UOP 673-2012 Linear Alkylbenzene Isomer Distribution in Detergent Alkylate by GC.pdf
《UOP 673-2012 Linear Alkylbenzene Isomer Distribution in Detergent Alkylate by GC.pdf》由会员分享,可在线阅读,更多相关《UOP 673-2012 Linear Alkylbenzene Isomer Distribution in Detergent Alkylate by GC.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 1968, 1985, 1988, 2006, 2012 UOP LLC.
3、All rights reserved. Nonconfidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9
4、555 FAX, or 610.832.9585 PHONE. Linear Alkylbenzene Isomer Distribution in Detergent Alkylate by GC UOP Method 673-12 Scope This method is for determining the isomer distribution in C15 through C20 mono-substituted linear alkylbenzenes (LABs). The 6-phenyltridecane and 7-phenyltridecane isomers are
5、not resolved from each other and are determined as a composite. The lower limit of detection for a single component is 0.1 mass-%. Components other than linear alkylbenzenes are not normally determined by this method; however, an estimate of the total LAB content of the sample may be obtained by mea
6、suring the components present in the sample other than the LABs. A calculation is included to determine the number average molecular weight of the LABs in the sample, if needed. Reference UOP Method 999, “Precision Statements in UOP Methods,“ www.astm.org Outline of Method The sample to be analyzed
7、is injected into a gas chromatograph that is equipped with a fused silica capillary column. The mass-% distribution of LABs is obtained by the internal normalization technique of quantitation wherein the peak areas of the LABs are normalized to 100%. By summing the non-LAB components, an estimate of
8、 the linearity of the detergent alkylate may be obtained. Apparatus References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. Chromatographic column, 30 m of 0.32-mm ID fused silica capillary, internally coated to a film thickness of 0
9、.25 m with DB-17 bonded (50%-phenyl)-methylpolysiloxane, Agilent Technologies, Cat. No. 123-1732 Gas chromatograph, temperature programmable, built for capillary column chromatography, utilizing a split injection system having a glass injection port insert and equipped with a flame 2 of 7 673-12 ion
10、ization detector (FID), Agilent Technologies, Model 7890A, with autosampler capable of injecting a repeatable 0.2-L volume of sample, Agilent Technologies, Model 7693A Data system, electronic, for obtaining peak areas. This device must integrate areas at a sufficiently fast rate so that narrow peaks
11、 typically resulting from the use of a capillary column can be accurately measured. Agilent Technologies, ChemStation. Leak detector, gas, Grace Davison, Cat. No. 60229 Regulator, air, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-590 Regulator,
12、 hydrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-350 Regulator, nitrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Reagents and Materials References to catalog numbers and s
13、uppliers are included as a convenience to the method user. Other suppliers may be used. Air, zero gas, total hydrocarbons less than 2.0 ppm as methane Gas purifier, for hydrogen, to remove oxygen and moisture from carrier gas, VICI Mat/Sen, Cat. No. P200-1, and (optional) indicating oxygen trap, Res
14、tek, Cat. No. 22010, see Procedure, Chromatographic Technique, Step 1 Hydrogen, zero gas, 99.99% minimum purity, total hydrocarbons less than 0.5 ppm as methane Nitrogen, zero gas, 99.99% minimum purity, total hydrocarbons less than 0.5 ppm as methane Syringe, replacement, for recommended sample inj
15、ector, 10-L, Agilent Technologies, Cat. No. 5181-1267 Vials, autosampler, for recommended sample injector, with caps, Agilent Technologies, Cat. No. 5182-0864 Procedure The analyst is expected to be familiar with general laboratory practices, the technique of gas chromatography, and the equipment be
16、ing used. Dispose of used reagents, materials, and samples in an environmentally safe manner according to local regulations. Chromatographic Technique 1. Install the gas purifier in the supply line between the carrier gas source and the carrier gas inlets on the gas chromatograph. Column life is sig
17、nificantly reduced if the gas purifier is not used. Replace the gas purifier at intervals determined by good laboratory practice. An indicating oxygen trap may be placed downstream of the gas purifier. When the indicator shows one-half used, replace both the gas purifier and the indicating trap. 2.
18、Install the fused silica capillary column in the gas chromatograph according to the column and gas chromatograph manufacturers instructions. CAUTION: Hydrogen leakage into the confined volume of the column oven can cause a violent explosion. Therefore, it is mandatory to check for leaks each time a
19、connection is made and periodically thereafter. 3 of 7 673-12 3. Establish the recommended operating conditions as given in Table 1. Different conditions may be used provided they produce the required sensitivity and chromatographic separations equivalent to those shown in the Typical Chromatogram (
20、see Figure). 4. Program the column oven to 250C (see Table 1) and maintain this temperature until a stable baseline has been obtained at the required sensitivity. Table 1 Recommended Operation Conditions Carrier gas hydrogen Mode constant flow Flow rate 1.4 mL/min flow rate Column head pressure 140C
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP6732012LINEARALKYLBENZENEISOMERDISTRIBUTIONINDETERGENTALKYLATEBYGCPDF

链接地址:http://www.mydoc123.com/p-1074519.html