UOP 910-2007 Total Chloride in LPG and Gaseous Hydrocarbons by Microcoulometry.pdf
《UOP 910-2007 Total Chloride in LPG and Gaseous Hydrocarbons by Microcoulometry.pdf》由会员分享,可在线阅读,更多相关《UOP 910-2007 Total Chloride in LPG and Gaseous Hydrocarbons by Microcoulometry.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1992, 1997, 2007 UOP LLC. All rights r
3、eserved. Nonconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or
4、610.832.9585 PHONE. Total Chloride in LPG and Gaseous Hydrocarbons by Microcoulometry UOP Method 910-07 Scope This method is for determining total chloride in gaseous hydrocarbons or liquefied petroleum gas (LPG) at concentrations ranging from approximately 1 to 1000 ng/mL for gas or mass-ppm (g/g)
5、for LPG. Except for fluoride, other halogens present are calculated as chloride. Chloride cannot be determined quantitatively if sulfur is present at concentrations greater than approximately 1 mass-%, see Note 1. References Perrys Chemical Engineering Handbook UOP Method 516, “Sampling of Gasolines
6、, Distillate Fuels and C3- C4Fractions”, www.astm.org Outline of Method A commercial microcoulometer is set up according to the manufacturers specifications. Gaseous and LPG samples are collected in stainless steel cylinders. Measured volumes of the LPG or gaseous samples are analyzed by injecting t
7、hem into a heated combustion tube containing oxygen blended with argon. The organic halides are converted quantitatively to hydrogen halide, hydrogen oxyhalide, carbon dioxide and water. The gaseous combustion products are passed through a dryer and then pass into a coulometric cell containing a buf
8、fer, in which the halide ions are dissolved. The instrument automatically titrates the halide ions to a coulometric endpoint and calculates the number of equivalents of halide in the sample. The halide is reported as the equivalent number of chloride ions, given in nanograms. The concentration of ch
9、loride in the sample is then calculated in g/g for LPG or in ng/mL for gases. Any HCl content is non-quantitative because of losses to the vessel walls, especially in the presence of water, and special care must be exercised to determine the HCl content quantitatively (see Note 2). The other halogen
10、s that respond are bromine and iodine. Fluorine does not react with silver so it is not detected. In the case of bromine and iodine, only the HX form titrates with silver, while for chlorine, both HCl and HOCl react with silver. All chlorine is determined, but only approximately 50% of the bromine a
11、nd iodine is detected. In the chlorine mode, the apparent response to bromine and iodine is 2 of 9 910-07 further reduced to 25 and 15% of the chloride response, respectively, due to the differences in their atomic weights. Apparatus References to catalog numbers and suppliers are included as a conv
12、enience to the method user. Other suppliers may be used. Additional items may be required depending on the specific instrument used. Bottle, Teflon, 1000-mL, Fisher Scientific, Cat. No. 02-923-30D Cylinder, graduated, 500-mL, borosilicate glass, Fisher Scientific, Cat. No. 08-552-4E Flask, Erlenmeye
13、r, 1000-mL, borosilicate glass, Fisher Scientific, Cat. No. 10-041C Flowmeter, digital, soap film, range 0.50- to 500-mL/min, Alltech Associates, Cat. No. 4068 Gauge, pressure, 0 to 2760 kPa gauge (0 to 400 psig), Matheson Tri-Gas, Cat. No. 63-2242 Microcoulometer, with attached furnace, cell, gas/L
14、PG module, controls and computer; a printer is optional, but recommended, Mitsubishi TCL-100 (Cosa Instruments), Thermo ECS-3000, or Analytik Jena Multi EA 3100. A Dohrmann MCTS-120, no longer manufactured, may also be used. Follow the manufacturers recommendations for maintaining a supply of spare
15、parts and consumables. Regulator, argon, two-stage, high purity, with stainless steel diaphragm, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Cat. No. 3122-580 Regulator, nitrogen, two-stage, high purity, with stainless steel diaphragm, delivery pressure range 35-2400 kPa (5-350
16、 psi), Matheson Tri-Gas, Cat. No. 3128-580, only needed for analysis of LPG Regulator, oxygen, two-stage, high purity, with stainless steel diaphragm, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Cat. No. 3122-540 Reagents and Materials References to catalog numbers and supplier
17、s are included as a convenience to the method user. Other suppliers may be used. References to water mean deionized or distilled water. The following items are required to perform the analysis. Additional reagents and materials may be required depending on the specific microcoulometer utilized. Wher
18、e different reagents are specified in the instrument manual, follow the manufacturer recommendations. Argon, compressed gas, 99.99% minimum purity 2-Chlorobutane in isobutane, (LPG and Gas Reference Standard, see Note 3). Liquefied Petroleum Gas and Gas Reference Standards are available by special o
19、rder from most gas suppliers. Request preparation on a mass/mass basis at a concentration similar to that expected for the sample of LPG or on a nominal volume/volume basis at a concentration similar to that expected for the sample of gas. In addition, request that the LPG cylinder be supplied with
20、a full length dip tube to deliver the liquefied gas from the bottom of the cylinder and under 250 psig (1750 kPa) of nitrogen pressure. This will keep the LPG liquefied for instrumental loop-sampling, free of bubbles. The material is certified by the supplier to be stable for at least six months. Ma
21、theson Tri-Gas or local supply. Inorganic chloride standard (NaCl or HCl Solution), prepared to meet manufacturers recommendations, if needed Isobutane, (LPG and Gas Blank). Liquefied Petroleum Gas and Gas Reference Blanks are available on special order from a gas supplier and should be ordered when
22、 the Reference Standards are 3 of 9 910-07 ordered. Request that the LPG cylinder be supplied with a full length dip tube to deliver the liquefied gas from the bottom of the cylinder and under 250 psig (1750 kPa) of nitrogen pressure. This will keep the LPG liquefied for instrumental loop-sampling,
23、free of bubbles. Matheson Tri-Gas or local supply. Leak detector, liquid solution, Matheson Tri-Gas, Cat. No. BUBL-01 Nitrogen, compressed gas, 99.98% minimum purity, only needed for analysis of LPG Oxygen, compressed gas, 99.98% minimum purity Quartz combustion tube, replacement, as supplied for th
24、e instrument used Sulfuric acid, concentrated, Fisher Scientific, Cat. No. A300-212 Sulfuric acid, 90%, for drying combustion exit gases. Pour 50 mL of water into a 1000-mL Erlenmeyer flask cooled in an ice bath. Slowly and carefully add, from a 500-mL graduated cylinder, 450 mL of concentrated sulf
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9102007TOTALCHLORIDEINLPGANDGASEOUSHYDROCARBONSBYMICROCOULOMETRYPDF

链接地址:http://www.mydoc123.com/p-1074562.html