1、 IEC TR 62461 Edition 2.0 2015-01 TECHNICAL REPORT Radiation protection instrumentation Determination of uncertainty in measurement IEC TR 62461:2015-01(en) colour inside THIS PUBLICATION IS COPYRIGHT PROTECTED Copyright 2015 IEC, Geneva, Switzerland All rights reserved. Unless otherwise specified,
2、no part of this publication may be reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying and microfilm, without permission in writing from either IEC or IECs member National Committee in the country of the requester. If you have any questions about IEC
3、copyright or have an enquiry about obtaining additional rights to this publication, please contact the address below or your local IEC member National Committee for further information. IEC Central Office Tel.: +41 22 919 02 11 3, rue de Varemb Fax: +41 22 919 03 00 CH-1211 Geneva 20 infoiec.ch Swit
4、zerland www.iec.ch About the IEC The International Electrotechnical Commission (IEC) is the leading global organization that prepares and publishes International Standards for all electrical, electronic and related technologies. About IEC publications The technical content of IEC publications is kep
5、t under constant review by the IEC. Please make sure that you have the latest edition, a corrigenda or an amendment might have been published. IEC Catalogue - webstore.iec.ch/catalogue The stand-alone application for consulting the entire bibliographical information on IEC International Standards, T
6、echnical Specifications, Technical Reports and other documents. Available for PC, Mac OS, Android Tablets and iPad. IEC publications search - www.iec.ch/searchpub The advanced search enables to find IEC publications by a variety of criteria (reference number, text, technical committee,). It also giv
7、es information on projects, replaced and withdrawn publications. IEC Just Published - webstore.iec.ch/justpublished Stay up to date on all new IEC publications. Just Published details all new publications released. Available online and also once a month by email. Electropedia - www.electropedia.org
8、The worlds leading online dictionary of electronic and electrical terms containing more than 30 000 terms and definitions in English and French, with equivalent terms in 15 additional languages. Also known as the International Electrotechnical Vocabulary (IEV) online. IEC Glossary - std.iec.ch/gloss
9、ary More than 60 000 electrotechnical terminology entries in English and French extracted from the Terms and Definitions clause of IEC publications issued since 2002. Some entries have been collected from earlier publications of IEC TC 37, 77, 86 and CISPR. IEC Customer Service Centre - webstore.iec
10、.ch/csc If you wish to give us your feedback on this publication or need further assistance, please contact the Customer Service Centre: csciec.ch. IEC TR 62461 Edition 2.0 2015-01 TECHNICAL REPORT Radiation protection instrumentation Determination of uncertainty in measurement INTERNATIONAL ELECTRO
11、TECHNICAL COMMISSION ICS 13.280 ISBN 978-2-8322-2216-4 Registered trademark of the International Electrotechnical Commission Warning! Make sure that you obtained this publication from an authorized distributor. colour inside 2 IEC TR 62461:2015 IEC 2015 CONTENTS FOREWORD . 5 INTRODUCTION . 7 1 Scope
12、 8 2 Normative references 8 3 Terms and definitions 9 4 List of symbols 12 5 The GUM and the GUM S1 concept . 14 5.1 General concept of uncertainty determination . 14 5.1.1 Overview in four steps . 14 5.1.2 Summary of the analytical method for steps 3 and 4 15 5.1.3 Summary of the Monte Carlo method
13、 for steps 3 and 4 15 5.1.4 Which method to use: Analytical or Monte Carlo? 16 5.2 Example of a model function . 16 5.3 Collection of data and existing knowledge for the example 18 5.3.1 General . 18 5.3.2 Calibration factor for the example 19 5.3.3 Zero reading for the example . 20 5.3.4 Reading fo
14、r the example 21 5.3.5 Relative response or correction factor for the example 21 5.3.6 Comparison of probability density distributions for input quantities 23 5.4 Calculation of the result of a measurement and its standard uncertainty (uncertainty budget) 25 5.4.1 General . 25 5.4.2 Analytical metho
15、d 25 5.4.3 Monte Carlo method 26 5.4.4 Uncertainty budgets . 26 5.5 Statement of the measurement result and its expanded uncertainty 27 5.5.1 General . 27 5.5.2 Analytical method 28 5.5.3 Monte Carlo method 28 5.5.4 Representation of the output distribution function in a simple form (Monte Carlo met
16、hod) 31 6 Results below the decision threshold of the measuring device . 31 7 Overview of the annexes . 32 Annex A (informative) Example of an uncertainty analysis for a measurement with an electronic ambient dose equivalent rate meter according to IEC 60846-1:2009 . 33 A.1 General . 33 A.2 Model fu
17、nction 33 A.3 Calculation of the complete result of the measurement (measured value, probability density distribution, associated standard uncertainty, and the coverage interval) . 34 A.3.1 General . 34 A.3.2 Low level of consideration of measuring conditions 35 A.3.3 High level of consideration of
18、measuring conditions . 37 Annex B (informative) Example of an uncertainty analysis for a measurement with a passive integrating dosimetry system according to IEC 62387:2012 40 IEC TR 62461:2015 IEC 2015 3 B.1 General . 40 B.2 Model function 40 B.3 Calculation of the complete result of the measuremen
19、t (measured value, probability density distribution, associated standard uncertainty, and the coverage interval) . 41 B.3.1 General . 41 B.3.2 Low level of consideration of workplace conditions 41 B.3.3 High level of consideration of workplace conditions . 43 Annex C (informative) Example of an unce
20、rtainty analysis for a measurement with an electronic direct reading neutron ambient dose equivalent meter according to IEC 61005:2003 46 C.1 General . 46 C.2 Model function 46 C.3 Calculation of the complete result of the measurement (measured value, probability density distribution, associated sta
21、ndard uncertainty, and the coverage interval) . 47 C.3.1 General . 47 C.3.2 Analytical method 47 C.3.3 Monte Carlo method 48 C.3.4 Comparison of the result of the analytical and the Monte Carlo method . 49 Annex D (informative) Example of an uncertainty analysis for a calibration of radon activity m
22、onitor according to the IEC 61577 series 51 D.1 General . 51 D.2 Model function 51 D.3 Calculation of the complete result of the measurement (measured value, probability density distribution, associated standard uncertainty, and the coverage interval) . 51 Annex E (informative) Example of an uncerta
23、inty analysis for a measurement of surface emission rate with a contamination meter according to IEC 60325:2002 . 54 E.1 General . 54 E.2 Model function 54 E.3 Calculation of the complete result of the measurement (measured value, probability density distribution, associated standard uncertainty, an
24、d the coverage interval) . 54 E.3.1 General . 54 E.3.2 Effects of distance . 55 E.3.3 Contamination non-uniformity 55 E.3.4 Surface absorption 56 E.3.5 Other influence quantities 56 E.3.6 Uncertainty budget 56 Bibliography 59 Figure 1 Triangular probability density distribution of possible values n
25、for the calibration factor N 20 Figure 2 Rectangular probability density distribution of possible values g 0 for the zero reading G 021 Figure 3 Gaussian probability density distribution of possible values g for the reading G 21 Figure 4 Comparison of different probability density distributions of p
26、ossible values: rectangular (broken line), triangular (dotted line) and Gaussian (solid line) distribution . 24 Figure 5 Distribution function Q of the measured value 29 4 IEC TR 62461:2015 IEC 2015 Figure 6 Probability density distribution (PDF) of the measured value 30 Figure C.1 Results of the an
27、alytical (red dashed lines) and the Monte Carlo method (grey histogram and blue dotted and solid lines) for ) 10 ( * H . 50 Figure D.1 Result of the analytical (red dashed lines) and the Monte Carlo method (grey histogram and blue dotted lines) for K T53 Table 1 Symbols (and abbreviated terms) used
28、in the main text (excluding annexes) 12 Table 2 Standard uncertainty and method to compute the probability density distributions shown in Figure 4 24 Table 3 Example of an uncertainty budget for a measurement with an electronic dosemeter using the model function M = N K (G G 0 ) and low level of con
29、sideration of the workplace conditions, see 5.3.5.2 . 27 Table 4 Example of an uncertainty budget for a measurement with an electronic dosemeter using the model function M = N K (G G 0 ) and high level of consideration of the workplace conditions, see 5.3.5.3 . 27 Table A.1 Example of an uncertainty
30、 budget for a dose rate measurement according to IEC 60846-1:2009 with an instrument having a logarithmic scale and low level of consideration of the measuring conditions, see text for details 36 Table A.2 Example of an uncertainty budget for a dose rate measurement according to IEC 60846-1:2009 wit
31、h an instrument having a logarithmic scale and high level of consideration of the measuring conditions, see text for details 38 Table B.1 Example of an uncertainty budget for a photon dose measurement with a passive dosimetry system according to IEC 62387-1:2007 and low level of consideration of the
32、 workplace conditions, see text for details 42 Table B.2 Example of an uncertainty budget for a photon dose measurement with a passive dosimetry system according to IEC 62387-1:2007 and high level of consideration of the measuring conditions, see text for details 44 Table C.1 Example of an uncertain
33、ty budget for a neutron dose measurement according to IEC 61005:2003 using the analytical method. 48 Table C.2 Example of an uncertainty budget for a neutron dose rate measurement according to IEC 61005:2003 using the Monte Carlo method 49 Table C.3 Results of the analytical and the Monte Carlo meth
34、od 50 Table D.1 List of quantities used in formula (D.1) . 51 Table D.2 List of data available for the input quantities of formula (D.1) . 52 Table D.3 Example of an uncertainty budget for the calibration of a radon monitor according to IEC 61577, see text for details 52 Table E.1 Example of an unce
35、rtainty budget for a surface emission rate measurement according to IEC 60325:2002, see text for details . 57 Table E.2 Example of an uncertainty budget for a surface emission rate measurement according to IEC 60325:2002 for the determination of the uncertainty at a measured value of zero 58 IEC TR
36、62461:2015 IEC 2015 5 INTERNATIONAL ELECTROTECHNICAL COMMISSION _ RADIATION PROTECTION INSTRUMENTATION DETERMINATION OF UNCERTAINTY IN MEASUREMENT FOREWORD 1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising all national electrotechnical
37、committees (IEC National Committees). The object of IEC is to promote international co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and in addition to other activities, IEC publishes International Standards, Technical Specifications, Tech
38、nical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with may participate in this preparatory work. International, governme
39、ntal and non- governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for Standardization (ISO) in accordance with conditions determined by agreement between the two organizations. 2) The formal decisions or
40、agreements of IEC on technical matters express, as nearly as possible, an international consensus of opinion on the relevant subjects since each technical committee has representation from all interested IEC National Committees. 3) IEC Publications have the form of recommendations for international
41、use and are accepted by IEC National Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any misinterpretation by any end user. 4) In order to pr
42、omote international uniformity, IEC National Committees undertake to apply IEC Publications transparently to the maximum extent possible in their national and regional publications. Any divergence between any IEC Publication and the corresponding national or regional publication shall be clearly ind
43、icated in the latter. 5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any services carried out by independent certification bodies.
44、6) All users should ensure that they have the latest edition of this publication. 7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and members of its technical committees and IEC National Committees for any personal injury, property dam
45、age or other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC Publications. 8) Attention is drawn to the Normative references cited in this pub
46、lication. Use of the referenced publications is indispensable for the correct application of this publication. 9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of patent rights. IEC shall not be held responsible for identifying any or all
47、such patent rights. The main task of IEC technical committees is to prepare International Standards. However, a technical committee may propose the publication of a technical report when it has collected data of a different kind from that which is normally published as an International Standard, for
48、 example “state of the art“. IEC 62461, which is a technical report, has been prepared by subcommittee 45B: Radiation protection instrumentation, of IEC technical committee 45: Nuclear instrumentation. This second edition of IEC TR 62461 cancels and replaces the first edition, published in 2006, and
49、 constitutes a technical revision. The main changes with respect to the previous edition are as follows: add to the analytical method for the determination of uncertainty the Monte Carlo method for the determination of uncertainty according to supplement 1 of the Guide to the Expression of uncertainty in measurement (GUM S1), and add a very simple method to judge whether a measured result is significantly different from zero or not based on ISO 11929. 6 IEC TR 62461:2015 IEC 20