UOP 939-1996 BASIC NITROGEN IN LPG BY ION CHROMATOGRAPHY.pdf
《UOP 939-1996 BASIC NITROGEN IN LPG BY ION CHROMATOGRAPHY.pdf》由会员分享,可在线阅读,更多相关《UOP 939-1996 BASIC NITROGEN IN LPG BY ION CHROMATOGRAPHY.pdf(12页珍藏版)》请在麦多课文档分享上搜索。
1、BASIC NITROGEN IN LPG BY ION CHROMATOGRAPHYUOP Method 939-96SCOPEThis method is for determining basic nitrogen in liquefied petroleum gas (LPG). The method determinesbasic nitrogen species that are extractable by aqueous acid, including ammonia and alkyl amines, whichtypically adsorb on the stainles
2、s steel surface of a LPG cylinder. This method is limited to ammonia,primary amines containing four or less carbon atoms, secondary amines containing eight or less carbonatoms and tertiary amines with nine or less carbon atoms. All linear aliphatic substituents of the amineanalytes must contain less
3、 than five carbon atoms. Combined nitrogen can be determined using UOPMethod 936. Nitriles are not detected by this method. Amides and/or ionic sodium, if present at aconcentration of 10 mass-ppm or greater, will interfere with the quantitation of ammonia. Potassium ions inthe acid extract can also
4、interfere with the detection of methylamine and ethylamine. The procedure yieldsquantitative basic nitrogen results in the range of 0.1 mass-ppm to 3 mass-%.OUTLINE OF METHODNitrogen compounds are extracted from an LPG sample using a solution of acetic acid in water accordingto the procedure specifi
5、ed in UOP Method 936. The sample extract is diluted and injected into an ionchromatograph equipped with a cation exchange column, conductivity suppressor device and conductivitydetector. Analytes are separated by cation exchange. Analyte column retention is based primarily on ionicsize and charge. T
6、he column effluent enters a conductivity suppressor device, where the backgroundconductivity of the column effluent is reduced and the analytes are converted into more conductive forms.The analyte effluent from the conductivity suppressor device flows to a conductivity detector wheresolution conduct
7、ance is continuously measured. The solution conductance is proportional to ionicconcentration and the inherent mobility of each ion present in the detector flow cell.An external standard method of quantitation is used. The best quantitation is achieved by matchingabsolute calibration blends to known
8、 sample analyte composition. If the analyte composition of a sample isunknown, representative response factors are determined by analyzing calibration blends that provide arepresentative mean detector response.IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TODETE
9、RMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH ANDSAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THISPROCEDURE IN THE MANNER PRESCRIBED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS(MSDS) OR EXPERIMENTAL MATERIAL SAFETY
10、DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED INTHIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTIONEQUIPMENT (PPE). COPYRIGHT 1996 UOP LLCALL RIGHTS RESERVEDUOP Methods are available through ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohoc
11、ken PA 19428-2959,United States. The Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service atserviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE.2 of 12939-96APPARATUSReferences to catalog numbers and suppliers are included as a convenience to the met
12、hod user. Othersuppliers may be used.In addition to the equipment listed below, all of the acid extraction related apparatus specified in UOPMethod 936 is also required for this method.Balance, readability 0.1-mgCation MicroMembrane Suppressor, conductivity suppressor device, Dionex CMMS-II, Dionex,
13、 Cat. No.43021Cation MicroMembrane Suppressor installation kit, includes one 4-L pressurizable regenerant reservoir,one 0 to 30 psi pressure regulator, and all tubing and fittings required to install the CMMS-II, Dionex,Cat. No. 38018Chromatographic column, IonPac CS3 Analytical, dimensions 250-mm l
14、ong by 4-mm ID, Dionex, Cat.No. 37024Cylinders, graduated, borosilicate glass, 10-, 50-, 250-, 500- and 1000-mL, Fisher Scientific, Cat. Nos.08-556-A, C, E, F and G, respectivelyFilter apparatus, all glass filter, 47-mm diameter, 300-mL capacity, one-liter flask, Millipore, Cat. No.XX1504700Filters,
15、 Nylon-66, 47-mm diameter, 0.45-m pore size, Rainin Instrument, Cat. No. 38-114Flask, Erlenmeyer, 4-L Pyrex glass, Fisher Scientific, Cat. No. 10-040PFlask, filtering, 4-L Pyrex glass with tubulation, Fisher Scientific, Cat. No. 10-180HFlasks, volumetric, 2000-mL, with caps, Fisher Scientific, Cat.
16、No. 10-210-5HIntegrator, electronic, for obtaining peak areasIon chromatograph, equipped with a Dionex APM-2 analytical pump, a Dionex Universal Sampler withauxiliary sampling pump, a Dionex CDM-2 conductivity detector and a Dionex AdvancedChromatography Module CHA-2 system controllerMagnetic stir b
17、ar, 0.5-inch diameter x 3-inch length, octagonal shape, covered with Teflon fluorocarbonresin, Fisher Scientific, Cat. No. 14-511-85Mobile Phase Reservoir, 4-L pressurizable mobile phase reservoir and plastic tubing, Dionex, Cat. No.380183 of 12939-96Pasteur pipet, 146-mm length disposable borosilic
18、ate glass, Fisher Scientific, Cat. No. 13-678-20ARecorder, 10-mV full scaleRegulator, nitrogen, two-stage, Matheson Gas Products, Model 3122-580Reservoir, 4-L, propylene bottle, Dionex, Cat. No. 39164Sample vials, for dilution of acid extract, Kimble Scientific, Cat. No. 60957-1, with Quorpack black
19、phenolic caps and polyseal liners, Cat. No. 02-883-5G, all available from Fisher ScientificStir plate, magnetic stirrer with 7- x 7-inch ceramic stirring surface, Fisher Scientific, Cat. No. 11-496-79Test tubes, 16- x 100-mm borosilicate glass culture tubes with plain end, Fisher Scientific, Cat. No
20、. 14-961-29Tubing, vacuum, 1/4-inch ID (6.4-mm), Fisher Scientific, Cat. No. 14-173CVacuum pump, house vacuum can be substitutedValve, needle, Supelco, Cat. No. 2-2116Vials, 6 dramWater purification system, Barnstead NANOpure II, 3 module, equipped with Pretreatment (D0835),Ultrapure (D0809) and Org
21、anic Free (D0820) replacement cartridges, Fisher ScientificREAGENTS AND MATERIALSAll reagents shall conform to the specifications established by the Committee on Analytical Reagents ofthe American Chemical Society, when such specifications are available, unless otherwise specified.References to wate
22、r mean distilled and deionized water that is further treated with a Barnstead NANOpureII or other equivalent water purification system.References to catalog numbers and suppliers are included as a convenience to the method user. Othersuppliers may be used.In addition to the reagents listed below, th
23、e reagents specified in UOP Method 936 are also required forthis method.Acetic acid, approximately 99.7% purity, Fisher Certified, Fisher Scientific, Cat. No. A38-212Acetic acid, approximately 0.4-N. Prepare two liters by slowly adding 46 mL of concentrated acetic acidfrom a graduated cylinder to ap
24、proximately 1500 mL of water in a 2000-mL volumetric flask. After theheat of dilution dissipates, dilute to volume with water and mix thoroughly by capping the flask andinverting at least 10 times.4 of 12939-96Ammonium chloride, 99.998% pure, Aldrich Chemical, Cat. No. 25,413-4Dibutylamine, 99+% pur
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9391996BASICNITROGENINLPGBYIONCHROMATOGRAPHYPDF

链接地址:http://www.mydoc123.com/p-1074578.html