UOP 801-2014 Size Distribution of Charcoal Particles.pdf
《UOP 801-2014 Size Distribution of Charcoal Particles.pdf》由会员分享,可在线阅读,更多相关《UOP 801-2014 Size Distribution of Charcoal Particles.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. SAFETY DATA SHEETS (SDS) OR EXPERIMENTAL SAFETY DATA SHEETS (ESDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1980, 2014 UOP LLC. All rights reserved. Nonconfidential U
3、OP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE. Size D
4、istribution of Charcoal Particles UOP Method 801-14 Scope This method is for determining the size distribution of charcoal and charcoal supported catalyst particles by screening through appropriate ASTM Series testing sieves. Reference UOP Method 999, “Precision Statements in UOP Methods,” www.astm.
5、org Outline of Method A representative 100-mL sample is placed on the top of a nest of appropriate mesh size, 8-inch diameter sieves with a bottom pan. The nest is covered, placed in a sieve shaker and shaken for 5 minutes. The material retained on each sieve and in the pan is weighed, and the weigh
6、t-percent of is then calculated. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Balance, readability 0.01 g, with digital interface. See the shaker manufacturers software or representative for balance compatibility
7、. Brush, paint, VWR, Cat. No. 17230-060, used to clean the sieves Computer, compatible with shaker interface and software Riffler, sample splitter, VWR, Cat. No. 56723-250 Sieve bottom pan and cover, stainless steel, 203-mm (8-inch) diameter, 51-mm (2-inch) height, Retsch Part Nos. 697203050 and 326
8、620002, respectively, VWR, Cat. Nos. 66004-848 and 14214-620, respectively Sieve shaker, with clamp cover, Retsch Model AS 200 Control with Easy Clamp, VWR Cat. No. 66004-956. Easy Sieve software is included. This sieve shaker employs a specific throwing motion with angular momentum; different resul
9、ts may be obtained if other sieve shakers are used. Substitution is permitted only if equivalent results are demonstrated. 2 of 4 801-14 Sieves, ASTM Series, stainless steel wire cloth, stainless steel frame, 203-mm (8-inch) diameter, 51-mm (2-inch) height. Sieve sizes and nest designations are dete
10、rmined by the size of the material to be tested and any specifications related to the material. Table 1 is a list of suggested sieves to be used. Sieves may be added or deleted depending upon the materials to be tested and test requirements. Important: If the results will be compared between laborat
11、ories, the sieves must be “Matched” test sieves from the same manufacturer. Table 1 Part Numbers for Sieves ASTM Series Size Opening Retsch Part Number VWR Catalog Number No. 10 2.00 mm 60150002000 66004-896 No. 20 0.85 mm 60150000850 66004-886 No. 30 0.60 mm 60150000600 66004-882 Reagents and Mater
12、ials References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Gloves, disposable, nitrile, VWR, Cat. Nos. 32919-000 through -008, corresponding to hand sizes XS through XL Procedure The analyst is expected to be familiar with general
13、laboratory practices and the equipment being used. Wear gloves whenever incidental contact with sample is possible. Dispose of all supplies and samples in an environmentally safe manner according to local regulations. The steps below are written specifically for the Retsch Model AS 200 Control shake
14、r listed in Apparatus. 1. Using approximately 500 g of sample, riffle to obtain a 50 to 75 g representative sample of the material to be tested. Follow instructions supplied with the riffler for proper riffling procedure. 2. Open the Easy Sieve program. Click on “Sieve Analysis” in the menu bar. Cho
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP8012014SIZEDISTRIBUTIONOFCHARCOALPARTICLESPDF

链接地址:http://www.mydoc123.com/p-1074546.html