UOP 851-2008 Density of Powders and Solids by Helium Displacement.pdf
《UOP 851-2008 Density of Powders and Solids by Helium Displacement.pdf》由会员分享,可在线阅读,更多相关《UOP 851-2008 Density of Powders and Solids by Helium Displacement.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1984, 2008 UOP LLC. All rights reserve
3、d. Nonconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.83
4、2.9585 PHONE. Density of Powders and Solids by Helium Displacement UOP Method 851-08 Scope This method is for determining true density of powders and solids by displacement of helium gas. This method is written for use with a Micromeritics AccuPyc 1330 Pycnometer; however, equivalent equipment may b
5、e used. References Owners Manual and Quick Reference Guide, AccuPyc 1330 Pycnometer, Micromeritics Corporation UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline of Method The pycnometer works by measuring the pressure drop of a gas as it is expanded into a known volume. The
6、 observed pressure change is used to compute the volume of the sample. The measured sample volume excludes all open pore structure as the gas molecules will rapidly fill the pores of the sample; only the truly solid phase of the sample displaces the gas. If pretreatment of the sample such as calcina
7、tion or outgassing is required, it must be carried out prior to analysis. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Balance, readability, 0.1-mg Helium Pycnometer, Micromeritics, AccuPyc 1330 Regulator, helium
8、, two-stage, high purity, delivery pressure range 15-200 kPa (2-30 psi), Matheson Tri-Gas, Model 3121-580 Reagents and Materials Cloth, lint-free Helium, gas, cylinder, 99.995% minimum purity 2 of 4 851-08 Procedure The analyst is expected to be familiar with general laboratory procedures, the techn
9、ique of pycnometer measurement, and the equipment being used. Set up the Micromeritics AccuPyc 1330 Helium Pycnometer as outlined in the manufacturers instruction manual. Calibration volumes are included with the instrument. Be certain that gauges are calibrated and properly zeroed before proceeding
10、. Specific Instructions for Micromeritics AccuPyc 1330 Helium Pycnometer The following detailed procedure for using the Micromeritics AccuPyc 1330 Pycnometer describes the technique for making measurements and is used for the empty cell as well as the cell containing the sample. If a different instr
11、ument is used, follow instructions supplied with that instrument. 1. Ensure that the instrument is turned on and that the helium is flowing to it. 2. Remove the cell chamber cap. Do not remove the cell chamber cap when the pycnometer is pressurized. 3. Remove the sample cup from the chamber and obta
12、in a tare weight. 4. Fill sample cup approximately full with sample and reweigh. See Note 1. 5. Subtract the tare weight from the filled sample cup and record the sample weight to the nearest 0.1 mg. 6. Insert the sample cup into the cell chamber. 7. Replace the cell chamber cap. 8. Press ANALYZE. F
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP8512008DENSITYOFPOWDERSANDSOLIDSBYHELIUMDISPLACEMENTPDF

链接地址:http://www.mydoc123.com/p-1074554.html