1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1974, 1977, 1986, 1998, 2004, 2006 UOP
3、 LLC. All rights reserved. Nonconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.8
4、32.9555 FAX, or 610.832.9585 PHONE. Aromatics in Hydrocarbons by GC UOP Method 744-06 Scope This method is for determining individual C6through C10aromatic compounds in petroleum distillates or aromatic concentrates having a final boiling point of 210C or lower. C11and heavier aromatics are reported
5、 as a group. C10and heavier non-aromatics may interfere with the determination of benzene and toluene. When this occurs, benzene and toluene can be determined by ASTM Methods D 5443, D 5580, D 6729, or D 6839, or UOP Methods 690, or 870. Other applications for this method include the assay of any C1
6、0or lower boiling aromatics, such as benzene, toluene, mixed xylenes, etc. This method may also be used to provide a distribution of C8aromatics and/or C9and heavier aromatics to a value determined by a different method. When used for any of the above applications, the limit of quantitation for a si
7、ngle aromatic component is 0.01 mass-%. References ASTM Method D 4307, “Preparation of Liquid Blends for Use as Analytical Standards,” www.astm.org ASTM Method D 5443, “Paraffin, Naphthene, and Aromatic Hydrocarbon Type Analysis in Petroleum Distillates Through 200C by Multidimensional Gas Chromatog
8、raphy,” www.astm.org ASTM Method D 5580, “Determination of Benzene, Toluene, Ethylbenzene, p/m-Xylene, o-Xylene, C9and Heavier Aromatics, and Total Aromatics by Gas Chromatography,” www.astm.org ASTM Method D 6729, “Determination of Individual Components in Spark Ignition Engine Fuels by 100 Meter C
9、apillary High Resolution Gas Chromatography,” www.astm.org ASTM Method D 6839, “Hydrocarbon Types, Oxygenated Compounds, and Benzene in Spark Ignition Engine Fuels by Gas Chromatography,” www.astm.org Scanlon, J. T. and Willis, D. E., Journal of Chromatographic Science, 23, 333-340 (1985) UOP Method
10、 690, “Octanes and Lower Boiling Hydrocarbons in Olefin-Free Gasolines by GC,” www.astm.org UOP Method 870, “Carbon Number Distribution of Paraffins, Naphthenes, and Aromatics by GC,” www.astm.org Copyright by ASTM Intl (all rights reserved);Reproduction authorized per License Agreement with Monique
11、 Tyree (ASTMIHS Account); Tue Aug 8 10:43:31 EDT 20062 of 11 744-06 UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org Outline of Method The sample is injected into a gas chromatograph that is equipped with a flame ionization detector (FID) and a fused silica capillary column intern
12、ally coated with cross-linked polyethylene glycol. The mass-% composition of the sample is obtained by the normalization of peak areas after relative response factors are applied to correct for detector mass response differences. Alternatively, selected aromatic components, e.g., C8aromatics, may be
13、 normalized to their sum as determined by another technique. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Balance, readability 0.1-mg Chromatographic column, 60 m of 0.32-mm ID fused silica capillary, internally
14、coated to a film thickness of 0.5-m with cross-linked polyethylene glycol, Restek, Cat. No. 10642 Gas chromatograph, capable of multiple temperature ramping, built for capillary column chromatography utilizing a split injection system, having a glass injection port insert, and equipped with a flame
15、ionization detector that will give a minimum peak height response of 5 times the background noise for 0.01 mass-% of benzene when operated at the recommended conditions. Electronic pressure control (EPC) is recommended. Agilent Technologies, Model 6890 Data system, or electronic integrator, for obta
16、ining peak areas. This device must integrate areas at a sufficiently fast rate so that narrow peaks typically resulting from use of a capillary column can be accurately measured. Agilent Technologies, ChemStation Leak detector, gas, Alltech Associates, Cat. No. 21-250 Regulator, air, two-stage, high
17、 purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-590 Regulator, hydrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-350 Regulator, nitrogen or helium, two-stage, high purity, delivery pressure range 30-70
18、0 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Sample injector, any syringe or injector capable of injecting a 0.5-L volume of sample. An automatic injection device is recommended. Agilent Technologies, Model 7683 Reagents and Materials References to catalog numbers and suppliers are included a
19、s a convenience to the method user. Other suppliers may be used. Air, total hydrocarbons less than 2 ppm as methane Benzene, 99.9%, Aldrich Chemical, Cat. No. 27,070-9 Gas purifier, for hydrogen, to remove oxygen and moisture from carrier gas, VICI Mat/Sen, Cat. No. P200-1 Copyright by ASTM Intl (al
20、l rights reserved);Reproduction authorized per License Agreement with Monique Tyree (ASTMIHS Account); Tue Aug 8 10:43:31 EDT 20063 of 11 744-06 n-Heptane, 99+%, Aldrich Chemical, Cat. No. 27,051-2 Hydrogen, zero-gas, 99.95% minimum purity, total hydrocarbons less than 0.5 ppm as methane 1-Methyl-2-
21、ethylbenzene, 99%, Chemsampco, Cat. No. 2900.00 Methylindane isomers, if needed for peak identification, AccuStandard, special order Nitrogen or helium, zero-gas, 99.99% minimum purity, total hydrocarbons less than 0.5 ppm as methane, used for detector makeup gas Syringe, replacement, for recommende
22、d sample injector, 5-L, Agilent Technologies, Cat. No. 5181-1273 1,2,3,4-Tetramethylbenzene, 99%, Chemsampco, Cat. No. 8490.00 Toluene, 99.8%, Aldrich Chemical, Cat. No. 27,037-7 Xylenes (C8aromatics), mixed, 98.5+%, Aldrich Chemical, Cat. No. 24,764-2 Procedure Chromatographic Technique 1. Install
23、the gas purifier in the supply line between the carrier gas source and the carrier gas inlet on the gas chromatograph. Column life is significantly reduced if the gas purifier is not used. 2. Install the fused silica capillary column in the gas chromatograph. CAUTION: Hydrogen carrier gas leakage in
24、to the confined volume of the column oven can cause a violent explosion. It is, therefore, mandatory to check for leaks each time a connection is made and periodically thereafter. 3. Establish the recommended operating conditions as given in Table 1. Other conditions may be used provided they produc
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP7442006AROMATICSINHYDROCARBONSBYGCPDF