL listed and labeled for the purpose. SPEC WRITER NOTE: Due to the extremely sensitive nature of patent infringement do not modify the follow
22、ing paragraph. 2) All hoses shall be accessible at all times. Use bars or other restraining devices to control exposed hoses. A panel may cover Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-11-11 10 25 13 - 4 the hoses provided it can be easily rem
23、oved with out the use of special tools for hose inspection. b. Relocatable type: 1) Relocatable (snap-in) without the use of tools to any one of several different fixed locations. 2) Appropriate relocatable adapter can be used to access available gases from each fixed location. 3) Cover all unused l
24、ocations with a blank (no gas) adapter plate. F. Electrical receptacles and switches shall comply with the requirements in Section 26 27 26, WIRING DEVICES; grounding in Section 26 05 26, GROUNDING AND BONDING FOR ELECTRICAL SYSTEMS; and internal wiring in Section 26 05 21, LOW-VOLTAGE ELECTRICAL PO
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- VA1025132011PATIENTBEDSERVICEWALLSPDF