1、STD-SMPTE RP 15-ENGL 1997 8357LiOL 0002726 T3T M SMPTE RECOMMENDED PRACTICE Calibration of Densitometers Used for Black-and-white Photographic Density Measurements RP 15-1 997 Revision of RP 15-1988 1 Scope 1.1 The purpose of this practice is to specify the means to be employed in the calibration of
2、 densitometers utilized in the measurement of diffuse transmission densities. 1.2 This practice applies to densitometers utilized for the measurement of processed black- and-white photographic films and plates or cast colloidal carbon tablets. 2 Normative references The following standards contain p
3、rovisions which, through reference in this text, constitute provisions of this practice. At the time of publication, the editions indicated were valid. All standards are subject to revision, and parties to agreements based on this practice are encouraged to investigate the possibility of applying th
4、e most recent edition of the standards indicated below. ANSIASO 5-2-1 995, ANSI/NAPM IT2.19-1995, Photography - Density Measurements - Part 2: Geometric Conditions for Transmission Density ANSI/ISO 5-3-1 995, ANSI/NAPM IT2.18-1996, Photography - Density Measurements - Part 3: Spectral Conditions 3 T
5、ypes of densitometers 3.1 In general, only those densitometers which conform to the geometric and spectral conditions specified by ANSVIS0 5-2, ANSVNAPM IT2.19 and ANSIASO 5-3, ANSVNAPM IT2.18 are capable of giving accurate readings of American Pagel of 4 pages National Standard diffuse transmission
6、 visual density for all types of black-and-white photo- graphic materials. 3.2 If a nonconforming densitometer is to be used with a given type of photographic material, it may be calibrated from reference samples composed of the same material. In this way, any densitometer may be calibrated to read
7、American National Standard diffuse transmission visual density, on any single type of photo- graphic material to a degree of accuracy com- mensurate with the stability and reproducibility of the instrument itself. In general, a new calibration must be made to obtain accurate readings on a different
8、material when a non- conforming densitometer is used. 4 Reference specimen 4.1 A reference specimen shall be a calibrated gray scale which is stored with special care and used at intervals of three months, more or less, as a primary reference against which to control the working specimens (see 5.1).
9、 4.2 A densitometer conforming to the geometric and spectral conditions specified in ANSMSO 5-2, ANSVNAPM IT2.19 and ANSIASO 5-3, ANSVNAPM IT2.18, and measuring in American National Standard diffuse transmission visual density, shall be used to calibrate the reference specimen. (Calibrated reference
10、 specimens are sold by manufacturers of densitometers, sensi- tometers, and film.) 4.3 The reference specimen shall have a range of diffuse transmission densities from below 0.06 to 3.0 or greater. Copyright O 1997 by the SOCIETY OF MOTION PICTURE AND TELEVISION ENGINEERS 595 W. Hartsdale Ave., Whit
11、e Plains, NY 10607 (91 4) 761 -1 1 O0 Approved January 22,1997 STD-SMPTE RP 15-ENGL 1777 8357403 O002727 77b RP 15-1997 4.4 The density variation within each step or within each specified calibration area shall be 0.01 or less. (Care should be exercised in select- ing reference specimens. They shoul
12、d be free from dirt spots and abrasions. The step or cali- bration area should be large enough to accom- modate the largest aperture used for measuring the specimen and to allow for normal specimen- positioning variations. Good optical density stability is essential for reference specimens. In gener
13、al, this can be accomplished by keeping the processed specimen two months or longer under normal laboratory lighting and tempera- ture conditions before calibrating the specimen.) 4.5 A calibration chart shall accompany each reference specimen, giving the diffuse transmis- sion densities of each ste
14、p. It shall be noted on the calibration chart whether the diffuse trans- mission densities listed are American National Standard diffuse visual densities or American National Standard diffuse printing densities (ANSI/ISO 5-2, ANSI/NAPM IT2.19 and ANSVIS0 5-3, ANSVNAPM IT2.18). Each refer- ence speci
15、men and corresponding calibration chart shall be identified by a code or serial number. The chart shall also show the type of film of which the reference specimen is made. 4.6 Each step of each of three reference speci- mens shall then be measured carefully on the laboratory densitometer to be contr
16、olled. A mini- mum of 1 O step-by-step measurements should be made and each step averaged to establish the reference densities on the laboratory densi- tometers. The step-by-step measurements of one specimen shall be compared to the calibra- tion chart values for that specimen, and the deviations pl
17、otted versus the calibration chart values. (if, during the procedure, deviations in excess of the tolerances shown in 6.2 are obtained, the densitometer first could be cali- brated according to clause 6 and the procedure repeated. However, correlation curves obtained from a densitometer out of calib
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- SMPTERP151997CALIBRATIONOFDENSITOMETERSUSEDFORBLACKANDWHITEPHOTOGRAPHICDENSITYMEASUREMENT 黑白 摄影 密度 测量

链接地址:http://www.mydoc123.com/p-1046385.html