《UOP 936-1995 COMBINED NITROGEN IN LPG BY CHEMILUMINESCENCE.pdf》由会员分享,可在线阅读,更多相关《UOP 936-1995 COMBINED NITROGEN IN LPG BY CHEMILUMINESCENCE.pdf(16页珍藏版)》请在麦多课文档分享上搜索。
1、COMBINED NITROGEN IN LPG BY CHEMILUMINESCENCE UOP Method 936-95 SCOPE This method is for determining combined nitrogen in liquefied petroleum gas (LPG). The method determines all nitrogen compounds that are extractable by aqueous acid, including ammonia, amines, nitriles and amides. The method is pa
2、rticularly useful for LPG samples containing nitrogen species, such as ammonia and amines, which can readily adsorb on stainless steel surfaces in contact with the LPG. The procedure yields quantitative results in the range of approximately 0.1 mass-ppm to approximately 3 mass-%. OUTLINE OF METHOD N
3、itrogen compounds are quantitatively extracted from an LPG sample using a solution of 2-N acetic acid in water. A measured mass of LPG is collected in a cleaned, stainless steel, sample cylinder. A known amount of acid solution is added to a second cylinder, the two vessels are connected and the aci
4、d solution is transferred under helium pressure into the LPG sample cylinder. The nitrogen components are extracted into the acid solution by mixing the liquids for several minutes. The LPG and acid extract are allowed to separate. The acid extract is withdrawn and analyzed for total nitrogen by che
5、miluminescence, utilizing ASTM Method D 4629 with the optional boat inlet system. APPARATUS References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Adapter, male, 2 required, for charging acetic acid cylinder, available from Dearborn
6、 Valve and Fitting, Swagelok, Cat. No. SS-4-TA-1-4 (see also Connector) Balance, readability 0.1-mg Balance, readability 0.1-g, capacity 4000-g Beaker, polypropylene, 100-mL, Fisher Scientific, Cat. No. 02-591-17B IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO
7、 DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIBED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL
8、 SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1995 UOP LLC ALL RIGHTS RESERVED UOP Methods are available through ASTM International, 100 Barr Harbor Drive, PO Box C700, We
9、st Conshohocken PA 19428-2959, United States. The Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE. 2 of 16 936-95 Calculator, capable of least squares linear regression calculation, required
10、 only if the nitrogen analyzer does not permit direct calibration and direct reading of the nitrogen in the extract Clamps, for ring stand and sample cylinders, several required, with holders and cast iron ring, Fisher Scientific, Cat. Nos. 05-769-8, 05-754 and 14-050C, respectively Connector, stain
11、less steel, used to connect the 300-mL and the 1000-mL cylinders vertically together and consisting of a male connector, Swagelok, Cat. No. SS-400-1-4 joined together with one male adapter, Swagelok, Cat. No. SS-4-TA-1-4. All parts are available from Dearborn Valve and Fitting. Cylinders, for sample
12、 collection and analysis, type 304 stainless steel, high pressure, 12,400 kPa gauge (1800 psi) maximum working pressure, 1000- and 300-mL capacity, Whitey, Cat. Nos. SS-304L-HDF4-1000 and -300, respectively. Equip the cylinders with two stainless steel valves each, Whitey, Cat. No. SS-16DKM4-F4, and
13、 two pipe-plugs each, Cajon, quarter-inch P-male-NPT, Swagelok, Cat. No. SS-4-P. All parts are available from Dearborn Valve and Fitting. Cylinders, graduated, borosilicate glass, 50-, 100- and 250-mL, Fisher Scientific, Cat. Nos. 08-556C, D and E, respectively Extractor (optional), special construc
14、tion, see APPENDIX Flasks, volumetric, 100-, 1000- and 2000-mL, with caps, Fisher Scientific, Cat. Nos. 10-210-5C, -5G and -5H, respectively Gauge, pressure, 0-20, 700 kPa gauge (0-3000 psig), 2-inch diameter, 0.25-inch MNPT, used for testing the pressure of an unknown LPG sample gas cylinder, Mathe
15、son Gas Products, Model 63-2233 Jar-Mill (optional), 20-300 rpm, Model 755 RMV, Fisher Scientific, Cat. No. 08-381-1. Required only if optional extractor is used. Nitrogen analyzer, total nitrogen by chemiluminescence, with boat control system for aqueous sample, Mitsubishi Instruments, TN-10, avail
16、able from Cosa Scientific Oven, drying, capable of maintaining 105C Pipet, Mohr, 1-, 5- and 25-mL, Fisher Scientific, Cat. Nos. 13-665-F, -K and -N, respectively Quartz beads, -10 to + 60 mesh, Pyromatics, Cat. No. 7323 Regulators, helium, two-stage, Matheson Gas Products, Models 3104C-580 (used for
17、 the nitrogen analyzer) and 3030-580 (used for acetic acid transfer between cylinders) Regulator, oxygen, two-stage, high purity, Matheson Gas Products, Model 3813-CGA-540, cleaned for oxygen service by the supplier, used with Nitrogen Analyzer Ring stand, with rectangular base, 170- x 280-mm base w
18、ith 910-mm rod, Fisher Scientific, Cat. No. 14-670D 3 of 16 936-95 Sample vials, for extracted sample, Kimble Scientific Products, Cat. No. 60957-1, with Quorpack black phenolic caps and polyseal liners, Cat. No. 02-883-5G, available from Fisher Scientific Syringe, 50-L, Hamilton, Model 705N, Fisher
19、 Scientific Products, Cat. No. 14-824-5 Tubing, polyethylene, “POLYFLOW”, for charging acetic acid cylinder, 6.4-mm (0.25-inch) ID, Lesman Instruments, Cat. No. 44-P-N Tubing, Tygon, R3603, for charging acetic acid cylinder 6.4-mm (0.25-inch) ID, Fisher Scientific, Cat. No. 14-169-3C Tubing, vacuum,
20、 1/4-inch ID (6.4-mm), Fisher Scientific, Cat. No. 14-173C Vacuum pump, aspirator, Nalgene, Fisher Scientific, Cat. No. 09-960-2 Valve, check, 20,700 kPa gauge (3000 psig), Matheson Gas Products, Model 400V REAGENTS AND MATERIALS All reagents shall conform to the specifications established by the Co
21、mmittee on Analytical Reagents of the American Chemical Society, when such specifications are available, unless otherwise specified. References to water mean deionized or distilled water. References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers ma
22、y be used. Acetic acid, approximately 99.7%, Fisher Certified, Fisher Scientific, Cat. No. A38-212 Acetic acid, approximately 2-N. Prepare two liters by slowly adding 232 mL of concentrated acetic acid from a graduated cylinder to approximately 1500 mL of water in a 2000-mL volumetric flask. After t
23、he heat of dilution dissipates, dilute to volume with water and mix thoroughly by capping the flask and inverting at least 10 times. Acetone, Fisher Certified, Fisher Scientific, Cat. No. A18-4 Helium, Ultra High Purity, 99.999%, Matheson Gas Products Isooctane, Certified, Fisher Scientific, Cat. No
24、. 0299-4 Isooctane-acetone, 9 to 1, Measure 900 mL of isooctane into a 1000-mL volumetric flask and fill with acetone to the mark. Cap and mix thoroughly by inverting at least ten times. Methanol, Fisher Certified, ACS, Fisher Scientific, Cat. No. A412-4 Methanol-acetone, 1 to 1. Measure 500 mL of a
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9361995COMBINEDNITROGENINLPGBYCHEMILUMINESCENCEPDF

链接地址:http://www.mydoc123.com/p-1074575.html