UOP 915-1992 NORMAL PARAFFINS BY CAPILLARY GAS CHROMATOGRAPHY.pdf
《UOP 915-1992 NORMAL PARAFFINS BY CAPILLARY GAS CHROMATOGRAPHY.pdf》由会员分享,可在线阅读,更多相关《UOP 915-1992 NORMAL PARAFFINS BY CAPILLARY GAS CHROMATOGRAPHY.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、NORMAL PARAFFINSBY CAPILLARY GAS CHROMATOGRAPHYUOP Method 915-92SCOPEThis method is for determining total normal paraffins (TNP) and individual normal paraffins inhydrocarbons, such as the UOP Molexprocess streams, including samples containing the Molex processdesorbent. Samples are typically in the
2、 C8to C20carbon number range. By preheating the sample andadding a suitable solvent, the range can be extended up to C40. The mass-% TNP at the high and lowconcentrations tend to be biased slightly high because of incomplete compensation for non-normals at thenormal paraffin sites. A slightly more a
3、ccurate TNP value would be expected using UOP Method 411 (seeNOTE). The lower limit of detection is 0.01 mass-% for any compound.OUTLINE OF METHODThe sample to be analyzed is injected into a gas chromatograph that is equipped with a fused silicacapillary column, internally coated with DB-17 (bonded)
4、 methylphenyl-polysiloxane, and with a flameionization detector (FID). To determine the total normal paraffins at all concentrations requires analysis bytwo different procedures, designated Method A and Method B.In Method A, samples containing less than 75 mass-% TNP are injected using the appropria
5、te operatingconditions. The integrator events are programmed to force a baseline at the valley point before and aftereach normal paraffin to compensate for the underlying non-normals at the normal paraffin site (Figs. 1 and2). The mass-% of each normal paraffin is obtained by the internal standard m
6、ethod of quantitation, whereinpeak areas are compared to the peak area of a known amount of an internal standard. The individual normalparaffins are summed to obtain TNP.In Method B, samples containing 75 mass-% or greater concentrations of total normal paraffins areinjected using the appropriate op
7、erating conditions. The mass-% composition of the sample is obtained bythe internal normalization technique. The individual normal paraffins are summed to obtain TNP.APPARATUSReferences to catalog numbers and suppliers are included as a convenience to the method user. Othersuppliers may be used.Bala
8、nce, readability 0.1-mgIT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TODETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH ANDSAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THISPROCEDURE IN
9、THE MANNER PRESCRIBED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS(MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED INTHIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTIONEQUIPMENT (PPE). COPYRIGHT 1992 UOP LLCALL RIGHTS
10、 RESERVEDUOP Methods are available through ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken PA 19428-2959,United States. The Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service atserviceastm.org, 610.832.9555 FAX, or 610.832.9585
11、 PHONE.2 of 10915-92Chromatographic column, 30 m of 0.32-mm ID fused silica capillary, internally coated to a filmthickness of 0.25 m with DB-17 (bonded) 50% methyl-50% phenyl-polysiloxane, J therefore, it is mandatory to check for leaks each time a connection is made and periodicallythereafter.3. E
12、stablish the operating conditions listed in Table 1; except, set the column oven to 260C until thebaseline is stable. Other conditions may be used provided they produce the required sensitivity and chromatographicseparations equivalent to those shown in the Typical Chromatograms. If the carbon numbe
13、r range is significantly different from that shown in the Typical Chromatograms, thestarting temperature, final temperature or programming rate (Table 1) can be adjusted to optimizecomponent elution.4. Cool the column oven to a stabilized 32C.Method A: Normal Paraffins by Internal Standard Less Than
14、 75 Mass-% TNP1. Inject 0.5 L of the sample to be analyzed into the gas chromatograph and start the recorder,integrator and column temperature programming sequence. Typical chromatograms are shown inFigs. 1 and 2. Better precision is obtained when an autoinjector is used. This experimental run deter
15、mines the carbon number distribution of the sample, the internal standard to beused, and the approximate TNP composition. For samples containing equal to or greater than 75 mass-% TNP, use Method B.2. Choose an internal standard having one less or one more carbon number than the lowest or highestcon
16、tained in the sample.3. Add approximately 0.5 g of the internal standard, weighed to the nearest 0.1 mg, to approximately 5 gof the weighed liquefied sample. Samples containing solids at ambient temperature are heated in an oven at 85C until the entire sample isliquefied. A solvent such as n-hexane
17、can be added to the weighed sampleinternal standard solution, if necessary,to maintain a liquefied state at ambient temperature.4. Set the integrator events to force a baseline at the valley points before and after each normal paraffin(see Figs. 1 and 3). The baseline must be set in this way to comp
18、ensate for the underlying non-normals at the n-paraffin sites.5. Inject 0.5 L of the sample with the internal standard added and start the recorder, integrator andcolumn temperature programming sequence.4 of 10915-92 If the sample has been diluted, inject a proportionately larger sample volume.6. Id
19、entify the normal paraffins from the resultant chromatogram by comparing the chromatogramobtained to the Typical Chromatograms (Figs. 1 or 2), or to a known blend of normal paraffinsanalyzed under identical conditions.Method B: Normal Paraffins by Normalized Composition Equal To or Greater Than 75 M
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9151992NORMALPARAFFINSBYCAPILLARYGASCHROMATOGRAPHYPDF

链接地址:http://www.mydoc123.com/p-1074564.html