UOP 919-2011 Trace Hydrocarbons in Udex Solvent by GC.pdf
《UOP 919-2011 Trace Hydrocarbons in Udex Solvent by GC.pdf》由会员分享,可在线阅读,更多相关《UOP 919-2011 Trace Hydrocarbons in Udex Solvent by GC.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1992, 2011 UOP LLC. All rights reserve
3、d. Nonconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.83
4、2.9585 PHONE. Trace Hydrocarbons in Udex Solvent by GC UOP Method 919-11 Scope This method is for determining low concentrations of hydrocarbons in lean UOP Udex process solvent. Benzene, toluene, C8 aromatics composite, and C9+ aromatics composite are determined. Light non-aromatics through C9 para
5、ffins, if present, are reported as a composite. Higher non- aromatics, i.e., greater than C9, may interfere with the determination of benzene. The lower limit of quantitation for each aromatic component is 3 mg/kg (mass-ppm). References ASTM Practice D4307, “Preparation of Liquid Blends for Use as A
6、nalytical Standards,” www.astm.org Scanlon, J. T. and Willis, D. E., Journal of Chromatographic Science, 23, 333-340 (1985) UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline of Method The sample is injected into a gas chromatograph that is equipped with a fused silica capil
7、lary column internally coated with poly(ethylene glycol), an autosampler, and a flame ionization detector. The concentrations of individual or group impurities are determined by the external standard method of quantitation, wherein peak areas of the sample components are compared to the peak areas o
8、f a calibration blend analyzed under identical conditions and injection volumes. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Balance, readability 0.0001 g Chromatographic column, 60 m of 0.32-mm ID fused silica
9、capillary, internally coated to a film thickness of 0.50 m with cross-linked poly(ethylene glycol), Restek, Cat. No. 10642 Gas chromatograph, temperature programmable, built for capillary column chromatography, utilizing a split injection system having a glass injection port insert and equipped with
10、 a flame ionization detector that will give a minimum peak height response of 10 times the background noise for 3 mg/kg of toluene when operated at the recommended conditions, Agilent Technologies, Model 7890 2 of 8 919-11 Data system, electronic, for obtaining peak areas. This device must integrate
11、 areas at a sufficiently fast rate so that narrow peaks typically resulting from use of a capillary column can be accurately measured. Agilent Technologies, ChemStation. Leak detector, gas, Grace Davison, Cat. No. 60229 Refrigerator, flammable-materials storage, Fisher Scientific, Cat. No. 97-938-1
12、Regulator, air, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-590 Regulator, hydrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-350 Regulator, nitrogen, two-stage, high purity, delivery
13、 pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Sample injector, any syringe or injector capable of injecting a repeatable 1.0-L volume of sample. The use of an automatic injection device is required to achieve necessary repeatable injection volumes. Agilent Technologies, Mo
14、del 7693. Vortex mixer, VWR, Cat. No. 58816-121 Reagents and Materials References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Air, zero gas, total hydrocarbons less than 2.0 ppm as methane Benzene, 99.9% minimum purity, Sigma-Aldric
15、h, Cat. No. 270709 Cumene, 98.0% minimum purity, Sigma-Aldrich, Cat. No. 28230 Ethylbenzene, 99.8% minimum purity, Sigma-Aldrich, Cat. No. 296848 Gas purifier, for hydrogen, to remove oxygen and moisture from carrier gas, VICI Mat/Sen, Cat. No. P200-1, and (optional) indicating oxygen trap, Restek,
16、Cat. No. 22010, see Procedure, Chromatographic Technique, Step 1 Hydrogen, zero gas, 99.99% minimum purity, total hydrocarbons less than 0.5 ppm as methane Nitrogen, zero gas, 99.99% minimum purity, total hydrocarbons less than 0.5 ppm as methane Pipets, disposable, Pasteur, VWR, Cat. No. 14673-043
17、Pipet bulbs, VWR, Cat. No. 15001-362 Propylene glycol (1,2-propanediol), 99.5% minimum purity, Sigma-Aldrich, Cat. No. 398039 Syringe, replacement, for recommended sample injector, 10-L, Agilent Technologies, Cat. No. 5181-1267 Tetraethylene glycol, 99% minimum purity, Sigma-Aldrich, Cat. No. 110175
18、 Toluene, 99.9% minimum purity, Sigma-Aldrich, Cat. No. 650579 Vials, autosampler, for recommended sample injector, with caps, Agilent Technologies, Cat. No. 5182-0864 3 of 8 919-11 Procedure The analyst is expected to be familiar with general laboratory practices, the technique of gas chromatograph
19、y, and the equipment being used. Chromatographic Technique 1. Install the gas purifier in the supply line between the carrier gas source and the carrier gas inlets on the gas chromatograph. Column life is significantly reduced if the gas purifier is not used. Replace the gas purifier at intervals de
20、termined by good laboratory practice. An indicating oxygen trap may be placed downstream of the gas purifier. When the indicator shows one-half used, replace both the gas purifier and the indicating trap. 2. Install the fused silica capillary column in the gas chromatograph according to the column a
21、nd gas chromatograph manufacturers instructions. CAUTION: Hydrogen leakage into the confined volume of the column oven can cause a violent explosion. Therefore, it is mandatory to check for leaks each time a connection is made and periodically thereafter. 3. Establish the recommended operating condi
22、tions as given in Table 1. Different conditions may be used provided they produce the required sensitivity and chromatographic separations equivalent to those shown in the Typical Chromatogram (see Figures 1 and 2). Table 1 Recommended Operating Conditions Carrier gas hydrogen Mode constant flow Flo
23、w rate 2.5 mL/min Head pressure 50C 74 kPa gauge (10.7 psig) Linear velocity 50C 41 cm/sec Split flow 250 mL/min Injection port Temperature 280C Autoinjector viscosity delay 5 sec Column temperature program Initial temperature 50C Initial hold time 0 min Programming rate 10C/min Final hold temperatu
24、re 250C Final Time 20 min Detector flame ionization Detector temperature 280C Hydrogen flow ratea 28 mL/min Air flow ratea 400 mL/min Makeup gas nitrogen Makeup gas flow ratea 30 mL/min Sample size 1.0 L, repeatable aConsult the manufacturers instrument manual for suggested flow rates. 4. Program th
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9192011TRACEHYDROCARBONSINUDEXSOLVENTBYGCPDF

链接地址:http://www.mydoc123.com/p-1074568.html