UOP 912-2006 Fluoride in Catalysts Molecular Sieves and Aqueous Solutions by Ion Selective Electrode.pdf
《UOP 912-2006 Fluoride in Catalysts Molecular Sieves and Aqueous Solutions by Ion Selective Electrode.pdf》由会员分享,可在线阅读,更多相关《UOP 912-2006 Fluoride in Catalysts Molecular Sieves and Aqueous Solutions by Ion Selective Electrode.pdf(12页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1991, 1994, 2006 UOP LLC. All rights r
3、eserved. Nonconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or
4、610.832.9585 PHONE. Fluoride in Catalysts, Molecular Sieves, and Aqueous Solutions by Ion Selective Electrode UOP Method 912-06 Scope This method is for determining total fluoride in aluminosilicate catalysts and molecular sieves, and inorganic fluoride in aqueous samples. The method is particularly
5、 useful for catalysts containing zirconia or other metals that interfere with alternative procedures. The range of fluoride typically determined is from 0.01 to approximately 20 mass-% for catalysts or molecular sieves, and 0.01 to approximately 50 mass-% for aqueous samples. References UOP Method 9
6、18, “Collection and Grinding of Adsorbents and Catalysts for Analysis,” www.astm.org UOP Method 954, “Loss on Ignition (LOI) for Fresh, Regenerated, Used, and Spent Catalysts, Catalyst Supports, and Adsorbents,” www.astm.org UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline
7、 of Method Fluoride is quantitatively separated from a solid sample by fusion with an accelerant at 1000C in a steam atmosphere. The fluoride is converted to hydrofluoric acid (HF) and is collected with the steam condensate. The condensate flows into a polypropylene beaker containing a sodium hydrox
8、ide solution. The sodium hydroxide in the solution prevents volatilization of the HF by converting it to sodium fluoride. The condensate solution is then diluted to a specific volume with water. The solution is mixed with an ionic strength activity buffer, and the fluoride concentration is measured
9、potentiometrically by comparison to standard fluoride solutions using an ion-selective electrode. Inorganic fluoride is determined in aqueous samples by dilution to a specific volume with water. The solution is mixed with an ionic strength activity buffer and the fluoride concentration is determined
10、 potentiometrically as described above. 2 of 12 912-06 Apparatus References to catalog numbers are included as a convenience to the method user. Other suppliers may be used. Aluminum press and die, for forming nickel boats and boat covers to fit in the heavy wall nickel boat, fabricate locally Balan
11、ce, readability 0.1-mg Beaker, graduated, 150-mL, polypropylene, Fisher Scientific, Cat. No. 02-591-10C Boat puller, LECO Corporation, Cat. No. 501-062 Clamps, tubing, Fisher Scientific, Cat. No. 14-198-5A, or cable ties, to secure tubing to fittings Connector, polypropylene, T-shaped, 7.9-mm OD, Fi
12、sher Scientific, Cat. No. 15-319D Crucible, porcelain, high form, 10-mL, Fisher Scientific, Cat. No. 07-965C Electrode, combination fluoride, Orion, Model 96-09, Fisher Scientific, Cat. No. 13-641-904 Flasks, volumetric, polypropylene, Class B, 100-, 250-, 500- and 1000-mL, Fisher Scientific, Cat. N
13、os. 10-198-50B, D, E and F, respectively Flask, round bottom, borosilicate glass with three vertical necks and 24/40 joints, 3-liter, Fisher Scientific, Cat. No. 10-164D Furnace, split-tube, capable of maintaining 1100C, Fisher Scientific, Cat. No. 10-559-15 Heating mantle, with controller, for 3-li
14、ter flask, Fisher Scientific, Cat. No. 11-474-105G Jack, laboratory, 200- x 200-mm, Fisher Scientific, Cat. No. 14-673-52, two required Magnetic stir bar, Fisher Scientific, Cat. No. 14-512-126 Magnetic stirrer, Fisher Scientific, Cat. No. 14-493-120S Meter, benchtop, pH/mV/ISE, Orion, Model EA 940,
15、 Fisher Scientific, Cat. No. 13-641-741 Nickel boats, heavy wall, Fisher Scientific, Cat. No. 07-645 Nickel ribbon, nickel No. 201 ASTM B 162, 0.002- x 1.00-inch, Allegheny Ludlum. Fabricate disposable boats and lids from nickel ribbon using the aluminum press and die. Pinchcock, Day, Fisher Scienti
16、fic, Cat. No. 05-867 Pipet filler, Fisher Scientific, Cat. No. 03-692-35 Pipets, volumetric, Class A, 1-, 5-, 10-, 25- and 50-mL, Fisher Scientific, Cat. Nos. 13-660A, E, F, J and K, respectively Reaction chamber-condenser assembly (nickel pyrohydrolysis tube), handle assembly, and sample port closu
17、re, (see Figures 1, 2, and 4), fabricate locally Rubber stopper, black, solid, size No. 4, Fisher Scientific, Cat. No. 14-130G, two required Rubber stopper, black, two-hole, size No. 4, Fisher Scientific, Cat. No. 14-140M Spatula, micro, Fisher Scientific, Cat. No. 21-401-15 Support ring, cork, 140-
18、mm OD, Fisher Scientific, Cat. No. 07-835B 3 of 12 912-06 Tubing, borosilicate glass, standard wall, 6-mm OD, Fisher Scientific, Cat. No. 11-362D Tubing, clear, silicone rubber, 6.4-mm ID, Fisher Scientific, Cat. No. 14-169-3C, for steam lines Tubing, laboratory, Tygon R-3603, 6.4-mm ID, Fisher Scie
19、ntific, Cat. No. 14-169-3C, for water lines Tubing clamp, flatjaw pinchcock, Fisher Scientific, Cat. No. 05-850A Reagents and Materials All reagents shall conform to the specifications established by the Committee on Analytical Reagents of the American Chemical Society, when such specifications are
20、available, unless otherwise specified. References to water mean deionized or distilled water. References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Accelerant, Com-Cat Accelerator, LECO Corporation, Cat. No. 502-321. Vanadium pento
21、xide, 99.99% purity, Aldrich Chemical, Cat. No. 20,485-4, may be substituted and provides equivalent performance. Care must be taken when handling vanadium pentoxide including the wearing of gloves because it is toxic and an irritant. Electrode filling solution, fluoride combination electrode, Fishe
22、r Scientific, Cat. No. 13-641-899 Fluoride standard, 100 g/mL, AccuStandard, Cat. No. IC-F-1X-5 Methyl purple indicator, Aldrich Chemical, Cat. No. 27,733-9 Phosphate rock, reference standard, approximately 3.82% fluorine, NIST, Cat. No. SRM 120c Pipet dropper bulbs, Fisher Scientific, Cat. No. 03-4
23、48-26 Pipets, transfer, disposable, Fisher Scientific, Cat. No. 13-711-7 Sodium hydroxide, 0.1-M, Aldrich Chemical, Cat. No. 319481 TISAB II (total ionic strength activity buffer), fluoride electrode activity standard, Fisher Scientific, Cat. No. 13-641-873 Toothpaste, non-gel type containing fluori
24、de Water, deionized or distilled Water, tap, for cooling Procedure The analyst is expected to be familiar with general laboratory practices, the technique of titration, and with the equipment being used. Preparation of Fluoride Standards Fluoride Standard, Approximately 100-g/mL F This is the reagen
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9122006FLUORIDEINCATALYSTSMOLECULARSIEVESANDAQUEOUSSOLUTIONSBYIONSELECTIVEELECTRODEPDF

链接地址:http://www.mydoc123.com/p-1074563.html