UOP 649-2010 Total Oxygen in Solid Semi-solid and High Boiling Point Liquid Hydrocarbons by Pyrolysis.pdf
《UOP 649-2010 Total Oxygen in Solid Semi-solid and High Boiling Point Liquid Hydrocarbons by Pyrolysis.pdf》由会员分享,可在线阅读,更多相关《UOP 649-2010 Total Oxygen in Solid Semi-solid and High Boiling Point Liquid Hydrocarbons by Pyrolysis.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 2010 UOP LLC. All rights reserved. Non
3、confidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585
4、 PHONE. Total Oxygen in Solid, Semi-solid, and High Boiling Point Liquid Hydrocarbons by Pyrolysis UOP Method 649-10 Scope This method is for determining total oxygen in solid, semi-solid, and high boiling point liquid hydrocarbons at concentrations ranging from approximately 0.1 to 50 mass-%. Sampl
5、ing methods are described for several types of sample matrices. The method determines the total oxygen, including organic oxygen compounds, dissolved water and dissolved molecular oxygen. The contributions of oxygen from water and dissolved molecular oxygen can be measured using other methods (see R
6、eferences below) and subtracted from this result. Volatile liquid samples should be analyzed using UOP Method 730, “Total Oxygen in Liquid Hydrocarbons by Pyrolysis.” References ASTM Method D 1364, “Water in Volatile Solvents (Karl Fischer Reagent Titration Method),” www.astm.org ASTM Method D 4052,
7、 “Density and Relative Density of Liquids by Digital Density Meter,” www.astm.org UOP Method 481, “Water in Liquid Hydrocarbons by Coulometry,” www.astm.org UOP Method 678, “Dissolved Molecular Oxygen in Liquid Hydrocarbons by Electrochemical Detection,” www.astm.org UOP Method 730, “Total Oxygen in
8、 Liquid Hydrocarbons by Pyrolysis,” www.astm.org Outline of Method An oxygen analyzer (CHN/O instrument) is set up and calibrated with liquid standards or solid standards, as required for the sample matrix. Liquid samples are diluted in toluene and directly injected into the pyrolysis reactor in a h
9、elium carrier. Solid or semi-solid samples are sealed in silver foil capsules before introduction into the pyrolysis reactor. Samples are pyrolyzed at a temperature in excess of 1000C. A metalized carbon reagent helps convert all of the oxygen in the sample into carbon monoxide (CO). Acid gas is rem
10、oved from the product stream with an appropriate trap material (soda lime). The pyrolysis product gas is passed through a GC column to separate the CO. The CO content is measured using thermal conductivity detection. 2 of 9 649-10 Apparatus References to catalog numbers and suppliers are included as
11、 a convenience to the method user. Other suppliers may be used. Balance, readability 0.1-mg Capsule press, CE Elantech, Cat. No. 205-030-02 CHN/O Analyzer, Thermo Scientific, Model FLASH 2000, available from CE Elantech. Instrument is equipped with a liquid autosampler and/or a solids autosampler, d
12、epending on the types of samples analyzed. Follow the manufacturers recommendations for maintaining a supply of spare parts and consumables, such as autosampler vials. Any use of an alternative instrument must be tested to determine if it provides equivalent sensitivity and precision. Flasks, volume
13、tric, Class A, borosilicate glass, 10-, 25-, and 50-mL, Fisher Scientific, Cat. Nos. 20-812D, 10-210-8A, and -8B, respectively Hot plate, Fisher Scientific, Cat. No. 11-100-16H Microbalance, readability 1-g, Mettler-Toledo, Model MX5. The microbalance is normally purchased with the instrument and co
14、nfigured to automatically transfer the sample weights to the instrument control program. Pipet, automatic delivery with disposable plastic tips, 20L, Fisher Scientific, Cat. No. 21-377-817 Pipet, volumetric, Class A, 1-, 2-, 5- and 10-mL, Fisher Scientific, Cat. No. 13-650-2B, -2C, -2F, and -2L, res
15、pectively Pipet filler, Fisher Scientific, Cat. No. 13-681-51 Regulator, helium, two-stage, high purity, with stainless steel diaphragm, delivery pressure range 30-700 kPa (4-100 psi), Matheson, Cat. No. 3810-580 Syringe, gastight, 250-L, CE Elantech, Cat. No. 365-040-42 Reagents and Materials Refer
16、ences to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used. References to water mean deionized or distilled water. The following items are required to perform the analysis. Additional reagents and materials may be required depending on the sp
17、ecific oxygen analyzer utilized. Where different reagents are specified in the instrument manual, follow the manufacturers recommendations. BBOT (2,5-Bis(5-tert-butyl-benzoxazol-2-yl)thiophene), CE Elantech, Cat. No. 338-352-10. For calibration, nominal oxygen content 7.43 mass-%. Certificate of Ana
18、lysis sent with material. Benzoic acid, Aldrich, Cat. No. 242381-100G. For calibration, nominal oxygen content 26.2 mass-%. Certificate of Analysis sent with material. Cystine, CE Elantech, Cat. No. 338-400-18. For calibration, nominal oxygen content 26.6 mass-%. Certificate of Analysis sent with ma
19、terial. Helium, compressed gas, 99.999% minimum purity, such as Matheson Ultra High Purity. Magnesium perchlorate (anhydrone), CE Elantech, Cat. No. 338-219-00 Methionine, CE Elantech, Cat. No. 338-352-20. For calibration, nominal oxygen content 21.5 mass-%. Certificate of Analysis sent with materia
20、l. 3 of 9 649-10 Nickel coated carbon, CE Elantech, Cat. No. 338-238-00 Pipet, transfer, disposable plastic, 152-mm length, Fisher Scientific, Cat. No. 13-711-SA Quartz turnings, CE Elantech, Cat. No. 338-223-10 Quartz wool, CE Elantech, Cat. No. 338-222-50 Silver capsules for liquids, CE Elantech,
21、Cat. No. 240-051-00 Silver foil capsules, CE Elantech, Cat. No. 240-054-00 Soda lime, CE Elantech, Cat. No. 338-352-30 Toluene, anhydrous, Burdick otherwise troubleshoot the leak before proceeding. Calibration of Apparatus Part A Liquid Calibration 1. Select the appropriate calibration standards for
22、 a linear calibration over the range of interest. It is recommended to include at least 3 standards and a solvent blank (anhydrous toluene). Follow instrument manufacturers guidelines to determine maximum injectable volume. 5 of 9 649-10 The linear calibration is made from injections of the same vol
23、ume of standards of different concentrations. It is recommended that an autosampler be used with the instrument. The sample must be completely injected into the pyrolysis reactor. Small volumes and/or slow injection rates can result in a loss of sample to the walls of the pyrolysis reactor, resultin
24、g in a bias to lower results. The use of larger injection volumes shortens the useful life of the pyrolysis reactor. The injection of 20-30 L from a 250L syringe is the best compromise for meeting the analytical requirements while maximizing the reactor life. 2. Create a calibration program sequence
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP6492010TOTALOXYGENINSOLIDSEMISOLIDANDHIGHBOILINGPOINTLIQUIDHYDROCARBONSBYPYROLYSISPDF

链接地址:http://www.mydoc123.com/p-1074518.html